Evaluation of air cavities on dose distributions with air-filled apparatuses having different volumes using Gafchromic EBT3 films in brachytherapy

Published online by Cambridge University Press:  15 August 2018

Mehmet Bahadır Çelik
Affiliation:
Department of Radiation Oncology, Mersin State Hospital, Mersin, Turkey
Nezahat Olacak
Affiliation:
Department of Radiation Oncology, Ege University Faculty of Medicine, İzmir, Turkey
Songül Barlaz Us*
Affiliation:
Department of Radiation Oncology, Mersin University Faculty of Medicine, Mersin, Turkey
Emin Tavlayan
Affiliation:
Department of Radiation Oncology, Ege University Faculty of Medicine, İzmir, Turkey
*
Author for correspondence: Songül Barlaz Us, Mersin University, 33100, Mersin, Turkey, E-mail: barlaz@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Aim

The data used in brachytherapy planning are obtained from homogeneous mediums. In practice, the heterogeneous tissues and materials affect the dose distribution of brachytherapy. It is aimed to investigate the effect of air cavities on brachytherapy dose distribution using a specially designed device.

Material and methods

In this study, the special device designed with different volumes of air and water to be irradiated and measured at different depths using EBT3 Gafchromic films. EBT3 Gafchromic films were preferred for this study because they can be cut to the shape of the experimental geometry, are water resistance and double directional usability.

Results

In our study, sudden dose increases and decreases were observed at the water–air–water interfaces. Increases were 9, 11·8 and 15% in the 13, 18 and 22 mm apparatus, respectively. These effects were expected and the results were consistent with the literature and within the tolerance limits stated in the clinical dose guidelines. The most important result is that the percent depth–dose curve of the radiation passing through the air to the water and only passing through the water medium is different. The average differences were 1·97, 2·97 and 2·31% for the 13, 18 and 22 mm apparatus, respectively.

Conclusion

Although the effect of heterogeneity may be neglected according to clinical guidelines, it is suggested that the dose effect of heterogeneity is taken into account so that the dose can be estimated sensitively. Brachytherapy plans using dose data without considering air gaps may cause erroneous dose distributions due to heterogeneity of tissue.

Type
Original Article
Copyright
© Cambridge University Press 2018 

Introduction

Brachytherapy (BT) plays an important role in the curative management of cancers in a variety of disease sites, most prominently, gynaecological malignancies, breast cancer and prostate cancer.Reference Beauleu, Tedgren and Carrier 1 BT provides advantages over external beam radiotherapy in that the tumour can be treated with very high doses of localised radiation, while reducing the probability of unnecessary damage to the surrounding healthy tissues. The successful and effective treatment with lesser side effects for healthy tissue strongly depends on the dose distribution in the tumour and surrounded healthy tissues. Dose distributions depend on photon energies, atomic number, density and electron density in BT.Reference Chiavassa, Buge and Herve 2 Reference Zhang and Das 4

The human body is not homogenous. It consists of different types of tissues like bone, liver, lung, spleen, muscle, air cavities, etc. But TG-43 formalism for treatment planning arise from the neglect of applicator, shielding and contrast solution perturbations, scatter dose changes in the absence of a full scatter medium, and tissue heterogeneity effects.Reference Rivard, Coursey and DeWerd 5 Reference Mashouf, Lechtman and Lai 8 In these circumstances the dose distribution in the patient cannot be accurately assessed.Reference Moura, Micka and Hammer 9 , Reference Palmer, Di Pietro and Alobaidli 10 Actually the isodose distribution is affected by the heterogeneous tissues and materials, because backscatter electrons increase the dose between different density mediums.Reference Ravikumar and Lakshminarayana 11

Although recently, three-dimensional computed tomography based treatment planning for BT applications has been popularly adopted, as it shows more realistic dosimetric outcomes in patient anatomy, there are only a few study that have considered the impact of tissue heterogeneity in BT.Reference Ravikumar and Lakshminarayana 11 Qualitative recognition of the effect of the heterogeneous medium in BT is very useful in assessing the dose distribution in the patient. Therefore, in this study, a special designed device was constructed and irradiated with Iridium-192 (Ir-192), a radioactive source of high dose rate (HDR) BT, and the effects of body air cavities on dose distributions were evaluated with air-filled apparatuses having different volumes using EBT3 Gafchromic films. EBT3 Gafchromic film is water resistant, has double directional usability, weak energy dependent, improved film uniformity, short time period of post-exposure optical density growth, high dose sensitivity, requires no chemical processing and is insensitive to ambient light. The most important feature of EBT3 Gafchromic film is that it can be cut to the exact shape of the experimental geometry. Therefore EBT3 Gafchromic film was used in the study.Reference Gholami, Mirzaei and Jabbary Arfaee 12 , Reference Chiu-Tsao, Medich and Munro 13

