Dosimetric feasibility of an anthropomorphic three-dimensional PRESAGE® dosimeter for verification of single entry hybrid catheter accelerated partial breast brachytherapy

Published online by Cambridge University Press:  20 July 2018

Khalid Iqbal*
Affiliation:
Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA Department of Physics, The Islamia University, Bahawalpur, Pakistan Department of Radiation Oncology, Shaukat Khanum Cancer Hospital & Research Center, Lahore, Pakistan
Geoffrey S. Ibbott
Affiliation:
Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA
Ryan G. Lafratta
Affiliation:
Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA
Kent A. Gifford
Affiliation:
Department of Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA
Muhammad Akram
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan
Saeed A. Buzdar
Affiliation:
Department of Physics, The Islamia University, Bahawalpur, Pakistan
*
Author for correspondence: Khalid Iqbal, Department Radiation Physics, The University of Texas MD Anderson Cancer Center, Houston, TX, USA. Tel: +924235905000. Fax: +924235945206. E-mail: khalid_phy@yahoo.com
Rights & Permissions [Opens in a new window]

Abstract

Purpose

To determine the feasibility of an anthropomorphic breast polyurethane-based three-dimensional (3D) dosimeter with cavity to measure dose distributions and skin dose for a commercial strut-based applicator strut-adjusted volume implant (SAVI™) 6–1.

Materials and methods

An anthropomorphic breast 3D dosimeter was created with a cavity to accommodate the SAVI™ strut-based device. 2 Gy was prescribed to the breast dosimeter having D95 to planning target volume evaluation (PTV_EVAL) while limiting 125% of the prescribed dose to the skin. Independent dose distribution verification was performed with GAFCHROMIC® EBT2 film. The dose distribution from the 3D dosimeter was compared to the distributions from commercial brachytherapy treatment planning system (TPS) and film. Point skin doses, line profiles and dose–volume histogram (DVHs) for the skin and PTV_EVAL were compared.

Results

The maximum difference in skin dose for TPS and the 3D dosimeter was 4% whereas 41% between the TPS and EBT2 film. The maximum dose difference for line profiles between TPS, 3D dosimeter, and film was 4·1%. DVHs of skin and PTV_EVAL for TPS and 3D dosimeter differed by a maximum of 4% at 5 mm depth and skin differed by a maximum 1·5% between TPS and 3D dosimeter. The criterion for gamma analysis comparison was 92·5% at ±5%±3 mm criterion. The TPS demonstrated at least ±5% comparability in predicting dose to the skin, PTV_EVAL and normal breast tissue.

Conclusions

3D anthropomorphic polyurethane dosimeter with cavity gives comparable results to the TPS dose predictions and GAFCHROMIC® EBT2 film results in the context of HDR brachytherapy.

Type
Original Article
Copyright
© Cambridge University Press 2018 

Introduction

Accelerated partial breast irradiation (APBI) provided by high dose-rate (HDR) brachytherapy offers an excellent compact course of radiation due to a small number of fractions for early stage breast carcinoma.Reference Manoharan, Rodriguez, Bobba and Mukka 1 APBI has been used by multiple methods which includes three-dimension conformal external radiotherapy, MammoSite® balloon internal radiation therapy and multiple catheter brachytherapy.Reference Guo, Adamovics and Oldham 2 Reference Palmer, Di Pietro and Alobaidli 5 The MammoSite® brachytherapy system (Hologic, Marlborough, MA, USA) with a single dwell position has been widely accepted due to technically much easier to perform and axially symmetric dose distribution.Reference Sakhalkar, Adamovics, Ibbott and Oldham 6 But one matter of concern in MammoSite® is inadequate balloon to skin distance even using multiple dwell positions which can change the shape of the dose distribution up to some extent. The recent development in APBI is strut-adjusted volume implant (SAVI™) device (Cianna Medical Inc., Aliso Viejo, CA, USA) which has 6, 8 or 10 peripheral source channels with one centre channel.Reference Petrokokkinos, Zourari and Pantelis 7 MammoSite® and multiple catheters of SAVI™ with struts adjusted volume combine the advantages of APBI treatment.Reference Major, Polgár, Lövey and Fröhlich 8 , Reference Pantelis, Papagiannis and Karaiskos 9

APBI treatment with HDR brachytherapy delivered dose is accurate and reliable using Ir-92. Ir-192 is the most common source for remote after loader in HDR with advantages of high specific activity (450 Ci/g) that allow the construction of high activity source (10 Ci) of small diameter (0·6–1·1 mm).

