Review on the feasibility of using PRESAGE® dosimeter in various radiotherapy techniques

Published online by Cambridge University Press:  16 March 2020

Qurat-ul-ain Shamsi*
Affiliation:
Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Saeed Ahmad Buzdar
Affiliation:
Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Shagufta Jabeen
Affiliation:
Bioloby Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan
Khalid Iqbal
Affiliation:
Clinical and Radiation Oncology Department, Shaukat Khanum Memorial Cancer Hospital and Research Centre, Lahore, Pakistan
*
Author for correspondence: Qurat-ul-ain Shamsi, Physics Department, The Islamia University of Bahawalpur, Bahawalpur, Punjab, Pakistan. Tel: +923216827959. E-mail: annieshamsi@hotmail.com
Rights & Permissions [Opens in a new window]

Abstract

The emergence of advanced radiotherapy techniques, such as intensity-modulated radiotherapy (IMRT), brachytherapy, conformal radiotherapy, magnetic resonance-guided radiotherapy (MRgRT), stereotactic synchrotron radiotherapy (SSRT) and microbeam radiotherapy (MRT), has increased the importance of the verification of volumetric dose distribution. The verification of dose distribution is usually done by 2D films and 3D gel dosimeters, but PRESAGE® due to its affordability, reproducibility, precision, accuracy, unique dosimetric and physical properties is considered as an effective candidate in providing 3D dose data. PRESAGE® is insensitive to oxygen contamination, machinable and can be molded to a variety of shapes and sizes. It is absorbing rather than scattering light which facilitates high-accuracy readout by optical computed tomography (OP-CT). This review focuses on the feasibility of using PRESAGE® in various complicated radiotherapy techniques by comparing its measured doses with 2D films and treatment planning system (TPS) calculated doses.

Type
Literature Review
Copyright
© The Author(s), 2020. Published by Cambridge University Press

Introduction

The introduction of advanced radiotherapy techniques has escalated the significance of validating 3D dose distribution specified by steep dose gradients and complicated shapes. Among various other dosimeters such as polymer and Fricke gels, required for 3D dosimetry, PRESAGE® (Heuris Pharma LLC, Skillman, NJ, USA) due to its dosimteric and physical properties is chosen in 3D dosimetry to give volumetric dose information. Reference Khezerloo, Nedaie and Takavar1 PRESAGE is a radiochromic 3D dosimeter that was first fabricated in 2006. Reference Oldham, Godfrey, Das and Wolbars2 It is basically a plastic material which is composed of leuco dye and an optically clear, solid polyurethane matrix with free radical initiators. Halogens and leucomalachite green (LMG) are commonly used as radical initiators and leuco dye, respectively. Fabrication details of PRESAGE® are documented in previously published studies and also in US Patent Application Publication No. 2007/0020793A1. Reference Juang, Newton, Niebanck, Benning, Adamovics and Oldham3Reference Clift, Thomas, Adamovics, Chang, Das and Oldham5

When radiation beam is administered to PRESAGE®, oxidation of LMG in malachite green occurs due to the production of free radicals by radical initiators and local optical density (OD) of the dosimeter changes. Reference Newton, Thomas, Ibbott and Oldham6 In the literature presented by Adamovics et al., the change in OD is directly proportional to the amount of radiation dose delivered and can be determined for cuvettes by spectrophotometer and for larger volumetric dosimeter by optical computed tomography (OP-CT) scanners. Reference Adamovics and Maryanski7

PRESAGE® possesses numerous significant advantages over other polymer and Fricke gel dosimeters. Presence of polyurethane matrix enables the PRESAGE® dosimeter to retain its solid shape and insensitivity to atmospheric oxygen. Reference Alqathami, Blencowe and Ibbott8 This eliminates the need of any supporting chamber. Furthermore, it has also become convenient to fabricate PRESAGE® into various shapes by using different molds. Bache et al. have interpreted the possibility of molding PRESAGE® into different shapes and producing regions of various densities in their investigation. Reference Bache, Juang and Adamovics9 The spontaneous polymerisation of PRESAGE® is not affected by the temperature below 80°C. Reference Farhood, Khezerloo, Zadeh, Nedaie, Hamrahi and Khezerloo10 Various investigations by Guo et al., Sakhalkar et al. and Wang et al. have described the dosimetric properties of PRESAGE® dosimeters and mentioned its suitability as a relative dosimeter in clinical applications. According to these studies, the dosimeter response remained linear till 100 Gy dose, insignificant dependence of photon beam on dose rates varying from 100 to 600 cGy/min. Reference Guo, Adamovics and Oldham11Reference Jackson, Juang, Adamovics and Oldham17

This study is intended to review previous work on the feasibility of PRESAGE® dosimeter in various radiotherapy techniques such as intensity-modulated radiotherapy (IMRT), brachytherapy, 3D conformal partial radiotherapy, magnetic resonance-guided radiotherapy (MRgRT) and stereotactic radiotherapy (SRT). For this purpose, dose distribution measured by PRESAGE® was verified by comparing it with external beam therapy (EBT) film measured and treatment planning system (TPS) calculated dose distribution. Isodose line profiles, dose volume histograms (DVHs) and gamma maps were used as tools for evaluating the performance of PRESAGE® dosimeter in various radiation treatment techniques.