Materials and Method

Design of experimental setup

A special designed device was prepared and was constructed, the polypropylene has a close density to water (0·909 g/cm3) and considered water equivalent, between the tubes were air gaps. The cylindrical apparatus was prepared with fixing proper junctions and adding inside styrofoam size of 2×20×30 cm and separately into 13, 18 and 22 mm. Each catheter which has 2 mm diameter was placed 1 cm laterally from the apparatus surface inside styrofoam. The styrofoam was used for immobilised the catheters and apparatuses. And then, a special device was placed in the water-filled plastic box (Figure 1).

Figure 1 Special designed device for measuring the effect of air cavities on brachytherapy dose distributions.

BT planning

Before starting the BT planning, computer tomography (CT) images of special designed device were obtained with 5 mm slice thickness for using BT planning system (Asteion, TSX-021B, Toshiba, Japan) and CT images were transferred to treatment planning system (TPS-Oncentra, Elekta, The Netherlands). The length of each catheter was measured with a check ruler. The starting point of the catheters was marked and the irradiation distance was determined as 4 cm with 5 mm steps. The origin was defined at the centre of the catheter on the X, Y, Z coordinates. A dose normalisation point was marked at a distance of 2·5 cm from the midpoint of the 4 cm catheter length (origin) to the periphery and the dose was optimised to give 700 cGy to evaluate the change in water–air–water interfaces on a single film. And then, the BT planning was loaded on the BT (MicroSelectron HDR, Nucletron, The Netherlands) (Figure 2).

Figure 2 Dose distributions on the X, Y and Z axes (for the 13, 18 and 22 mm apparatuses, respectively).

Irradiation

On the Gafchromic EBT3 films, the holes were opened to fit the diameters of the apparatus and catheter. The film was placed in the centre of the 4 cm catheter length. According to the plan made in the TPS, the special designed device was irradiated with Ir-192 source (Figure 3). The measurement was repeated three times to increase accuracy for each apparatus.

Figure 3 Irradiation of the special designed device.

Acquisition of data

The irradiated films were scanned using the Landscape method and red colour channel mode using a film scanner (Epson Perfection V750 Pro, Epson, USA). The ‘OmniPro IMRT’ programme was used to generate percent depth dose (PDD) curves from the scanned films. The average values of the three measurements were calculated. PDD curves at various 2-mm increment distances from the centre of the catheter, depending on the diameter lengths of the cylinders, were obtained from films with and without apparatus and the differences in PDD values between the air–water interface were calculated. This process repeated for each apparatus of 13, 18 and 22 mm. Due to damage to the film that occurred when creating openings for their insertion through the polypropylene apparatus, the dose values obtained from the vicinity of the hole as well as the dose densities originating from the scattered on the sides of the hole were not included in the measurements.

Results

Isodose distributions according to air spaces simulated with 13 mm (a), 18 mm (b) and 22 mm (c) cylinders are showed in Figure 4. As can be seen from the Figure 4, when the air gap increases, the deformation in the shape of the circle becomes more pronounced. The PDD curves in the X direction are given for 13 mm (a), 18 mm (b) and 22 mm (c) cylinders in Figure 5. In Figure 5, X axes show the axis length from the centre of the catheter and Y axes represent the PDD. According to these PDD distributions, the radiation emitted from the Ir-192 source is less than the dose calculated according to the inverse square law. The intensity of a wave transitioning from water to air is increased by the backscattering of radiation at the interface. The dose increases are 2, 3 and 4% for 13, 18 and 22 mm cylinder devices, respectively. Hence, the dose increased as the air gap widened. The signal plateaued after the first transitional increase; the plateau distances were 0·10, 0·12 and 0·14 mm for the 13, 18 and 22 mm devices, respectively, when including the air-equivalent polypropylene. These values were small and negligible. When a wave travels from air to water, the higher density of water slows the wave and thereby increases its interaction with the medium. We found that a plateau formed at the interface between the air and water (0·06 mm for the 13 mm device, 0·07 mm for 18 mm device and 0·10 mm for 22 mm device); however, even when neglecting this plateau zone, the wave was exponentially absorbed by the sudden change in medium; these increases were 9, 11·8 and 15% in the 13, 18 and 22-mm devices, respectively, indicating that the signal increased in direct proportion to the air gap.