Skin dose can be the limiting factor in radiation therapy treatments, and it is a fairly common cause of toxicity in radiation therapy treatment. The chosen planning optimisation strategy can also affect the skin dose.Reference Poon and Verhaegen 10 Reference Lee, Chuang and Quivey 18 Despite the clinical importance of skin dose, the literature contains scanty detail concerning the expected accuracy of radiation treatment planning and skin dose calculations. Using radiochromic film, Chung et al.Reference Chung, Jin and Dempsey 11 reported that two treatment planning systems (TPS’s) (Pinnacle3, Royal Philips Electronics, Eindhoven, The Netherlands and CORVUS, North American Scientific, Chatsworth, CA, USA) overestimated surface dose by 7·4–18·5%. Zourari et al.Reference Zourari, Major, Herein, Peppa, Polgár and Papagiannis 12 in 2015, made a dosimetric comparison using a contemporary (MBDCA) model-based dose calculation algorithm following TG 186 protocol and TG 43 techniques in HDR Ir-192 breast brachytherapy and found a maximum percentage difference of the order of 6% for D 10 cc for the skin. Shahid et al. presented that dosimetric differences between a commercially available treatments planning system utilising the TG43U1 dose calculation algorithm did not differ from measurements by more than 5–6% for points on the skin.Reference Awan, Dini, Hussain, Soleimani‐Meigooni and Meigooni 13

In this work, we want to compare two parameters:

First, comparison of skin dose calculations for TPS and dosimeter. Physicians often ask planners this important clinical question. For optimisation to be reliable, calculated and measured skin dose must be correlated, even if the calculation contains a systematic error.Reference Dickler, Kirk and Choo 16 , Reference Hsu, Lessard, Weinberg and Pouliot 17

Second, can the PRESAGE® (Heuris Pharma LLC, Skillman, NJ, USA) dosimeter be feasible for HDR brachytherapy dosimetry for breast having a cavity? Furthermore, we wished to determine the accuracy of the commercial brachytherapy TPS in predicting dose to the skin, planning target volume for plan evaluation (PTV_EVAL) and normal tissues. The literature has yet to address these important clinical questions for the Oncentra® (Nucletron, an Elekta Company, Elekta AB, Stockholm, Sweden) Brachy TPS. To answer our questions, we used PRESAGE® and GAFCHROMIC® EBT2 (International Specialty Products Inc., Wayne, NJ, USA) film to measure dose distributions delivered with the SAVI™ device. As EBT film is the first radiochromic film suitable for the use with doses as the typical doses occurring in radiotherapy.Reference Aland, Karin and Kenny 14 The resulting knowledge will be useful when evaluating and comparing measured and calculated plans and also when developing optimisation strategies for the SAVI™ device.

Methods

SAVI™ devices

The SAVI™ is an interstitial type and APBI device with single entry brachytherapy. It includes the benefits of both MammoSite and multi-catheter and is available in 6, 8, 10 peripheral struts with a central strut. SAVI™ 6-1 was used in this work and inserted inside the lumpectomy cavity through a small incision and is inserted in a collapsed state and by the clockwise rotation of the knurled knob at the proximal end of the device; the peripheral struts are expanded as shown in Figure 1. The SAVI™ has 3·0 cm diameter and 6·1 cm long axis. The optimisation of dwell positions and weights is possible by multiple catheters to account for close vicinity of ribs, skin and pectoralis muscle. Dose conformity is obtained with additional peripheral catheters of the device which permit the user to more closely model in an interstitial implant.Reference Awan, Dini, Hussain, Soleimani‐Meigooni and Meigooni 13 , Reference Aland, Karin and Kenny 14

Figure 1 SAVI™ 6-1 size with peripheral struts expanded (Courtesy: Cianna Medical).