Dosimetric Verification Using GAFCHROMIC EBT Film

The purpose of including independent dose measurement by EBT film is to resolve discrepancies between PRESAGE® measured and TPS calculated doses. In radiotherapy, EBT films are mainly used for the verification of dose distribution. Reference Casanova Borca, Pasquino and Russo18 Gafchromic EBT films play a tremendous role in verifying and assuring the quality of treatment plans. Reference Devic19 There is a large number of publications on the properties of EBT films like their homogeneity, dose response, post-colouration behaviour, absorption spectra, sensitivity to light, temporal stability, directional independence and convenience of self-developing. Reference Brown, Bailey, Willowson and Baldock20Reference Devic, Seuntjens and Sham30

Linearity in dose response of PRESAGE dosimeter

Dose response linearity of PRESAGE® dosimeter has been confirmed in the previously published studies. Reference Zhao, Newton, Oldham, Das, Cheng and Adamovics31Reference Abdul Rahman, Bräuer-Krisch and Brochard33 In accordance with the investigation by Iqbal et al., breast-shaped anthropomorphic PRESAGE® dosimeter with EBT2 films inserted was irradiated with 6 MV beam with the 5-field IMRT plan. PRESAGE® had Z eff and physical density of 7·6 and 1·07 g/cm3, respectively. It was scanned with Duke midsized optical scanner dedicated for RPC (Duke University, Durham, NC, USA). Reference Newton, Thomas, Ibbott and Oldham21 OP-CT scans of PRESAGE® did not give rise to any variation in the OD of the PRESAGE®. In Figure 1, the radiochromic response was found linear showing sensitivity of 0·0059 OD change for a path length of 1 mm. PRESAGE® dosimeter showed a coefficient of variation of 1·0%. EBT2 had 0·75% coefficient of variation. Reference Iqbal, Gifford, Ibbott, Grant and Buzdar23

Figure 1. OD to dose curve for PRESAGE (a); OD to dose curve for EBT2 film (b).

Further studies by Iqbal et al. on brachytherapy, there appeared an uncertainty of 0·8% in OD, estimated for breast-shaped PRESAGE® dosimeter and 0·7% for EBT2 film. A Strut-Adjusted Volume Implant (SAVI) device was used for irradiating the PRESAGE® dosimeter. While calibrating PRESAGE®, radiation dose was delivered to the small volumes from the same batch of breast-shaped PRESAGE® with an intent to detect errors when comparing dose distribution. Reference Aland, Karin and Kenny34,Reference MacDougall, Miquel and Keevil35 The volume effect uncertainty of PRESAGE® necessitates the utilisation of normalisation as to compare dose distribution of PRESAGE® with Oncentra TPS and EBT2 film. Reference Iqbal, Ibbott, Lafratta, Gifford, Akram and Buzdar36

For the two film planes, the maximum dose difference between PRESAGE® and Oncentra TPS was 4·5% and between Oncentra TPS and EBT2 film was 2·42%. Further, mean percentage differences of skin dose between PRESAGE® and Oncentra TPS were 3·54 and 2·85% for films 1 and 2, respectively. At various chosen number of points, mean percentage differences between Oncentra and EBT2 film were 0·97 and 0·53% for films 1 and 2, respectively (Figure 2).

Figure 2. (a) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) and EBT2 film for film 1. (b) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS and EBT2 film for surface film 2.

In 3D conformal radiotherapy, radio-chromic response of an EBT2 film was linear with sensitivity of 0·0057OD variation for a path length of 1 mm. The uncertainty in the net OD was 1·8% in case of EBT2 films with a reference dose of 3 Gy. The sources of errors responsible for introducing uncertainty in OD have been discussed in detail in the literature [3-field]. In the case of PRESAGE®, there was an uncertainty of 0·8% for the same reference dose. The power of dose–response linearity of PRESAGE® did not cause any limitation on the data analysis. Reference Iqbal, Ibbott, Grant, Gifford and Buzdar37

According to the exploration of Gye Won Choi in MRgRT, dose response curve of EBT3 film illustrated the difference of less than 2% in net OD between the two cases of 0 magnetic field and when adjusted to 1·5 T. Reference Choi38 The performance of PRESAGE® was evaluated by establishing EBT3 film as a standard. PRESAGE® curve displayed under dose response of 9% in the presence of 1·5 T magnetic field. The dose response of PRESAGE® dosimeter showed linearity when used as a relative dosimeter indicating that magnetic field effects would not disturb its dosimetry in this case.

Verification of PRESAGE® Dosimeter in IMRT

Gamma analysis

Gamma analysis is a well-established verification tool required to compare the agreement between the measured and computed dose distributions. Reference Low, Harms, Mutic and Purdy39 This analysis is essential in evaluating the performance of radiotherapy clinics to deliver the prescribed doses. For research purposes, gamma maps can do an assessment of the efficiency of new dosimeters against standard dosimeters or treatment plans.

