Hostname: page-component-745bb68f8f-lrblm Total loading time: 0 Render date: 2025-02-06T11:34:05.871Z Has data issue: false hasContentIssue false

In vivo experiences with magnetic resonance imaging scans in Vibrant Soundbridge type 503 implantees

Published online by Cambridge University Press:  23 April 2018

I Todt*
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
P Mittmann
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
A Ernst
Affiliation:
Department of Otolaryngology, Unfallkrankenhaus Berlin, Germany
S Mutze
Affiliation:
Department of Radiology, Unfallkrankenhaus Berlin, Germany
G Rademacher
Affiliation:
Department of Radiology, Unfallkrankenhaus Berlin, Germany
*
Author for correspondence: Dr Ingo Todt, Department of Otolaryngology, Unfallkrankenhaus Berlin, Warener Str. 7, D-12683 Berlin, Germany Fax: +49 30 5681 4303 E-mail: todt@gmx.net
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To observe the effects of magnetic resonance imaging scans in Vibrant Soundbridge 503 implantees at 1.5T in vivo.

Methods:

In a prospective case study of five Vibrant Soundbridge 503 implantees, 1.5T magnetic resonance imaging scans were performed with and without a headband. The degree of pain was evaluated using a visual analogue scale. Scan-related pure tone audiogram and audio processor fitting changes were assessed.

Results:

In all patients, magnetic resonance imaging scans were performed without any degree of pain or change in pure tone audiogram or audio processor fitting, even without a headband.

Conclusion:

In this series, 1.5T magnetic resonance imaging scans were performed with the Vibrant Soundbridge 503 without complications. Limitations persist in terms of magnetic artefacts.

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2018 

Introduction

Magnetic resonance imaging (MRI) scanners have become a regular diagnostic tool in hospitals and out-patient departments, with specific advantages over other radiological options. It has been estimated that 50–75 per cent of pacemaker wearers will need to have an MRI during their lifetime.Reference Kalin and Stanton1 Investigation of the MRI behaviour of implantable hearing systems is also very important.

Although cochlear implant systems have their brand-specific advantages and disadvantages in terms of demagnetisation, pain occurrence and magnet dislocations,Reference Nospes, Mann and Keilmann2Reference Kim, Kim, Park, Kim, Kim and Choi4 the situation for active middle-ear implants is different. In contrast to the non-MRI-compatible active middle-ear systems (i.e. the middle-ear transducer, Carina and direct acoustic cochlear stimulator (Cochlear, Sydney, Australia) systems), the Vibrant Soundbridge® implant system is the only MRI-compatible system. The Vibrant Soundbridge requires more detailed observations to determine its MRI behaviour.

The Vibrant Soundbridge contains two magnets: the audio processor containing a holding magnet, and the magnetic floating mass transducer. As the previous Vibrant Soundbridge type 502 generation showed side effects, including pain and transfer function affecting floating mass transducer dislocations,Reference Todt, Wagner, Goetze, Scholz, Seidl and Ernst5 the new Vibrant Soundbridge type 503 generation was designed to offer better compatibility with MRI scanning. In order to prevent implant receiver dislocations caused by the magnetic field, wings were attached to the body of the implant to allow the implant receiver to be screwed into the skull. Additionally, the internal floating mass transducer magnet was changed to a tri-polar magnet to direct itself in the case of an MRI scan in the magnetic field, in order to minimise torsion-related floating mass transducer movements.Reference Schnabl, Wolf-Magele, Pok, Hirtler, Heinz and Sprinzl6

The current study aimed to investigate the consequences of an MRI scan performed in patients implanted with the Vibrant Soundbridge 503 system.

Materials and methods

Five patients implanted with a 503 Vibrant Soundbridge between 2014 and 2016 underwent an MRI scan of the head (Ingenia MRI system; Philips Medical Systems, Best, Netherlands). A headband was attached before the scan. If the first scan was performed without any complications in terms of pain, discomfort, hearing ability with the audio processor, and hearing ability without the audio processor, the same MRI scan was repeated without a headband.