PRESAGE® dosimeter for optical-CT scanning

A PRESAGE® dosimeter (Heuris Pharma LLC, Skillman, NJ, USA) was moulded from a pre-mould mixture comprised of a solvent, leuco dye and free radical initiator. The formulation of PRESAGE® used in this study has Zeff of 7·6 and a physical density of 1·07 g/cm3. The physical dimensions of breast PRESAGE® dosimeter were 13·8 cm (length) by 12·4 cm (width) and 4·6 cm (height)Reference Patel, Becker, Das and Mackie 15 Reference Vidovic, Juang and Meltsner 25 as shown in Figure 2a. The dosimeter was scanned with the Duke midsized optical scanner dedicated for the RPC (DMOS-RPC) (Duke University, Durham, NC, USA) using 1° per step to produce 360 projection images.Reference Chung, Jin and Dempsey 23 , Reference Susan and Ramiro 26 The parameters used during the optical scanning of PRESAGE dosimeter were refractive indices of the matching fluid and the dosimeter, attenuation coefficients of the fluid and the medium, wavelength of the light used, radius and position of the dosimeter. The refractive index of the matching fluid was set equal to the refractive index of the PRESAGE dosimeter which was 1·503. The optical scanning and irradiation of the dosimeter and cuvettes were performed at room temperature (22°C). Transverse images were reconstructed by filtered back projection to a 1 mm voxel edge.Reference Newton, Thomas, Ibbott and Oldham 27 Radio-opaque markers are present on struts two, four and six of the device to allow struts identification in the treatment planning as shown in Figure 2b.

Figure 2 (a) PRESAGE®, dosimeter with EBT2 films inserted. (b) Brachytherapy treatment plan with dose distribution in three dimensions.

Treatment planning and delivery

A treatment planning X-ray CT scan with a slice thickness of 1·25 mm of the breast PRESAGE® dosimeter was acquired using a GE CT scanner (GE Healthcare Technologies, Waukesha, WI, USA). CT data were exported to the Oncentra® Masterplan version, 4.1 brachy planning workstation. The cavity was contoured as the periphery of the struts of the SAVI™ device. The skin was generated by contracting the external contour by 2 mm.Reference Hadley, Kelly and Lam 19 The PTV for plan optimisation (PTV_OPT) was generated by expanding the cavity 10 mm isotropically and limiting it by the skin. The PTV_EVAL was created by subtracting the cavity from the PTV_OPT. Figure 2 shows the external, PTV_EVAL and skin contours on an axial CT slice. Subvolumes were created by contracting the (external-PTV_OPT) by 1 mm, 3 mm and 5 mm isotropically. The inverse planning simulated annealing (IPSA) algorithm was used to optimally prescribe 2 Gy to the PTV_EVAL while limiting the maximum skin dose (D 0·1 cc) to or below 125% of the prescribed dose. An IPSA is simulated annealing optimisation algorithm and is based on the patient’s anatomy contoured from the CT scan and with the help of volumetric or surface dose constraints. IPSA is capable of rapid generation of conformal plans by giving the dwell times distribution within the catheters.Reference Iqbal, Isa, Buzdar, Gifford and Afzal 20 2 Gy was prescribed to the breast dosimeter having D 95 to PTV_EVAL while limiting 125% of the prescribed dose to the skin using the Nucletron HDR microSelectron after loader. PTV_EVAL is considered as the difference between the expanded and the cavity volume.Reference Court and Tisher 21 PTV_EVAL was prescribed to receive 95% of the prescription dose which was equal to 2 Gy in each fraction. Cuvettes of PRESAGE® with a volume of 1×1×3 cm were irradiated in high impact polystyrene with 6 MV beam on a Varian 21EX linear accelerator. Dose levels were 0, 3, 6 and 9 Gy. The absorption of the material was determined by a Genesys 20 spectrophotometer (Thermo Fisher Scientific, Houston, TX, USA) before and post-irradiation and the optical density (OD) was compared with the dose delivered to calculate the PRESAGE® calibration.