Mark Oldham et al. delivered 11-field IMRT plan to the cylindrical PRESAGE® of 16 cm in diameter and 11 cm in height. PRESAGE® measured dose distributions were compared with EBT and Eclipse TPS (Varian Medical Systems, Palo Alto, CA, USA) calculated distributions. Isodose line profiles, gamma maps and DVHs were employed to evaluate the agreement among three dose distributions. Reference Oldham, Sakhalkar, Guo and Adamovics40 A good agreement was found with a maximum difference of 3% among them at all points excluding within 3 mm outer ring due to edge artefacts. Gamma pass ratio was 96% between the Eclipse and PRESAGE® measured dose distributions. Guo et al. evaluated the functionality of PRESAGE for 5-field radiotherapy by comparing the agreement among the Eclipse TPS calculated, PRESAGE® and EBT film measured dose distributions. The gamma criteria of 4% dose difference and 4 mm distance to agreement (DTA) appeared. Owing to the presence of edge artefacts, the outer layer comprising 10% of the radius of PRESAGE® dosimeter was not included. Reference Guo, Adamovics and Oldham41 From the investigation by Oldham et al., gamma criteria of 3%/3 mm occurred with a pass rate of 96% for Eclipse TPS versus PRESAGE® dosimeter without considering edge artefacts in 11-field IMRT. Even taking into account the region of edge artefact, PRESAGE® dosimetric characteristics showed similarity with that of film dosimeters. The gamma pass rates for PRESAGE® versus Eclipse and EBT film versus Eclipse were 91·4 and 94%, respectively. Reference Oldham, Sakhalkar, Guo and Adamovics42 Sakhalkar et al. achieved a good agreement among EBT film, PRESAGE® and Eclipse TPS with passing rate of more than 98% using head and neck phantom in IMRT. This analysis had a gamma criteria of 4%/3 mm by excluding 4 mm outer rim of the PRESAGE® due to edge artefacts. Reference Sakhalkar, Adamovics, Ibbott and Oldham43 Thomas and Newton et al. achieved a gamma criteria of 3%/3 mm with 97·9% passing rate for PRESAGE® versus TPS in 4-field IMRT. Reference Thomas, Newton, Adamovics and Oldham44 Moutsatsos et al. employed PRESAGE® dosimeter for verifying dose distribution generated in helical tomotherapy using small fields for head and neck treatment plans. A comparison was made among TPS calculated, PRESAGE® and film measured dosimetry results. Gamma index pass rates were achieved more than 90% and uncertainties remained within 2%. Reference Moutsatsos, Pantelis and Antypas45

Iqbal et al., for 2D axial gamma comparison, determined the pass rates of 91·2, 90·6 and 88·4% for EBT2 versus Pinnacle, PRESAGE® versus Pinnacle and EBT2 versus PRESAGE®, respectively, in breast IMRT. Reference Iqbal, Gifford, Ibbott, Grant and Buzdar23 In all comparisons, failures were found near the dosimeter periphery in the 8 mm outer layer of the PRESAGE®. In this area, PRESAGE® measured doses are possibly imprecise owing to edge artefacts, and Pinnacle determined doses are likely to be erroneous due to complexity in modelling the buildup region. Without this 8 mm outer layer, gamma pass rates increase to 95% for PRESAGE® versus Pinnacle comparison (Figure 3).

Figure 3. Gamma maps (a) (±3%/3 mm) between Pinnacle, EBT2 and PRESAGE for a region of PTV 5 mm on film plane; (b) (±3%/3 mm) in the transverse and sagittal planes for PTV 5 mm intersecting film plane.

DVH analysis

From studies by Iqbal et al., in 5-field breast IMRT plan, DVHs of the planning target volume (PTV) determined by PRESAGE® were different from that calculated by the Pinnacle TPS (Philips Radiation Oncology Systems, Fitchburg, WI, USA). Reference Iqbal, Gifford, Ibbott, Grant and Buzdar23 In addition to this, a slightly less homogeneity was observed in PRESAGE® determined DVH than that calculated by Pinnacle TPS. This is due to aberrations occurring at the edges of the breast-shaped PRESAGE®. A maximum dose difference of 5% for the PTV DVH was seen at 5 and 95% of the fractional volume. For PTV 1 mm, PTV 3 mm and PTV 5mm, sub-volume DVHs showed a maximum dose difference of 3, 2 and 1%, respectively. DVH of the breast-shaped PRESAGE showed a deviation from Pinnacle calculated DVH occurring between 0·5 and 3 Gy dose. Edge artefacts are responsible for this discrepancy because the breast region of interest encircles the whole PRESAGE.

Comparison between isodose line profiles of PRESAGE, TPS and EBT film

In IMRT by Iqbal et al., Pinnacle TPS computed, PRESAGE® and EBT2 film measured line dose distribution gave a good agreement among them with a maximum dose difference of 5% which occurs at the periphery of the breast. Reference Iqbal, Gifford, Ibbott, Grant and Buzdar23 However, it is complicated to investigate about the dose distribution of which dosimeter (PRESAGE® and EBT) correlates more with Pinnacle calculated distribution. Moreover, within 80% of the central field width, a maximum difference of 2% was produced among three line dose distributions. PRESAGE® and EBT2 film measured dose distributions are related to two independent deliveries of the same radiotherapy plan. Line profiles are presented in Figure 4.

Figure 4. Line profiles of Pinnacle, PRESAGE and EBT2 film dose distributions from axial slices.

PRESAGE Dosimeter in Brachytherapy

Verification through dose profiles

In 2010, Pierquet et al. determined a satisfactory agreement between PRESAGE® measured and Eclipse calculated isodose line profiles in brachytherapy. PRESAGE® dosimeter with Z eff of 8·1 was irradiated with 7 Gy dose at 1·5 cm from the centre of the source, and line dose distributions were obtained along the transverse axis from the planes. Edge artefacts were present in the ± 3 mm region from the source centre, but on the whole measured and calculated dose distributed agreed well with each other in the region between 5 and 20 mm from the source centre. Reference Pierquet, Craciunescu, Steffey, Song and Oldham46 In the studies by Iqbal et al. in 2018, for taking measurements of isodose profiles, PRESAGE® was cut at its central long axis and EBT2 film pieces were inserted between the two halves of the PRESAGE. The maximum dose difference in dose profiles was 2·48% between Oncentra TPS and PRESAGE® using both films. Between EBT2 film and Oncentra TPS, the maximum difference was 4·06%. The mean dose differences in the line plots of both films 1 and 2 were 2·25 and 1·80% between Oncentra and PRESAGE® and 3·31 and 2·82% between Oncentra and EBT2 film. Reference Iqbal, Ibbott, Lafratta, Gifford, Akram and Buzdar36

DVH of PTV and normal tissue

Iqbal et al. in their studies showed a DVH comparison between Oncentra TPS and PRESAGE® which showed a maximum percentage difference of 4%. This difference occurred because of the reflection artefacts as described in previous studies. Reference Guo, Adamovics and Oldham11 The PTV_EVAL V90 and V95 of breast-shaped PRESAGE® and Oncentra were 97, 96·59% and 99·6, 98·95%, respectively.