Figure 4 Isodose distributions according to air spaces simulated with 13 mm (a), 18 mm (b) and 22 mm (c) cylinders.

Figure 5 The per cent depth–dose curves in devices with simulated air gaps using 13 mm (a), 18 mm (b) and 22 mm (c) cylinders [X axes: length from the centre of catheter (cm), Y axes: per cent depth dose].

When the radiation transverses through the air medium and reaches the water medium, the exponential decrease in the depth dose is different from that of the water medium. The average differences were 1·97, 2·97 and 2·31% for the 13, 18 and 22 mm devices, respectively.

Discussion

When radiation beams travel through a homogenous medium, the dose decreases in inverse proportion to the square of the distance from the source. However, when radiation passes through mediums of different atomic number, density and electron density, dose fluctuations are caused. If radiation passes through an interface between materials with low atomic number and a high atomic number, a significant fraction of the secondary electrons is backscattered toward the interface by the tissue with the high atomic number. This effect considerably increases the dose at the interface.Reference Binger, Seifert, Blass, Bormann and Rücker 14 Such changes in dose intensity are most apparent at the interfaces between mediums of different densities. Transitioning from a low-density medium (air) to a dense medium (water) causes a sudden drop in dosage at the interface. The reverse situation results in dose increases at the interface. Anatomically, such disruptions cause a breakdown in dose distribution by increasing the PDD values within the BT treatment area.Reference Engelsman, Damen, Koken, Veld, Ingen and Mijnheer 15 , Reference Azam, Harter, Thobejone and Bertrant 16

Zabihzadeh et al.Reference Zabihzadeh, Yadollahpour and Kargar 17 have found air heterogeneity dose increases as 9·11–10·2, 9·11–10·0, 8·62–10·08 and 8·5–10·07%, respectively, for 5, 10, 20 and 30 mm distances from the source. In a study by Terribilini et al.,Reference Terribilini, Manser, Frei, Volken, Mini and Fix 18 air at a distance of 10 mm from the source produced heterogeneity that was 7% higher than the corresponding value of the control water medium. In a similar study, Chandola et al.Reference Chandola, Tiwari, Kowar and Choudhary 19 have found that the corresponding doses behind heterogeneities were 5·5–6·5% higher and 4·5–5% lower than water phantom values in the presence of air and cortical bone, respectively. Graf et al.Reference Graf, Scanderbeg, Yashar and Jiang 20 have showed a heterogeneous dose increase of 10% at a distance of 2 mm above the water–air interface and a dose decrease of 30% in the homogeneous medium at distances of under 2 mm.

In our study, sudden dose increases and decreases were observed at the water–air–water interfaces. These effects were expected and the results were consistent with the literature and within the tolerance limits stated in the clinical dose guidelines. The most important point of our results is that the PDD curve of the radiation in the air medium through to the water medium is different from the only water medium.

Conclusion

Even if the heterogeneous dose changes according to the AAPM TG-43 protocol are neglected but according to results of our study, heterogenity changes in dose distributions should be taken into consideration in BT plannings. It is recommended that the dose effect of air gaps be considered so that the dose can be accurately estimated.

Acknowledgement

None.

