Reliability of temperature determination from curve-fitting in multi-wavelength pyrometery

Published online by Cambridge University Press:  09 July 2013

P.A. Ni*
Affiliation:
Lawrence Berkeley National Laboratory, Berkeley, California
R.M. More
Affiliation:
Lawrence Berkeley National Laboratory, Berkeley, California
F.M. Bieniosek
Affiliation:
Lawrence Berkeley National Laboratory, Berkeley, California
*
Address correspondence and reprint requests to P.A. Ni, Lawrence Berkeley National Laboratory, Berkeley, CA 94720. E-mail: pani@lbl.gov
Rights & Permissions [Opens in a new window]

Abstract

This paper examines the reliability of a widely used method for temperature determination by multi-wavelength pyrometry. In recent warm dense matter experiments with ion-beam heated metal foils, we found that the statistical quality of the fit to the measured data is not necessarily a measure of the accuracy of the inferred temperature. We found a specific example where a second-best fit leads to a more realistic temperature value. The physics issue is the wavelength-dependent emissivity of the hot surface. We discuss improvements of the multi-frequency pyrometry technique, which will give a more reliable determination of the temperature from emission data.

Type
Research Article
Copyright
Copyright © Cambridge University Press 2013 

INTRODUCTION

Warm dense matter (WDM) can be produced by several heating technologies including lasers, pulsed electric current, electron, ion and X-ray beam heating, shock-wave heating, etc. (DOE, 2009; Hoffmann et al. Reference Hoffmann, Blazevic, Ni, Rosmej, Roth, Tahir, Tauschwitz, Udrea, Varentsov, Weyrich and Maron2005; Sasaki et al. Reference Sasaki, Yano, Nakajima, Kawamura and Horioka2006; Ng et al. Reference Ng, Ao, Perrot, Dharma-Wardana and Foord2005; Lee et al. Reference Lee, Baldis, Cauble, Landen, Wark, NG, Rose, Lewis, Riley, Gauthier and Audebert2002). Temperature measurements are essential to WDM science. In most cases, the deposited energy is determined accurately but at present there is no universal method to directly measure the temperature. As a consequence, the specific heat C = dE/dT is not accurately determined and this is a key unknown aspect of the equation of state (EOS). There is also a class of experiments, which aims to study phase transitions: these may be high-temperature evaporation of refractory metals, a liquid-vapor critical point, or an abrupt change of electronic structure (metal-insulator transition). Such transitions typically occur at definite temperatures and an accurate measurement of the transition temperature is obviously a key step for building up the quantitative science of WDM. Today the critical points of refractory metals are uncertain to something like 50% and this uncertainty is a leading source of EOS uncertainty for WDM expanded below solid density. At higher energy-density (HEDP), shock Hugoniot measurements can accurately determine for the material density, energy-density and pressure, but different EOS models predict different temperatures for the same (energy/pressure/density) shocked conditions, so again temperature is a key uncertainty for HEDP EOS data. Finally, almost all of the computer codes used to predict the properties of WDM (or hotter plasmas) use density and temperature as inputs and for this reason it will be easier to directly compare theory and experiment if one can accurately measure the temperatures attained in any given experiment.

Pyrometry (Dewitt, Reference Dewitt1998), the determination of surface temperature by analysis of thermally emitted light, is a well-known and commonly used method for measuring temperature of hot matter. While being a popular technique, pyrometry is known for two drawbacks: a limitation to measure only the surface temperature, and limited accuracy in experiments where surface emissivity is not known. This paper will discuss the second issue.

The normal surface emissivity E(λ) is (by definition) the ratio of the observed emission intensity to the black-body function B λ(T):

$$I(\lambda)=E(\lambda)\times B_{\lambda}(T)$$

The difficult point is that while the surface emissivity is known for cold matter (tabular data exists), it is generally unknown above melting (T ≥ 2000 K–3000 K) temperatures, i.e., in the WDM regime. The fundamental reason for this is the fact that the high-frequency (optical) alternate current conductivity, or equivalently the optical constants n, k, which determine the emissivity, are not known for material at WDM conditions.