EBT2 film dosimetry

Independent dose distribution verification was performed with GAFCHROMIC® EBT2 film. Temporal stability, directional independence and convenience of the self-developing radiochromic film were the basic reasons to use EBT2.Reference Chung, Jin and Dempsey 23 A PRESAGE® dosimeter was cut with a scalpel blade at its central, long axis and pieces of EBT2 film were placed between the PRESAGE® halves for line profile measurement and also on top of dosimeter for skin dose calculations. The PRESAGE dosimeter was irradiated with the films inserted. An OD to dose curve was measured in solid water with 6 MV photons from a Varian 21EX. The films were scanned in three colours (48-bit RGB) at a scanning resolution of 75 dpi in transmission mode on flatbed photo-scanner Epson-10000XL (Epson America, Inc., Long Beach, CA, USA). Each film was scanned in transmission mode but only the red channel was extracted for analysis because it has more sensitive response to dose as compared with blue and green channels.Reference Brady, Brown, Clift, Yoo and Oldham 28 Reference Iqbal, Gifford, Ibbott, Grant and Buzdar 32

Data registration and dose analysis

The transverse images were reconstructed in terms of change in OD by the DMOS Matlab program (Duke University). Theses transverse images with dose distribution from DMOS and Oncentra® treatment plan were exported to the computational environment for radiotherapy research program (CERR) (Memorial Sloan Kettering Cancer Center, New York, NY, USA), Matlab-based software which is used to analyse and display the radiotherapy plans. CERR scaled the change in OD values to dose values with the help of the scale factor taken from the calibration curve. The dose distributions were generated by the datasets that were loaded and registered into CERR and then normalised to relative dose distributions. The calculated dose distribution from Oncentra® was compared with the measured distributions from PRESAGE® and EBT2. EBT2 scans were analysed using Image-J software (National Institutes of Health, USA).

Results and Discussion

Film and PRESAGE® calibration

Figure 3a shows the calibration curve (OD to dose) for the PRESAGE®. Figure 3b shows the calibration curve (OD to dose) for the EBT2 film. The uncertainty in net OD was estimated at 0·8% (1 SD) for PRESAGE® and 0·7% (1 SD) for EBT2 film. The net OD uncertainty was estimated by subtracting the pre-irradiation scans from the post-irradiation scans. Sakhalkar et al. in 2009 and Guo et al. in 2006 have reported various uncertainties associated with the calibration process of EBT2 film and Presage intradosimeter consistency, temporal stability and temperature gradients in the literature.Reference Sakhalkar, Sterling, Adamovics, Ibbott and Oldham 24 , Reference Guo, Adamovics and Oldham 33 , Reference Iqbal, Iqbal, Akram, Altaf and Buzdar 34 In the case of PRESAGE® calibration, small volumes were irradiated for particular dose using from same batch of the PRESAGE®. The determination of using small volumes to calibrate large volumes was to introduce errors when comparing to dose distribution.Reference Iqbal, Gillin, Summers, Dhanesar, Gifford and Buzdar 29 , Reference MacDougall, Miquel and Keevil 30 , Reference Iqbal, Ibbott, Lafratta, Gifford and Buzdar 35 The volume effect uncertainty of the PRESAGE® required the use of normalisation as compare dose distribution to PRESAGE® dosimeter with EBT2 film and Oncentra® treatment planning. However, the dose-response linearity of the PRESAGE® has been extensively confirmed.Reference Patel, Becker, Das and Mackie 15 In this paper, the PTV_EVAL dose of the PRESAGE® was within the calibration uncertainty of planned dose as measured by the films. We therefore conclude that the PRESAGE® is relative dosimeter, by normalising D 95 to PTV_EVAL and PRESAGE® linearity response did not introduce any limitation on data analysis.Reference Gifford, Iqbal, Grant, Buzdar and Ibbott 36

Figure 3 (a) Breast contours figures [external breast body, cavity, planning target volume evaluation (PTV_EVAL), skin]. (b) External body-PTV for plan optimisation: −1 mm, −3 mm, −5 mm.

Figure 4 illustrates the comparison between Oncentra®, PRESAGE® and EBT2 film for the two film planes. The maximum dose difference of both films was 4·5% between Oncentra® and PRESAGE® and 2·42% between Oncentra® and EBT2 film. The mean percentage differences of skin dose for film 1 and film 2 were 3·54 and 2·85%, respectively, between Oncentra® and PRESAGE®. Oncentra® and EBT2 mean differences of film 1 and film 2 were 0·97 and 0·53% at different selected number of points, respectively, as shown in Table 1.

Figure 4 (a) Calibration curve [optical density (OD) to dose] for the PRESAGE®. (b) Calibration curve (OD to dose) for the EBT2 film.