The dose distribution determined by the PRESAGE® varied from Oncentra calculated dose distribution particularly near the edges of the dosimeter due to edge artefacts in the PRESAGE® dose distribution. By considering DVH curve of PTV_OPT, PRESAGE®, measured dose distribution was found insignificantly less homogeneous than Oncentra calculated dose distribution because of the presence of small regions of relatively under and over lying closer to the dosimeter edges. A maximum dose difference of 4% appeared at the lower and upper volume ends. From the DVHs of 1, 3 and 5 mm sub-volumes of the normal breast tissue, the maximum dose difference of 3·2, 2·5 and 1·5% occurred between the PRESAGE® and Oncentra treatment plans.

3D Conformal Radiotherapy

In 3D conformal therapy, Guo et al. verified the cylindrical-shaped PRESAGE® dosimeter by comparing its dose distribution with EBT film measured and Eclipse TPS calculated dose distribution. PRESAGE® dosimeter had a Z eff of 8·3, physical density of 1·07 g/cm3 and a OP-CT number of 200, and it was irradiated with 6 MV beam from Varian 21EX linac with a prescription dose of 15 Gy. All distributions were normalised at treatment plan isocentre found centrally in the high-dose region. There found a good agreement among all the three dose distributions with a maximum dose difference of 4%. However, under dosage was noticed at the periphery of the PRESAGE® due to edge artefacts. Reference Guo, Adamovics and Oldham11

Further in 2017, Iqbal et al. analysed the comparison of gamma pass rates between Pinnacle calculated and PRESAGE® measured dose distribution, and it was 97·4% in 3-field partial radiotherapy. An axial 2D gamma map comparisons of EBT2 versus Pinnacle and EBT2 versus PRESAGE® were 95·3 and 97·6%, respectively. About 95·3% pass rate can be increased up to 96%, if the outer boundary of 3 mm is eliminated.

From the investigations by Iqbal et al., in this treatment modality, independent 2D line dose measurements using EBT2 films in two selected levels simplified resolving any dose difference between PRESAGE® and Pinnacle. Reference Iqbal, Ibbott, Grant, Gifford and Buzdar37 A good correlation was found among Pinnacle, PRESAGE® and EBT2 film measured dose distribution showing maximum difference of 1·5% (Figure 5).

Figure 5. Line profiles of the Pinnacle3, PRESAGE® and EBT2 film dose distributions of axial slice.

DVHs comparison between PRESAGE and Pinnacle TPS

DVH curve of the PRESAGE® PTV showed less homogeneous dose distribution than Pinnacle calculated dose distribution with a maximum dose difference of 2·2%. Furthermore, gross tumor volume (GTV) and clinical target volume (CTV) manifested 0·8 and 1·5% dose difference. This difference again happened due to edge artefacts similar to that in other treatment techniques (IMRT and brachytherapy) (Figure 6).

Figure 6. GTV, CTV and PTV DVH comparison between the PRESAGE® and Pinnacle3 Planning dose distributions.

Edge artefacts

Dose distribution measurements with PRESAGE®/OP-CT often undergo deterioration due to edge artefacts. This effect has been reported in detail in the literature by Doran et al. Reference Doran and Yatigammana47 The sensitivity of PRESAGE® is highly affected by any mismatch in the refractive indices between the matching fluid and the dosimeter such that the mismatch of only 0·3% can produce the edge artefact of several millimeters. In the studies involving PRESAGE®/OP-CT dosimetry, edge artefacts regions have been excluded which extend from the outer 3 to 8 mm rim of the PRESAGE dosimeter. Reference Oldham, Sakhalkar, Guo and Adamovics48,Reference Thomas, Newton, Adamovics and Oldham49

Magnetic Resonance Imaging-Guided Radiotherapy

In radiation therapy techniques, which are guided by magnetic resonance imaging, there occurs an alteration in the delivered dose distribution in the presence of strong magnetic field created during magnetic resonance (MR) imaging. In this treatment mode, electron return effect (ERE) becomes more pronounced as it is responsible for the loss of electronic equilibrium and enhancement of dose in the original medium near the interface. This phenomenon is shown to cause major clinical concerns which are presented in detail by Raaijmakers et al. Reference Raaijmakers, Raaymakers and Lagendijk50,Reference Raaijmakers51

Dosimeters response in magnetic field

For the measurement of dose distribution in magnetic field, the response of various dosimeters was investigated. Meijsing et al. determined that by changing the strength of magnetic field from 0 to 1·2 T, the response of farmer-type ionisation chamber varies by 10–15%. Reference Meijsing, Raaymakers and Raaijmakers52 Conventional 2D dosimeters like thermo luminescent dosimeters (TLDs), optically stimulated luminescent dosimeters OSLD and EBT3 film were not affected by magnetic field, but dosimetric results with PRESAGE showed an under dose response. Reference Mathis, Sawakuchi and Flint53Reference Zhao, Newton, Oldham, Das, Cheng and Adamovics57