An attempt to improve the situation by simultaneously measuring emission at many wavelengths (multi-wavelength pyrometery) (Dewitt, Reference Dewitt1998; Ni et al., Reference Ni, Kulish, Mintsev, Nikolaev, Ternovoi, Udrea, Hug, Tahir and Hoffmann2008; Reference Ni and W.L.2011) does not solve the problem because there is a new unknown emissivity E(λ) at each new wavelength λ and there always N + 1 unknowns, where N is the number of direct measurements. Thus from the mathematical point of view, multi-wavelength pyrometry cannot produce a unique temperature. If emissivity is not measured and there is no information about its behavior, experimenters can assume a linear or quadratic dependence of E(λ) on wavelength λ and optimize this “gray-body” fit to the multi-frequency emission measurements. The temperature produced by this fit is often reported as the temperature “observed” in the experiment.

REAL-WORLD EXAMPLE OF “BLIND” CURVE FITTING

Unfortunately, the accuracy of temperature determination does not reflect the accuracy of gray-body fits to spectral data. A recent real-life example, which motivates this paper, was encountered during WDM experiments at the NDCX-I, heavy ion induction LINAC (Bieniosek et al., Reference Bieniosek, Barnard, Friedman, Henestroza, Jung, Leitner, Lidia, Logan, More, Ni, Roy, Seidl and Waldron2010a). In the experiments, a Au foil of thickness 0.15 µm was heated by a 300 keV K+, 250 kW/cm2 beam and the foil temperature was measured by a standard high-speed streak-camera optical pyrometer (SOP) technique (Bieniosek et al. Reference Bieniosek, Henestroza, Lidia and Ni2010b): thermal emission was recorded by a system consisting of a spectral grating and a fast optical streak camera. The entire SOP setup was calibrated absolutely with a tungsten ribbon lamp, which is a NIST-traceable calibration source. Absolute emission spectra from 400 nm to 850 nm were recorded with about 2% accuracy and a time-resolution of approximately 50 ns (Fig. 1). The data were analyzed by the usual method of optical pyrometry, i.e., the emission spectrum was fit to a blackbody function multiplied by an emissivity E(λ), which was assumed to be a polynomial (in our case quadratic) function of a wavelength in the visible range.

Fig. 1. (Color online) Absolute emission spectrum as a function of time from a beam heated Au target by NDCX-I. Temperature is determined from spectrum at each moment of time. Lines correspond to collisional Au-I vapor excitation by ion beam.

The resulting best fit, shown in Figure 2, leads to a temperature of 7200 K (the spectrum shown corresponds to a time of 800 ns, where the radiation is a maximum). The best least-square fit is determined by the global minimum of the sums of residuals. The deviation of the fit from the experimental data is very small, and if one had stopped here one would have believed the temperature.

Fig. 2. (Color online) (a) Best fit (with a fitted temperature T = 7200 K) of the spectrometer data (from the data in Fig. 1b) using an emissivity assumed to have a quadratic form. (b) Inferred best-fit emissivity as function of wavelength.

However the energy deposited by the NDCX-I beam is well-known (fluence is about 250 kW/cm2), and with all the energy accounted for, the highest expected temperature could not exceed 3400 K (Bieniosek et al., 2010). The energy required to heat a 0.150 micron Au foil up to the boiling point (not counting the latent heat of evaporation) is approximately 0.2 J/cm2. To deliver this energy in 1 µs requires 200 kW/cm2. A thick target would be cooled by thermal conduction, but our thin targets do not cool in this way (lateral thermal conduction is negligible on our time-scale). However, the target cools by evaporation at high temperatures. We estimate the cooling rate at a temperature equal to the normal boiling point, T b = 3240 K for Au. At this temperature the saturation pressure of Au vapor would be 1 atmosphere. This corresponds to a gas density of approximately 2.24 × 1018 atoms/cm3. In equilibrium, the evaporation rate would just balance condensation from the saturated vapor, but in vacuum there is little condensation. Assuming a sticking coefficient close to unity, that balance implies an evaporation rate of 3.3 × 1022 atom/cm2/s. Each evaporated atom carries away energy of approximately the latent heat of evaporation, about 3.51 eV for Au. Thus the evaporation cooling heat flux is about 18.5 kW/cm2. This should be multiplied by a factor of 2 for a thin foil or a factor of 4 for spherical droplets (ratio of evaporating surface area to projected area facing the beam). If the ion heating does not deposit more energy than 270 kW/cm2 (= 200 kW/cm2 for initial temperature rise + 70 kW/cm2 for evaporative cooling), the target is expected to reach a temperature lower than the normal boiling point of 3240 K. Based on these numbers, a temperature of 7200 K is not possible with the NDCX-I ion beam.