The imaging parameters for the two-dimensional T2-weighted turbo spin echo sequences were: repetition time = 3300 ms, echo time = 120 ms, slice thickness = 1.5 mm, reconstruction resolution of 0.55 × 0.55 × 1.5 mm, field of view = 120  × 120, 12 slices. The imaging parameters for the two-dimensional T1-weighted turbo spin echo sequences were: repetition time = 550 ms, echo time = 20 ms, slice thickness = 3 mm, reconstruction resolution of 0.23 × 0.23 × 3 mm, field of view = 120 × 120, 20 slices.

A pure tone audiogram was acquired before the first scan and after the second MRI scan. Individual attachments of the floating mass transducer are described in Table I.

Table I Individual attachments of Vibrant Soundbridge

The transfer function of the Vibrant Soundbridge was evaluated subjectively.

The study was reviewed and supported by the institutional review board for Unfallkrankenhaus Berlin (‘IRB-UKB-HNO-2016/11’), and has been conducted according to the principles expressed in the Declaration of Helsinki. This study was supported by Med-El, Innsbruck, Austria.

Results

After the first series of scans, all five patients with a headband reported no pain related to the attachment of the headband. During scanning, the five patients showed no pain as evaluated using the visual analogue scale. Subjective hearing without the audio processor was unchanged, and even hearing with the audio processor evaluated by pure tone audiogram was unchanged.

In the second series of scans performed without the headband, no pain was reported. During scanning, all patients reported no pain as evaluated using a visual analogue scale. Hearing measured without the audio processor was unchanged, and, again, even hearing with the audio processor was unchanged compared to pre-scan assessment (Table II). Magnet strength was subjectively unchanged in all cases.

Table II Mean hearing thresholds pre- and post-scanning

Data represent hearing thresholds (in dB). SDM = standard deviation of the mean

The two Vibrant Soundbridge system magnets caused scanning artefacts that consisted of two artefact centres. The floating mass transducer related centre made assessment of the internal auditory canal and cochlea impossible (Figure 1).

Fig. 1 Magnetic resonance imaging artefact of a Vibrant Soundbridge 503 system in the (a) axial and (b) sagittal planes.

Discussion

The MRI behaviour of implants is of increasing importance, as the use of MRI scans is widespread as a common diagnostic radiological tool, and the number of patients with active middle-ear implants is rising.

Observations of the MRI behaviour of active middle-ear implants are rare. Apart from studies on the Soundtec systems,Reference Dyer, Nakmali and Dormer7 only a few observations of the Vibrant Soundbridge have been published.

The Vibrant Soundbridge 502 was initially investigated in vitro. It was observed that changes can occur to the floating mass transducer position in the middle ear and to the transfer function, but no disarticulation of the ossicular chain was found at 1.5 T.Reference Todt, Rademacher, Wagner, Schedlbauer, Wagner and Basta8, Reference Jesacher, Kiefer, Zierhofer and Fauser9

Clinically, it was shown in a group of 63 Vibrant Soundbridge 502 implantees that 13 implantees (20.6 per cent) had undergone 19 MRI scans.Reference Todt, Wagner, Goetze, Scholz, Seidl and Ernst5 Twelve of the patients were implanted with the floating mass transducer fixed to the long process of the incus. One patient was bilaterally implanted with the floating mass transducer attached to the stapes head. The further occurrence of major side effects such as pain (41.7 per cent), the need to refit (50 per cent) or an operative relocation of the floating mass transducer (16.7 per cent) was recorded. Three patients (25 per cent) described a ‘loud bang’ during the scan, and two patients (17 per cent) complained of a loud, continuous noise.

In a second clinical study of Vibrant Soundbridge 502 implantees with a floating mass transducer alternatively attached, four patients underwent five MRI scans.Reference Renninger, Ernst and Todt10 Complications included pain at the receiver site and a change in transfer function in one patient.

The observed side effects led to changes in the design of the type 502 implant, including a tri-polar design of the floating mass transducer magnet and the option to fix the receiver with a screw.

Because magnetic field forces lead to torsional force on the receiver implant magnet, which turns up the magnet/implant, pain occurs at the scalp. Screws prevent such movement and thus the accompanying pain. The use of a tri-polar floating mass transducer magnet eliminates torsional and demagnetisational forces on the magnet in the magnetic field.