Table 1 Point skin dose differences between Oncentra®, PRESAGE® and EBT2

Figure 5 illustrates the DVH comparison between and Oncentra® and PRESAGE® for the (a) skin and (b) A plot of dose versus relevant PTV_EVAL coverage parameters. The maximum per cent dose difference was 4% between Oncentra® and PRESAGE®. Preliminary investigations suggest the cause may be a reflection artefact of laser light from the underside of the top and bottom of the dosimeter.Reference Iqbal, Gifford, Ibbott, Grant and Buzdar 32

Figure 5 (a) Comparison of the calculated and measured point skin doses for PRESAGE®, Oncentra® Brachy Planning TPS and EBT2 film for film 1. (b) Comparison of the calculated and measured point skin doses for PRESAGE®, Oncentra® Brachy Planning TPS and EBT2 film for film 2.

The PTV_EVAL V 90 and V 95 of Oncentra® and PRESAGE® were 99·6, 98·95% and 97, 96·59%, respectively (V 95 is the volume receiving 95% of the prescribed dose). The PTV_EVAL V 150 and V 200 were well below the optimisation goals of V 150<50 cc and V 200<20 cc of the absolute volume (Protocol B-39/RTOG Protocol 0413) 37 as shown in Table 2.

Table 2 Planning target volume evaluation (PTV_EVAL) parameters for Oncentra® and PRESAGE®

Isodose line profiles

Figure 6 illustrates the dose line profile comparisons between Oncentra®, PRESAGE® and EBT2. The maximum line profile dose difference of both films was 2·84% between Oncentra® and PRESAGE® dosimeter and 4·06% between Oncentra® and EBT2 film. The mean percentage differences of line dose profiles of film 1 and film 2 were 2·25 and 1·80% between Oncentra® and PRESAGE®. Oncentra® and EBT2 mean percentage differences of film 1 and film 2 were 3·31 and 2·82% as shown in Table 3.

Figure 6 (a) Dose–volume histogram comparison between Oncentra® and PRESAGE® for skin. (b) Plot of dose versus relevant planning target volume evaluation coverage parameters for Oncentra® and PRESAGE®.

Table 3 Line profile comparisons between Oncentra®, PRESAGE® and EBT2 film

Normal breast tissue DVHs

Figure 7 illustrates the DVH comparisons between Oncentra® and PRESAGE® for normal breast tissue and normal breast tissue subvolumes. There are regions that show differences with Oncentra® near the edge of the dosimeter. This is likely due to edge artefacts in the PRESAGE® distribution. The edge artefacts were reduced by effective refractive index matching between the dosimeter and the matching fluid. The external-PTV_OPT DVH curve indicates that the PRESAGE® dose estimation was slightly less homogenous than that calculated by Oncentra®, with small regions of relative over and under dose occurring near the edges of the dosimeter. For the external-PTV_OPT, a 4% maximum dose difference was observed at upper and lower volume ends as shown in the Figure 7a. DVHs of the normal breast tissue subvolumes show that there are some edge artefacts resulting in maximum differences of 3·2 and 2·5% for the 1 and 3 mm subvolumes, respectively, as many studies reported earlier.Reference Iqbal, Gillin, Summers, Dhanesar, Gifford and Buzdar 29 , Reference MacDougall, Miquel and Keevil 30 The normal breast tissue 5 mm subvolume DVH showed excellent agreement between Oncentra® and PRESAGE® with a maximum difference of 1·5% (Figure 8).

Figure 7 (a) Line profile comparison between Oncentra®, PRESAGE® and EBT2 film dose distributions of axial slice. (b) Line profile comparison between Oncentra®, PRESAGE® and EBT2 film dose distributions of axial slice.

Figure 8 Dose–volume histogram comparisons between Oncentra® and PRESAGE® dose distributions. (a) External-planning target volume for plan optimisation (PTV_OPT). (b) External-PTV_OPT, −1 mm. (c) External-PTV_OPT, −3 mm. (d) External-PTV_OPT, −5 mm.

Gamma map comparison

Figure 9 is gamma comparisons of PRESAGE®/optical-CT, EBT2 film measurement and dose calculation of Oncentra® TPS at ±5%±3 mm criterion. The amount for the axial 2D gamma comparisons of EBT2 versus PRESAGE®, PRESAGE® versus Oncentra® and EBT2 versus Oncentra® were 92·4, 93·5 and 91·8%, respectively.