Validity of PRESAGE dosimeter in MRgRT

Gye Won Choi demonstrated the viability of employing PRESAGE® dosimeters in MRgRT. In this treatment modality, PRESAGE® cuvettes (1 cm x 1 cm x × 4 cm) were irradiated with 1, 4, 7 and 10 k MU by adjusting magnetic field to 1·5 T. From the studies by Mathic et al., magnetic field did not produce any considerable effect on the EBT3 film response, so it was chosen as the standard for the verification of the PRESAGE® dosimeter in this modality. Reference Mathis, Sawakuchi and Flint53 EBT3 and the PRESAGE® measured dose distributions when compared with each other showed a gamma criteria of 5%/3 mm. Despite the modifications in dose distributions caused by ERE, there observed still a satisfactory agreement between EBT3 and PRESAGE® measured doses with passing rate of more than 90%. Reference Choi38

In the radiotherapy techniques currently in use, all the differences among the EBT, PRESAGE measured and TPS calculated doses were found within the tolerance limits of ±5% specified by ICRU and also in the literature. 54Reference Shamsi, Buzdar, Atiq, Atiq, Altaf and Iqbal56

PRESAGE® in Proton Beam Therapy

Zhao et al. determined the dosimetric characteristics of PRESAGE® dosimeter in proton beam therapy. They compared PRESAGE® measured dose profiles with that of ion chamber (IC) in water phantom. PRESAGE® showed a linearity in dose response to proton beam. Reference Zhao, Newton, Oldham, Das, Cheng and Adamovics57 Mitchell Carroll et al. studied on the feasibility of PRESAGE® dosimeter in measuring dose distribution generated by proton beam using anthropomorphic head phantom. A comparison of low linear energy transfer-dependent PRESAGE® measured dose distribution was made with EBT2, radiochromic film, TLD measured and TPS calculated dose distribution. Digital multi-camera optical surveillance system was implemented to read out PRESAGE® dosimetric distribution. Isodose line profiles of PRESAGE® depicted an agreement within 3 and 4 mm in the coronal and sagittal planes, respectively. Beam doses measured by radiochromic film manifested agreement to 2 mm in both the given planes. Gamma pass ratio of 95% was achieved between TPS computed and PRESAGE® measured doses with gamma criteria of 5%/3 mm and 5%/4 mm in both planes. Reference Carrol, Ibbott, Gillin and Adamovics58

PRESAGE® Dosimetry in Stereotactic Synchrotron Radiotherapy

Through investigations by Alqathami et al. and Gorjiara et al., it was determined that minor changes in the chemical composition of PRESAGE® can alter its dosimetric properties to become radiologically equivalent to water with an energy range appropriate for kV and MV dosimetry. Reference Alqathami, Blencowe, Geso and Ibbott59,Reference Gorjiara, Hill and Kuncic60 Jackson et al. and Brady et al. have documented that water equivalent PRESAGE® is capable of verifying 3D dose distribution in IMRT, volumetric modulated arc therapy and external beam gated treatments with 6 MV beam. Reference Brady, Brown, Clift, Yoo and Oldham61,Reference Jackson, Juang, Adamovics and Oldham62 It is reported in the literature by Gagliardi et al. that water-equivalent PRESAGE® is well suited for high-resolution 3D imaging of synchrotron-generated micro beams in stereotactic synchrotron radiotherapy (SSRT). Reference Gagliardi, Liam, Poole, Franich and Moshi63

Gagliardi et al. have investigated in their other exploration about the validation of the radiological response of water equivalent PRESAGE in 3D dosimetry for synchrotron X-ray photon energies. PRESAGE® was shaped into cylindrical rods with 70 mm in length and 43 mm in width. They compared PRESAGE measured dose data with GAFCHROMIC EBT3, IC and Monte Carlo (MC) GEANT4 (Version 9.4.6) calculated data. PRESAGE® measured percentage depth dose (PDD) profiles and IC measured profiles agreed with each other within 2% for 6 and 18 MV beams showing an under response at the entrance of the PRESAGE® due to the affects given in the literature. Reference Alqathami, Blencowe, Geso and Ibbott64 IC measured and MC calculated PDD profiles showed an agreement within 1% with PRESAGE® measured profiles above 5 mm depth using 10 × 10 mm field size. There was an agreement between MC data and PRESAGE® measured data within 2% below 5 mm depth (buildup region). PRESAGE® showed an inability to measure field penumbra, but PRESAGE® possesses the potential to determine dose profiles which were calculated by MC model and verified by EBT3 films in SSRT and microbeam radiotherapy (MRT). Reference Gagliardi, Liam, Poole, Franich and Moshi63

Conclusion

PRESAGE® dosimetric results were found comparable to the Gafchromic EBT film measured and TPS calculated results as evaluated by DVHs, Isodose line profiles and Gamma map analysis. These results revealed that PRESAGE® can reliably be used in verifying volumetric dose distribution data in highly advanced and complicated radiotherapy techniques. From the comparison of anthropomorphic breast-shaped PRESAGE® measured dose distribution with EBT2 film measured and TPS calculated doses, it was confirmed that PRESAGE® can be fabricated into anthropomorphic shapes for validating breast IMRT, brachytherapy and 3D partial breast radiotherapy. PRESAGE® dosimeter should be fabricated into other anthropomorphic shapes for assuring the quality of radiotherapy plans of other tumour sites.

Acknowledgement

None.

Financial support

This review article has no financial support.

Conflicts of interest

None.