A recent temporal bone study of the Vibrant Soundbridge 503 implant showed that neither dislocation of the floating mass transducer, disarticulation of the ossicular chain nor a change in transfer function occurs at different coupling sites of the floating mass transducer.Reference Schnabl, Wolf-Magele, Pok, Hirtler, Heinz and Sprinzl6 Likewise, the implant function is not affected by the scan.

  • Magnetic resonance imaging (MRI) behaviour of active middle-ear implants is problematic in terms of pain, transfer function and artefacts

  • With the new generation of implants, MRI behaviour has improved, but artefact-related limitations persist

This experimental finding is consistent with the clinical observations made in our study. Neither the transfer function, indicating dislocation of the floating mass transducer, nor hearing without the audio processor were changed by the scan. Pain as the most frequent complicationReference Todt, Wagner, Goetze, Scholz, Seidl and Ernst5, Reference Renninger, Ernst and Todt10 was not observed, either with or without a headband, which is related to the internal screw fixation of the implant.

The artefacts of the receiver implant and the floating mass transducer persist, despite the change in design of the Vibrant Soundbridge 503. Even a positionally changed or artefact-reduced MRI sequence would not allow an accurate assessment of the internal auditory canal or the cochlea, because the floating mass transducer artefact does not allow any variations in middle-ear position.

Limitations concerning interpretation of the results include the fact that not all of the various coupling opportunities of the floating mass transducer in the middle ear were observed. However, it can be assumed that the new construction of the floating mass transducer and its expected behaviour in the MRI scanner are not dependent on the method of coupling.

Conclusion

Magnetic resonance imaging was performed with the Vibrant Soundbridge 503 implant in vivo without complications. Limitations persist in terms of magnetic artefacts.

References

1Kalin, R, Stanton, MS. Current clinical issues for MRI scanning of pacemaker and defibrillator patients. Pacing Clin Electrophysiol 2005;28:326–8.Google Scholar
2Nospes, S, Mann, W, Keilmann, A. Magnetic resonance imaging in patients with magnetic hearing implants: overview and procedural management [in German]. Radiologe 2013;53:1026–32.Google Scholar
3Grupe, G, Wagner, J, Hofmann, S, Stratmann, A, Mittmann, P, Ernst, A et al. Prevalence and complications of MRI scans of cochlear implant patients [in German]. HNO 2016;64:156–62.Google Scholar
4Kim, BG, Kim, JW, Park, JJ, Kim, SH, Kim, HN, Choi, JY. Adverse events and discomfort during magnetic resonance imaging in cochlear implant recipients. JAMA Otolaryngol Head Neck Surg 2015;141:4552.Google Scholar
5Todt, I, Wagner, J, Goetze, R, Scholz, S, Seidl, R, Ernst, A. MRI scanning in patients implanted with a Vibrant Soundbridge. Laryngoscope 2011;121:1532–5.Google Scholar
6Schnabl, J, Wolf-Magele, A, Pok, SM, Hirtler, L, Heinz, G, Sprinzl, GM. Magnetic resonance imaging compatibility of a new generation of active middle ear implant: a clinically relevant temporal bone laboratory study. Otol Neurotol 2016;37:222–7.Google Scholar
7Dyer, RK Jr, Nakmali, D, Dormer, KJ. Magnetic resonance imaging compatibility and safety of the SOUNDTEC Direct System. Laryngoscope 2006;116:1321–33.Google Scholar
8Todt, I, Rademacher, G, Wagner, F, Schedlbauer, E, Wagner, J, Basta, D et al. MRI safety of the floating mass transducer. Otol Neurotol 2010;9:1435–40.Google Scholar
9Jesacher, MO, Kiefer, J, Zierhofer, C, Fauser, C. Torque measurements of the ossicular chain: implication on the MRI safety of the hearing implant Vibrant Soundbridge. Otol Neurotol 2010;31:676–80.Google Scholar
10Renninger, D, Ernst, A, Todt, I. MRI scanning in patients implanted with a round window or stapes coupled floating mass transducer of the Vibrant Soundbridge. Acta Otolaryngol 2016;136:241–4.Google Scholar
Figure 0

Table I Individual attachments of Vibrant Soundbridge

Figure 1

Table II Mean hearing thresholds pre- and post-scanning

Figure 2

Fig. 1 Magnetic resonance imaging artefact of a Vibrant Soundbridge 503 system in the (a) axial and (b) sagittal planes.