We use cookies to distinguish you from other users and to provide you with a better experience on our websites. Close this message to accept cookies or find out how to manage your cookie settings.
To save content items to your account,
please confirm that you agree to abide by our usage policies.
If this is the first time you use this feature, you will be asked to authorise Cambridge Core to connect with your account.
Find out more about saving content to .
To save content items to your Kindle, first ensure no-reply@cambridge.org
is added to your Approved Personal Document E-mail List under your Personal Document Settings
on the Manage Your Content and Devices page of your Amazon account. Then enter the ‘name’ part
of your Kindle email address below.
Find out more about saving to your Kindle.
Note you can select to save to either the @free.kindle.com or @kindle.com variations.
‘@free.kindle.com’ emails are free but can only be saved to your device when it is connected to wi-fi.
‘@kindle.com’ emails can be delivered even when you are not connected to wi-fi, but note that service fees apply.
Transcatheter device closure has become the established standard of care for suitable atrial septal defects. Device erosion has been a recent focus and has prompted changes in the Instructions for Users documentation released by device companies. We reviewed our entire local experience with atrial septal defect device closure, focussing on the evolution of this procedure in our centre and particularly on complications.
Methods
We carried out a retrospective review of 581 consecutive patients undergoing attempted transcatheter device closure of an atrial septal defect in Auckland from December 1997 to June 2014. We reviewed all complications recorded and compared our outcomes with the current literature. We sought to understand the impact of the evolution in recommendations and clinical practice on patient outcomes in our programme.
Results
There were a total of 24 complications (4.1%), including 10 device embolisations (1.7%), nine arrhythmias (1.5%), two significant vascular access-related complications (0.3%), one device erosion (0.2%), one malposed device (0.2%), and one probable wire perforation of the left atrial appendage (0.2%). There was one mortality related to device embolisation. All device embolisations occurred following the change in Instructions for Users after publication of the first device erosion report in 2004. This increase in embolisation rate was statistically significant (p-value 0.015).
Conclusions
In our series, the incidence of device embolisation was higher than that anticipated, with a significant increase following changes to the Instructions for Users. This highlights the need for ongoing data collection on complication incidence and for ongoing review of the impact of changes in clinical practice on complication rates.