After a careful analysis triggered by this discrepancy, a “second-best” optimum fit to the same data, which yields a much lower temperature, 2400 K (Fig. 3) was found. Both Newton-Gauss and Levenberg-Marquardt, algorithms commonly used least-square optimization algorithms (Ni et al., Reference Ni, Kulish, Mintsev, Nikolaev, Ternovoi, Udrea, Hug, Tahir and Hoffmann2008), lead to the same results. The 2400 K fit is in much better agreement with the thermodynamic analysis but is not such a nice fit to the measured spectrum from a statistical point of view. However, it is important to note that the emissivity E(λ) resulting from the 2400 K fit (Fig. 3b) is very much like the emissivity of normal Au in this spectral range (Watanabe et al., Reference Watanabe, Susa, Fukuyama and Nagata2003), while the emissivity inferred from the high-temperature fit in Figure 2b has the opposite curvature.

Fig. 3. (Color online) (a) Second “best” fit of the spectrometer data (with a fitted temperature T = 2400 K) (from the data in Fig. 1b) using an emissivity assumed to have a quadratic form. (b) Inferred fit emissivity as function of wavelength. The parameters derived in figure are also at a minimum in the mean square deviation, but have a larger deviation (worse fit) than those in Figure 2.

The summary of the comparison is that the beautiful fit in Figure 1a is a 7200 K blackbody source times the emissivity of Figure 2b. The less beautiful fit in Figure 3a is a 2400 K blackbody times the emissivity of Figure 3b. But for two good reasons — the total energy available and the shape of Figure 3b — the second fit is much more likely to be correct. The important lesson is that, generally speaking, the accuracy of temperature determination has no relation to the accuracy of the fit to spectral data. One cannot measure the accuracy of a temperature extracted from conventional pyrometry by the quality of the curve-fit to the spectrum: looking only at Figure 1a, it would look like 1% precision, but seeing the whole picture, the temperature error from Figure 1a about 50%!

POLARIZATION OF SELF EMISSION AS A KEY TO SOLVE THE AMBIGUITY PROBLEM

An approach to address the temperature ambiguity, inherent to multi-wavelength pyrometers, is to find other manifestations of emissivity. It turns out that in addition to self-emission spectra, there is more information available that is generally overlooked, namely the polarization of the self-emission, occurring at a non-zero angle to a surface: generally s-polarized (parallel to surface) E-field emission is less intense than p-polarized emission. If a surface is perfectly smooth and clean, and the material beneath the surface is homogenous, the reflectivity, r = 1 − ɛ is related by Fresnel to the complex dielectric function, ξ (λ). For a non-zero angle from the surface normal, α, the reflectivity r is different for the p and s components of electrical vector, and the polarized spectral emission density can be described by

(1)$$\eqalign{I_p \lpar {\rm \lambda}\comma \; {\rm \alpha} \rpar &=\left({1 - \left\vert {\displaystyle{{{\rm \xi} \lpar n\comma \; k\rpar \cdot \cos \lpar {\rm \alpha} \rpar - \sqrt {{\rm \xi} \lpar n\comma \; k\rpar - \sin ^2 \lpar {\rm \alpha} \rpar } } \over {{\rm \xi} \cdot \cos \lpar {\rm \alpha} \rpar +\sqrt {{\rm \xi} - \sin ^2 \lpar {\rm \alpha} \rpar } }}} \right\vert ^2 } \right)\ast B_{\rm \lambda} \lpar T\rpar \comma \; \cr I_s \lpar {\rm \lambda}\comma \; {\rm \alpha} \rpar &=\left({1 - \left\vert {\displaystyle{{\cos \lpar {\rm \alpha} \rpar - \sqrt {{\rm \xi} \lpar n\comma \; k\rpar - \sin ^2 \lpar {\rm \alpha} \rpar } } \over {\cos \lpar {\rm \alpha} \rpar +\sqrt {{\rm \xi} \lpar n\comma \; k\rpar - \sin ^2 \lpar {\rm \alpha} \rpar } }}} \right\vert ^2 } \right)\ast B_{\rm \lambda} \lpar T\rpar \comma \; \cr {\rm \xi} &=\lpar n - ik\rpar ^2 - \hbox{complex dielectric function}.}$$

As one can see, if the absolute intensities of polarization components at (at least) two angles (four values in total) are measured, there is a unique solution for T, and optical constants, n and k. One can take data at more angles and obtain an over-determined system of equations, which is favorable when measurements have finite accuracy. In contrast to the multi-wavelength pyrometery, addition of more measurements at different wavelengths does not improve accuracy since there are always more unknowns than measured values. The combination of polarization and multi-wavelength pyrometer data, can thus reduce the overall ambiguity of temperature determination. A diagnostic based on the above principle, is called polarization pyrometer and is described in details in (More et al., Reference More, Goto, Graziani, Ni and Yoneda2010; Ni et al., Reference Ni, More, Yoneda and Bieniosek2012).