References

Khezerloo, D, Nedaie, HA, Takavar, A et al. PRESAGE® as a solid 3D dosimeter: a review article. Radiat Phys Chem 2017; 141: 8897.CrossRefGoogle Scholar
Oldham, M. Methods and techniques for comprehensive 3D dosimetry, In: Godfrey, D, Das, S, Wolbars, A (eds). Advances in Medical Physics. Madison, WI: Medical Physics Pub Corp, 2014: 7081.Google Scholar
Juang, T, Newton, J, Niebanck, M, Benning, R, Adamovics, J, Oldham, M. Customising PRESAGE® for diverse applications. J Phys Conf Ser 2013; 444: 012029.CrossRefGoogle ScholarPubMed
Bedford, JL, Lee, YK, Wai, P, South, CP, Warrington, AP. Evaluation of the Delta4 phantom for IMRT and VMAT verification. Phys Med Biol 2009; 54 (9): N167N176.CrossRefGoogle ScholarPubMed
Clift, C, Thomas, A, Adamovics, JA, Chang, Z, Das, IJ, Oldham, M. Toward acquiring comprehensive radiosurgery field commissioning data using the PRESAGE®/optical-CT 3D dosimetry system. Phys Med Biol 2010; 55: 12791293.CrossRefGoogle Scholar
Newton, J, Thomas, A, Ibbott, G, Oldham, M. Preliminary commissioning investigations with the DMOS-RPC optical-CT scanner. J Phys Conf Ser 2010; 250 (1): 012078.CrossRefGoogle ScholarPubMed
Adamovics, J, Maryanski, MJ. Characterization of PRESAGE: a new 3-D radiochromic solid polymer dosimeter for ionizing radiation. Radiat Prot Dosimetry 2006; 120 (1–4): 107112.CrossRefGoogle Scholar
Alqathami, M, Blencowe, A, Ibbott, G, An investigation into the potential influence of oxygen on the efficiency of the PRESAGE® dosimeter. IC3DDose 2014 – 8th International Conference 3D Radiation Dosimetry 2015: 3–6.Google Scholar
Bache, S, Juang, T, Adamovics, J et al. An investigation of the feasibility of rodentmorphic 3D dosimeters for verification of precision micro-irradiator treatment. Med Phys 2013; 40: 490.CrossRefGoogle Scholar
Farhood, B, Khezerloo, D, Zadeh, TM, Nedaie, HA, Hamrahi, D, Khezerloo, N. Evaluation of the effect of temperature variation on response of PRESAGE® dosimeter. J Caner Res 2017; 13: 118121.Google ScholarPubMed
Guo, P, Adamovics, J, Oldham, M. A practical three-dimensional dosimetry system for radiation therapy. Med Phys 2006; 33: 39623972.CrossRefGoogle ScholarPubMed
Guo, P, Adamovics, J, Oldham, M. Characterization of a new radiochromic three-dimensional dosimeter. Med Phys 2006; 33: 13381345.CrossRefGoogle ScholarPubMed
Sakhalkar, HS, Adamovics, J, Ibbott, G, Oldham, M. A comprehensive evaluation of the PRESAGE®/Optical-CT 3D dosimetry system. Med Phys 2009; 36: 7182.CrossRefGoogle Scholar
Wang, Z, Thomas, A, Newton, J, Ibbott, G, Deasy, J and Oldham, M. Dose verification of stereotactic radiosurgery treatment for trigeminal neuralgia with Presage® 3D dosimetry system. J Phys Conf Ser 2010; 250(1): 012058.CrossRefGoogle Scholar
Yates, ES, Balling, P, Petersen, JB et al. Characterization of the optical properties and stability of Presage following irradiation with photons and carbon ions. Acta Oncol 2011; 50: 829834. pmid:21767181.CrossRefGoogle Scholar
Juang, T, Newton, J, Niebanck, M, Benning, R, Adamovics, J, Oldham, M. Customising PRESAGE® for diverse applications. J Phys Conf Ser 2013; 444: 012029. pmid:24567739.CrossRefGoogle ScholarPubMed
Jackson, J, Juang, T, Adamovics, J, Oldham, M. SU-E-T-53: an investigation of the dosimetric characteristics of a novel radiochromic 3D dosimeter. Med Phys 2013; 40: 215.CrossRefGoogle Scholar
Casanova Borca, V, Pasquino, M, Russo, G et al. Dosimetric characterization and use of GAFCHROMIC EBT3 film for IMRT dose verification. J Appl Clin Med Phys 2013; 14 (2): 158171.CrossRefGoogle ScholarPubMed
Devic, S. Radiochromic film dosimetry: past, present, and future. Phys Med 2011; 27: 122134.CrossRefGoogle ScholarPubMed
Brown, S, Bailey, DL, Willowson, K, Baldock, C. Investigation of the relationship between linear attenuation coefficients and CT Hounsfield units using radionuclides for SPECT. Appl Radiat Isotopes 2008; 66: 12061212.CrossRefGoogle ScholarPubMed
Newton, J, Thomas, A, Ibbott, G, Oldham, M. Preliminary commissioning investigations with the DMOS-RPC optical-CT Scanner. J Phys Conf Ser 2010; 250 (1): 012078p.CrossRefGoogle ScholarPubMed
Devic, S, Seuntjens, J, Sham, E et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys 2005; 32 (7): 22452253.CrossRefGoogle ScholarPubMed
Iqbal, K, Gifford, KA, Ibbott, G, Grant, LR, Buzdar, SA. Comparison of an anthropomorphic PRESAGE® dosimeter and radiochromic film with a commercial radiation treatment planning system for breast IMRT: a feasibility study. J Appl Clin Med Phys 2014; 15: 363374.CrossRefGoogle ScholarPubMed