Figure 9 Gamma map comparisons of PRESAGE®/optical-CT , EBT2 film and Oncentra® treatment planning.

Conclusions

This work demonstrated that it is possible to fashion a breast PRESAGE® dosimeter with a cavity that will accommodate the SAVI™ device. This fact allows for patient-specific HDR treatment plan quality assurance. Further, these data show that it is also possible to obtain comparable dosimetry from an anthropomorphic PRESAGE® dosimeter. The Oncentra® TPS demonstrated comparable dose calculation up to ±5% for the skin, PTV_EVAL and normal breast tissue and 92·5% gamma map comparison at ±5%±3 mm criterion.

Acknowledgements

This work was supported by the NIH grant no. 5R01 CA 10083. The authors also very thankful to HEC (Higher education commission of Pakistan) which provided the scholarship for the completion of this Ph.D. research. Special thanks to Laura Rechner, M.S., for helping irradiate the breast PRESAGE® dosimeter.

References

1 Manoharan, S R, Rodriguez, R R, Bobba, V S, Mukka, C. Dosimetry evaluation of SAVI-based HDR brachytherapy for partial breast irradiation. J Med Phys 2010; 35: 131136.Google Scholar
2 Guo, P Y, Adamovics, J A, Oldham, M. Characterization of a new radiochromic three-dimensional dosimeter. Med Phys 2006; 33 (5): 13381345.Google Scholar
3 Scanderbeg, D J, Yashar, C, Rice, R, Pawlicki, T. Clinical implementation of a new HDR brachytherapy device for partial breast irradiation. Radiother Oncol 2009; 90: 3642.Google Scholar
4 Low, D A, Dempsey, J F. Evaluation of the gamma dose distribution comparison method. Med Phys 2003; 30 (9): 24552464.Google Scholar
5 Palmer, A L, Di Pietro, P, Alobaidli, S et al. Comparison of methods for the measurement of radiation dose distributions in high dose rate (HDR) brachytherapy: Ge-doped optical fiber, EBT3 Gafchromic film, and PRESAGE® radiochromic plastic. Med Phys 2013; 40 (6): 061707.Google Scholar
6 Sakhalkar, H S, Adamovics, J, Ibbott, G, Oldham, M. A comprehensive evaluation of the PRESAGE/optical-CT 3D dosimetry system. Med Phys 2009; 36 (1): 7182.Google Scholar
7 Petrokokkinos, L, Zourari, K, Pantelis, E et al. Dosimetric accuracy of a deterministic radiation transport based I192r brachytherapy treatment planning system. Part II: Monte Carlo and experimental verification of a multiple source dwell position plan employing a shielded applicator. Med Phys 2011; 38 (4): 19811992.Google Scholar
8 Major, T, Polgár, C, Lövey, K, Fröhlich, G. Dosimetric characteristics of accelerated partial breast irradiation with CT image–based multicatheter interstitial brachytherapy: a single institution’s experience. Brachytherapy 2011; 10 (5): 421426.Google Scholar
9 Pantelis, E, Papagiannis, P, Karaiskos, P et al. The effect of finite patient dimensions and tissue inhomogeneities on dosimetry planning of 192 Ir HDR breast brachytherapy: a Monte Carlo dose verification study. Int J Radiat Oncol Biol Phys 2005; 61 (5): 15961602.Google Scholar
10 Poon, E, Verhaegen, F. Development of a scatter correction technique and its application to HDR 192Ir multicatheter breast brachytherapy. Med Phys 2009; 36 (8): 37033713.Google Scholar
11 Chung, H, Jin, H, Dempsey, J F et al. Evaluation of surface and build‐up region dose for intensity‐modulated radiation therapy in head and neck cancer. Med Phys 2005; 32 (8): 26822689.Google Scholar
12 Zourari, K, Major, T, Herein, A, Peppa, V, Polgár, C, Papagiannis, P. A retrospective dosimetric comparison of TG43 and a commercially available MBDCA for an APBI brachytherapy patient cohort. Phys Med 2015; 31 (7): 669676.Google Scholar
13 Awan, S B, Dini, S A, Hussain, M, Soleimani‐Meigooni, D, Meigooni, A S. Cylindrical coordinate based TG_43U1 parameters for dose calculation around elongated brachytherapy sources. J Appl Clin Med Phys 2008; 9: 123142.Google Scholar
14 Aland, T, Karin, T, Kenny, J. Evaluation of a Gafchromic EBT2 film dosimetry system for radiotherapy quality assurance. Australas Phys Eng Sci Med 2011; 34: 251260.Google Scholar
15 Patel, R R, Becker, S J, Das, R K, Mackie, T R. A dosimetric comparison of accelerated partial breast irradiation techniques: multicatheter Interstitital brachytherapy, three-dimensional conformal radiotherapy and supine versus prone helical tomotherapy. Int J Radiat Oncol Biol Phys 2007; 68: 935942.Google Scholar
16 Dickler, A, Kirk, M, Choo, J et al. Treatment volume and dose optimization of Mammosite breast brachytherapy applicator. Int J Radiat Oncol Biol Phys 2004; 59: 469472.Google Scholar