References

1. Beauleu, L, Tedgren, AC, Carrier, CF et al. Report of the Task Group 186 on model-based dose calculation methods in brachytherapy beyond the TG-43 formalism: current status and recommendations for clinical implementation. Med Phys 2012; 39 (10): 62086236.Google Scholar
2. Chiavassa, S, Buge, F, Herve, C et al. Monte Carlo evaluation of the effect of inhomogeneities on dose calculation for low energy photons intra-operative radiation therapy in pelvic area. Phys Med 2015; 31: 956962.Google Scholar
3. Adelnia, A, Fatehi, D. Estimation and evaluation of tissue inhomogeneity effect on dose distribution for high dose rate iridium 192 source using Monte Carlo simulation and film dosimetery. J Nucl Med Radiat Ther 2016; 7 (6): 16.Google Scholar
4. Zhang, H, Das, IJ. Dosimetric perturbations at high-Z interfaces with high dose rate 192Ir source. Phys Med 2014; 30: 782790.Google Scholar
5. Rivard, MJ, Coursey, BM, DeWerd, LA et al. Update of AAPM Task Group No. 43 Report: A revised AAPM protocol for brachytherapy dose calculations. Med Phys 2004; 31 (3): 633674.Google Scholar
6. Poon, ES. Patient-specific dose calculation methods for high-dose-rate iridium-192 brachytherapy. McGill University, Montréal, Ph.D. Thesis, 2009.Google Scholar
7. Hofbauer, J, Kirisits, C, Resch, A et al. Impact of heterogeneity-corrected dose calculation using a grid-based Boltzmann solver on breast and cervix cancer brachytherapy. J Contemp Brachytherapy 2016; 8 (2): 143149.Google Scholar
8. Mashouf, S, Lechtman, E, Lai, P et al. Dose heterogeneity correction for low-energy brachytherapy sources using dual-energy CT images. Phys Med Biol 2014; 59: 53055316.Google Scholar
9. Moura, ES, Micka, JA, Hammer, CG et al. Development of a phantom to validate high-dose-rate brachytherapy treatment planning systems with heterogeneous algorithms. Med Phys 2015; 42 (4): 15661574.Google Scholar
10. Palmer, AL, Di Pietro, P, Alobaidli, S et al. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic. Med Phys 2013; 40 (6): 110.Google Scholar
11. Ravikumar, B, Lakshminarayana, S. Determination of the tissue inhomogeneity correction in high dose rate Brachytherapy for Iridium-192 source. J Med Phys 2012; 37 (1): 2731.Google Scholar
12. Gholami, S, Mirzaei, HR, Jabbary Arfaee, A et al. Dose distribution verification for GYN brachytherapy using EBT Gafchromic film and TG-43 calculation. Rep Pract Oncol Radiother 2016; 21 (5): 480486.Google Scholar
13. Chiu-Tsao, ST, Medich, D, Munro, J. The use of new GAFCHROMIC EBT film for 125I seed dosimetry in Solid Water phantom. Med Phys 2008; 35 (8): 37873799.Google Scholar
14. Binger, T, Seifert, H, Blass, G, Bormann, KH, Rücker, M. Dose inhomogeneities on surfaces of different dental implants during irradiation with high-energy photons. Dentomaxillofac Radiol 2008; 37: 149153.Google Scholar
15. Engelsman, M, Damen, EMF, Koken, PW, Veld, AAV, Ingen, KMV, Mijnheer, BJ. Impact of simple tissue inhomogeneity correction algorithms on conformal radiotherapy of lung tumours. Radiother Oncol 2001; 60: 299309.Google Scholar
16. Azam, NR, Harter, KW, Thobejone, S, Bertrant, K. Air cavity effects on the radiation dose to the larynx using Co-60, 6 MV and 10 MV photon beam. Int J Radiat Oncol Biol Phys 1994; 29: 11391146.Google Scholar
17. Zabihzadeh, M, Yadollahpour, A, Kargar, L. The effects of tissue heterogeneitieson dose distribution of iridium-192 source in brachytherapy treatments. Biomed Pharmacol J 2013; 6 (2): 205213.Google Scholar
18. Terribilini, D, Manser, P, Frei, D, Volken, W, Mini, R, Fix, M. Implementation of a brachytherapy Ir-source in an in-house system and comparison of simulation results with EGSnrc, VMC++ and PIN. J Phys Conf Ser 2007; 74: 012022.Google Scholar
19. Chandola, R, Tiwari, S, Kowar, M, Choudhary, V. Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation. J Cancer Res Ther 2010; 6 (1): 5457.Google Scholar
20. Graf, M, Scanderbeg, D, Yashar, C, Jiang, S. Monte Carlo dose comparison assessing material inhomogeneity effects in breast brachytherapy. Med Phys 2010; 37 (6): 1.Google Scholar
Figure 0

Figure 1 Special designed device for measuring the effect of air cavities on brachytherapy dose distributions.

Figure 1

Figure 2 Dose distributions on the X, Y and Z axes (for the 13, 18 and 22 mm apparatuses, respectively).

Figure 2

Figure 3 Irradiation of the special designed device.

Figure 3

Figure 4 Isodose distributions according to air spaces simulated with 13 mm (a), 18 mm (b) and 22 mm (c) cylinders.

Figure 4

Figure 5 The per cent depth–dose curves in devices with simulated air gaps using 13 mm (a), 18 mm (b) and 22 mm (c) cylinders [X axes: length from the centre of catheter (cm), Y axes: per cent depth dose].