Newton, J, Thomas, A, Ibbott, G, Oldham, M. Preliminary commissioning investigations with the DMOS-RPC optical-CT Scanner. J Phys Conf Ser 2010; 250 (1): 012078p.CrossRefGoogle ScholarPubMed
Desroches, J, Bouchard, H, Lacroix, F. Potential errors in optical density measurements due to scanning side in EBT and EBT2 Gafchromic film dosimetry. Med Phys 2010; 37 (4): 15651570.CrossRefGoogle ScholarPubMed
Andres, C, Del Castillo, A, Tortosa, R, Alonso, D, Barquero, R. A comprehensive study of the Gafchromic EBT2 radiochromic film. A comparison with EBT. Med Phys 2010; 37 (12): 62716278.CrossRefGoogle ScholarPubMed
Arjomandy, B, Tailor, R, Anand, A et al. Energy dependence and dose response of Gafchromic EBT2 film over a wide range of photon, electron and proton beam energies. Med Phys 2010; 37 (5): 19421947.CrossRefGoogle Scholar
Butson, MJ, Cheung, T, Yu, PKN, Alnawaf, H. Dose and absorption spectra response of EBT2 Gafchromic film to high energy X-rays. Australas Phys Eng Sci Med 2009; 32 (4): 196202.CrossRefGoogle ScholarPubMed
Devic, S, Aldelaijan, S, Mohammed, H et al. Absorption spectra time evolution of EBT-2 model GAFCHROMIC film. Med Phys 2010; 37 (5): 22072214.CrossRefGoogle ScholarPubMed
Devic, S, Seuntjens, J, Sham, E et al. Precise radiochromic film dosimetry using a flat-bed document scanner. Med Phys 2005; 32 (7): 22452253.CrossRefGoogle ScholarPubMed
Zhao, L, Newton, J, Oldham, M, Das, IJ, Cheng, CW, Adamovics, J. Feasibility of PRESAGE® for relative 3D dosimetry of small proton fields. Phys Med Biol 2012; 57 (22): 431443.CrossRefGoogle ScholarPubMed
Al-Nowais, S, Doran, S, Kacperek, A, Krstajic, N, Adamovics, J, Bradley, D. A preliminary analysis of LET effects in the dosimetry of proton beams using PRESAGE and optical CT. Appl Radiat Isotopes 2008; 67 (3): 415418.CrossRefGoogle ScholarPubMed
Abdul Rahman, AT, Bräuer-Krisch, E, Brochard, T et al. Sophisticated test objects for the quality assurance of optical computed tomography scanners. Phys Med Biol 2011; 56 (14): 41774199. doi: 10.1088/0031-9155/56/14/001 CrossRefGoogle ScholarPubMed
Aland, T, Karin, T, Kenny, J. Evaluation of a GAFCHROMIC EBT2 film dosimetry for radiotherapy quality assurance. Australas Phys Eng Sci Med 2011; 34: 251260.CrossRefGoogle ScholarPubMed
MacDougall, ND, Miquel, ME, Keevil, SF. Effects of phantom volume and shape on the accuracy on the accurancy of MRI BANG gel dosimetry using BANG3. Br J Radiol 2008; 81: 4650.CrossRefGoogle Scholar
Iqbal, K., Ibbott, G., Lafratta, R., Gifford, K., Akram, M., Buzdar, S. Dosimetric feasibility of an anthropomorphic three-dimensional PRESAGE® dosimeter for verification of single entry hybrid catheter accelerated partial breast brachytherapy. J Radiother Pract 2018; 17 (4): 403410.CrossRefGoogle Scholar
Iqbal, K, Ibbott, G, Grant, RL, Gifford, KA, Buzdar, SA. Dosimetric characterization of anthropomorphic PRESAGE® dosimeter and EBT2 film for partial breast radiotherapy. J Radiother Pract 2017; 17: 96103.CrossRefGoogle Scholar
Choi, GW. Measurement of the electron return effect using PRESAGE dosimeter. 2016. UT GSBS Dissertations and Theses. 690. The University of Texas, Graduate School of Biomedical Sciences, Houston.Google Scholar
Low, DA, Harms, WB, Mutic, S, Purdy, JA. A technique for the quantitative evaluation of dose distributions. Med Phys 1998; 25: 656661. PubMed PMID: 9608475.CrossRefGoogle ScholarPubMed
Oldham, M., Sakhalkar, H, Guo, P, Adamovics, J. An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques. Med Phys 2008; 35: 20722080.CrossRefGoogle ScholarPubMed
Guo, P, Adamovics, J, Oldham, M. A practical three-dimensional dosimetry system for radiation therapy. Med Phys 2006; 33: 39623972.CrossRefGoogle ScholarPubMed
Oldham, M., Sakhalkar, H, Guo, P, Adamovics, J. An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques. Med Phys 2008; 35: 20722080.CrossRefGoogle ScholarPubMed
Sakhalkar, HS, Adamovics, J, Ibbott, G, Oldham, M. A comprehensive evaluation of the PRESAGE®/optical-CT 3D dosimetry system. Med Phys 2009; 36: 7182.CrossRefGoogle Scholar
Thomas, A, Newton, J, Adamovics, J, Oldham, M. Commissioning and benchmarking a 3D dosimetry system for clinical use. Med Phys 2011; 38: 48464857.CrossRefGoogle ScholarPubMed
Moutsatsos, A, Pantelis, E., Antypas, C. et al. On the use of presage detectors for the dosimetry of helical tomotherapy small fields. Phys Medica 2016; 32 (Suppl 3): 332.CrossRefGoogle Scholar
Pierquet, M, Craciunescu, O, Steffey, B, Song, H, Oldham, M. On the feasibility of verification of 3D dosimetry near Brachytherapy sources using PRESAGE®/Optical-CTIC3DDose: The 6th international conference on 3D radiation dosimetry IOP publishing. J Phys Conf Ser 2010; 250: 012091.CrossRefGoogle Scholar