17 Hsu, I C, Lessard, E, Weinberg, V, Pouliot, J. Comparison of inverse planning simulated annealing and geometrical optimization for prostate high-dose-rate brachytherapy. Brachytherapy 2004; 3: 147152.Google Scholar
18 Lee, N, Chuang, C, Quivey, J M et al. Skin toxicity due to intensity-modulated radiotherapy for head-and-neck carcinoma. Int J Radiat Oncol Biol Phys 2002; 53: 630637.Google Scholar
19 Hadley, S W, Kelly, R, Lam, K. Effects of immobilization mask material on surface dose. J Appl Clin Med Phys 2005; 6: 17.Google Scholar
20 Iqbal, K, Isa, M, Buzdar, S A, Gifford, K A, Afzal, M. Treatment planning evaluation of sliding window and multiple static segments technique in intensity modulated radiotherapy. Rep Pract Oncol Radiother 2013; 18 (2): 101106.Google Scholar
21 Court, L E, Tisher, R B. Experimental evaluation of the impact of different head-and-neck intensity-modulated radiation therapy planning techniques on doses to the skin and shallow targets. Int J Radiat Oncol Biol Phys 2007; 69: 607613.Google Scholar
22 Panettieri, V, Smith, R L, Mason, NJ, Millar, J L. Comparison of IPSA and HIPO inverse planning optimization algorithms for prostate HDR brachytherapy. J Appl Clin Med Phys 2014; 15 (6): 256266.Google Scholar
23 Chung, H, Jin, H, Dempsey, J F et al. Evaluation of surface and build-up region dose for intensity-modulated radiation therapy in head and neck cancer. Med Phys 2005; 32: 26822689.Google Scholar
24 Sakhalkar, H, Sterling, D, Adamovics, J, Ibbott, G, Oldham, M. Investigation of the feasibility of relative 3D dosimetry in the Radiologic Physics Center Head and Neck IMRT phantom using Presage/optical‐CT. Med Phys 2009; 36 (7): 33713377.Google Scholar
25 Vidovic, A K, Juang, T, Meltsner, S et al. An investigation of a PRESAGE® in vivo dosimeter for brachytherapy. Phys Med Biol 2014; 59 (14): 3893.Google Scholar
26 Susan, L R, Ramiro, P. Dosimetric effects of an air cavity for the SAVI™ partial breast irradiation applicator. Med Phys 2010; 37: 39193926.Google Scholar
27 Newton, J, Thomas, A, Ibbott, G, Oldham, M. Preliminary commissioning investigations with the DMOS-RPC optical-CT Scanner. J Phys Conf Ser 2010; 250: 012078.Google Scholar
28 Brady, S L, Brown, W E, Clift, C G, Yoo, S, Oldham, M. Investigation into the feasibility of using PRESAGE® /optical CT dosimetry for the verification of gating treatments. Phys Med Biol 2010; 55: 21872201.Google Scholar
29 Iqbal, K, Gillin, M, Summers, P A, Dhanesar, S, Gifford, K A, Buzdar, S A. Quality assurance evaluation of spot scanning beam proton therapy with an anthropomorphic prostate phantom. Br J Radiol 2013; 86 (1031): 20130390.Google Scholar
30 MacDougall, N D, Miquel, M E, Keevil, S F. Effects of phantom volume and shape on the accuracy of MRI BANG gel dosimetry using BANG3. Br J Radiol 2008; 81: 4650.Google Scholar
31 Sakhalkar, H, Oldham, M. Fast, high-resolution 3D dosimetry utilizing a novel optical-CT scanner incorporating tertiary telecentric collimation. Med Phys 2008; 35: 101111.Google Scholar
32 Iqbal, K, Gifford, K A, Ibbott, G, Grant, L R, Buzdar, S A. Comparison of an anthropomorphic PRESAGE® dosimeter and Radiochromic film with a commercial radiation treatment planning system for breast IMRT: a feasibility study. J Appl Clin Med Phys 2014; 15: 363374.Google Scholar
33 Guo, P, Adamovics, J, Oldham, M. A practical three-dimensional dosimetry system for radiation therapy. Med Phys 2006; 33: 39623972.Google Scholar
34 Iqbal, K, Iqbal, M M, Akram, M, Altaf, S, Buzdar, S A. Dosimetric verification and quality assurance for intensity-modulated radiation therapy using Gafchromic® EBT3 film. J Radiother Pract 2018; 17 (1): 8595.Google Scholar
35 Iqbal, K, Ibbott, G S, Lafratta, R G, Gifford, K A, Buzdar, S A. Dosimetric characterisation of anthropomorphic PRESAGE® dosimeter and EBT2 film for partial breast radiotherapy. J Radiother Pract 2018; 17 (1): 96103.Google Scholar
36 Gifford, K A, Iqbal, K, Grant, R L, Buzdar, S A, Ibbott, G S. Dosimetric verification of a commercial brachytherapy treatment planning system for a single Entry APBI Hybrid Catheter Device by PRESAGE® and Radiochromic Film. Brachytherapy 2013; 12: 1177.Google Scholar
37 RTOG/NSABP. NSABP Protocol B-39/RTOG Protocol 0413: a randomized phase III study of conventional whole breast irradiation (WBI) versus partial breast irradiation (PBI) for women with stage 0, I, or II breast cancer. Philadelphia, PA: RTOG; 2009.Google Scholar
Figure 0