Doran, SJ, Yatigammana, DN. Eliminating the need for refractive index matching in optical CT scanners for radiotherapy dosimetry: I. Concept and simulations. Phys Med Biol 2012; 57: 665683.CrossRefGoogle ScholarPubMed
Oldham, M, Sakhalkar, H, Guo, P, Adamovics, J. An investigation of the accuracy of an IMRT dose distribution using two- and three-dimensional dosimetry techniques. Med Phys 2008; 35: 20722080.CrossRefGoogle ScholarPubMed
Thomas, A, Newton, J, Adamovics, J, Oldham, M. Commissioning and benchmarking a 3D dosimetry system for clinical use. Med Phys 2011; 38: 48464857.CrossRefGoogle ScholarPubMed
Raaijmakers, AJ, Raaymakers, BW, Lagendijk, JJ. Integrating a MRI scanner with a 6 MV radiotherapy accelerator: dose increase at tissue-air interfaces in a lateral magnetic field due to returning electrons. Phys Med Biol 2005; 50: 13631376.CrossRefGoogle Scholar
Raaijmakers, AJ. MR-Guided Radiotherapy: Magnetic Field Dose Effects. Netherlands: Utrecht University, 2008.Google Scholar
Meijsing, I, Raaymakers, BW, Raaijmakers, AJ et al. Dosimetry for the MRI accelerator: the impact of a magnetic field on the response of a Farmer NE2571 ionization chamber. Phys Med Biol 2009; 54: 29933002.CrossRefGoogle ScholarPubMed
Mathis, M, Sawakuchi, G, Flint, D et al. Effects of a strong magnetic field on selected radiation dosimeters (TLD, OSLD, EBT3 film, PRESAGE). Congress 2014 CSM, Poster No. 0715. doi: 10.1594/ranzcr2014/R-0175CrossRefGoogle Scholar
ICRU. Determination of Absorbed Dose in a Patient Irradiated by Beams 230of X- or Gamma-Rays in Radiotherapy Procedures. Bethesda, MD:International-Commission on Radiation Units and Measurement, 1976; 232ICRU Rep. 24.Google Scholar
Johansson, KA, Hariot, JC, Van Dam, J, Lepinoy, D, Setenac, I, Sernbo, G. Quality assurance control in the EORTC cooperative group of 235radiotherapy. Dosimetric intercomposition. Radiather Oncol 1986; 7: 269279.CrossRefGoogle Scholar
Shamsi, Q, Buzdar, SA, Atiq, A, Atiq, M, Altaf, S, Iqbal, K. Dosimetric determination of tissue maximum ratios in small fields. J Radiother Pract. April 2018. doi: 10.1017/S1460396918000055 CrossRefGoogle Scholar
Zhao, L, Newton, J, Oldham, M, Das, I, Cheng, C, Adamovics, J. Feasibility of PRESAGE® for relative 3D dosimetry of small proton fields. Phys Med Biol 2012; 57 (22): N431N443.CrossRefGoogle ScholarPubMed
Carrol, M, Ibbott, G, Gillin, M, Adamovics, J. Investigation of 3D dosimetry for proton therapy using PRESAGE. Med Phys 2013; 40 (6): 220.CrossRefGoogle Scholar
Alqathami, M, Blencowe, A, Geso, M, Ibbott, G. Characterization of Novel PRESAGE® dosimeters for megavoltage and kilovoltage X-ray beam dosimetry. Radiat Meas 2015; 74: 1219.CrossRefGoogle Scholar
Gorjiara, T, Hill, R, Kuncic, Z et al. Investigation of radiological properties and water equivalency of PRESAGE®dosimeters. Med Phys 2011; 38: 22652274.CrossRefGoogle Scholar
Brady, SL, Brown, WD, Clift, CG, Yoo, S, Oldham, M. Investigation into the feasibility of using PRESAGE®/optical-CT dosimetry for the verification of gating treatments. Phys Med Biol 2010; 55: 21872201.CrossRefGoogle ScholarPubMed
Jackson, J, Juang, T, Adamovics, J, Oldham, M. An investigation of PRESAGE®3D dosimetry for IMRT and VMAT radiation therapy treatment verification. Phys Med Biol 2015; 60: 22172230.CrossRefGoogle ScholarPubMed
Gagliardi, FM, Liam, D, Poole, CM, Franich, RD, Moshi, G. Water equivalent PRESAGE® for synchrotron radiation therapy dosimetry. Med Phys 2018; 45 (3): 12551265.CrossRefGoogle ScholarPubMed
Alqathami, M, Blencowe, A, Geso, M, Ibbott, G. Characterization of Novel PRESAGE® dosimeters for megavoltage and kilovoltage X-ray beam dosimetry. Radiat Meas 2015; 74: 1219.CrossRefGoogle Scholar
Figure 0

Figure 1. OD to dose curve for PRESAGE (a); OD to dose curve for EBT2 film (b).

Figure 1

Figure 2. (a) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS (Nucletron, an Elekta company, Elekta AB, Stockholm, Sweden) and EBT2 film for film 1. (b) Comparison of the calculated and measured point skin doses for PRESAGE, Oncentra Brachy Planning TPS and EBT2 film for surface film 2.

Figure 2

Figure 3. Gamma maps (a) (±3%/3 mm) between Pinnacle, EBT2 and PRESAGE for a region of PTV 5 mm on film plane; (b) (±3%/3 mm) in the transverse and sagittal planes for PTV 5 mm intersecting film plane.

Figure 3

Figure 4. Line profiles of Pinnacle, PRESAGE and EBT2 film dose distributions from axial slices.

Figure 4

Figure 5. Line profiles of the Pinnacle3, PRESAGE® and EBT2 film dose distributions of axial slice.

Figure 5

Figure 6. GTV, CTV and PTV DVH comparison between the PRESAGE® and Pinnacle3 Planning dose distributions.