Figure 1 SAVI™ 6-1 size with peripheral struts expanded (Courtesy: Cianna Medical).

Figure 1

Figure 2 (a) PRESAGE®, dosimeter with EBT2 films inserted. (b) Brachytherapy treatment plan with dose distribution in three dimensions.

Figure 2

Figure 3 (a) Breast contours figures [external breast body, cavity, planning target volume evaluation (PTV_EVAL), skin]. (b) External body-PTV for plan optimisation: −1 mm, −3 mm, −5 mm.

Figure 3

Figure 4 (a) Calibration curve [optical density (OD) to dose] for the PRESAGE®. (b) Calibration curve (OD to dose) for the EBT2 film.

Figure 4

Table 1 Point skin dose differences between Oncentra®, PRESAGE® and EBT2

Figure 5

Figure 5 (a) Comparison of the calculated and measured point skin doses for PRESAGE®, Oncentra® Brachy Planning TPS and EBT2 film for film 1. (b) Comparison of the calculated and measured point skin doses for PRESAGE®, Oncentra® Brachy Planning TPS and EBT2 film for film 2.

Figure 6

Table 2 Planning target volume evaluation (PTV_EVAL) parameters for Oncentra® and PRESAGE®

Figure 7

Figure 6 (a) Dose–volume histogram comparison between Oncentra® and PRESAGE® for skin. (b) Plot of dose versus relevant planning target volume evaluation coverage parameters for Oncentra® and PRESAGE®.

Figure 8

Table 3 Line profile comparisons between Oncentra®, PRESAGE® and EBT2 film

Figure 9

Figure 7 (a) Line profile comparison between Oncentra®, PRESAGE® and EBT2 film dose distributions of axial slice. (b) Line profile comparison between Oncentra®, PRESAGE® and EBT2 film dose distributions of axial slice.

Figure 10

Figure 8 Dose–volume histogram comparisons between Oncentra® and PRESAGE® dose distributions. (a) External-planning target volume for plan optimisation (PTV_OPT). (b) External-PTV_OPT, −1 mm. (c) External-PTV_OPT, −3 mm. (d) External-PTV_OPT, −5 mm.

Figure 11

Figure 9 Gamma map comparisons of PRESAGE®/optical-CT , EBT2 film and Oncentra® treatment planning.