CONCLUSION

The weak point of the pyrometry technique is the fact that, over the visible range, blackbody functions with different temperatures (combined with different emissivities) can fit the emission data almost equally well. The example in Figures 1 to 4 utilizes precisely the usual procedure in optical pyrometry and very similar methods are used in many laboratories for determining temperatures produced by shock-waves or other warm-dense matter experiments. A practical consequence from this finding is that if possible, emssivity should be measured independently and common sense verification of maximum temperature must be carried out. We were lucky to have a reference to emissivity value, but what about the experiments at much higher temperatures, where there is no emissivity or beam fluence reference? Addition of polarization resolved measurements to the spectral emission has a potential to solve the ambiguity problem.

ACKNOWLEDGMENTS

This work was performed under the auspices of the U.S. Department of Energy by LBNL under Contract No. DE-AC02–05CH11231. The authors would like to thank Steve Lidia for facilitating experiments at the NDCX-I. Finally, the authors sadly note the untimely passing of our friend, colleague, and co-author Frank Bieniosek. His clarity of scientific vision, his energy and enthusiasm, his guidance, and his companionship are much missed.

References

REFERENCES

Bieniosek, F.M., Barnard, J.J., Friedman, A., Henestroza, E., Jung, J.-Y., Leitner, M.A., Lidia, S., Logan, B.G., More, R.M., Ni, P.A., Roy, P.K., Seidl, P.A., Waldron, W.L. (2010 a). Ion-beam-driven warm dense matter experiments. J. Phys. Conf. Ser. 244 032028. HIFAN 1747, LBNL-2378E.CrossRefGoogle Scholar
Bieniosek, F.M., Henestroza, E., Lidia, S. & Ni, P.A. (2010 b). Diagnostics for ion beam driven high energy density physics experiments. Rev. Sci. Instrum. 81, 10E112.CrossRefGoogle ScholarPubMed
Dewitt, D.P. (1998). Theory and Practice of Radiation Thermometry. New York: John Wiley & Sons, Inc.Google Scholar
DOE (U.S. Department of Energy). (2009). Basic Research Needs for High Energy Density Laboratory Physics — Report of the Workshop on HEDLP Research Needs. Washington: U.S. Department of Energy, 124.Google Scholar
Hoffmann, D.H.H., Blazevic, A., Ni, P., Rosmej, O., Roth, M., Tahir, N., Tauschwitz, A., Udrea, S., Varentsov, D., Weyrich, K. & Maron, Y. (2005). Present and future perspectives for high energy density physics with intense heavy ion and laser beams. Laser Part. Beams 23, 395–395.CrossRefGoogle Scholar
Lee, R.W., Baldis, H.A., Cauble, R.C., Landen, O.L., Wark, J.S., NG, A., Rose, S.J., Lewis, C., Riley, D., Gauthier, J.-C. & Audebert, P. (2002). Plasma-based studies with intense X-ray and particle beam sources. Laser Part. Beams 20, 527536.CrossRefGoogle Scholar
More, R.M., Goto, M., Graziani, F., Ni, P.A. & Yoneda, H. (2010). Emission of visible light by hot dense metals. Plasma Fus. Res. 5, S2007.Google Scholar
Ng, A., Ao, T., Perrot, F., Dharma-Wardana, M.W.C. & Foord, M.E. (2005). Idealized slab plasma approach for the study of warm dense matter. Laser Part. Beams 23, 527537.CrossRefGoogle Scholar
Ni, P.A., Kulish, M.M., Mintsev, I., Nikolaev, D.N., Ternovoi, V.Ya., Udrea, S., Hug, A., Tahir, N.A. & Hoffmann, D.H.H. (2008). Temperature measurements of warm-dense-matter generated by heavy-ion beams. Laser Part. Beams 26, 583589.CrossRefGoogle Scholar
Ni, P.A., More, R.M., Yoneda, H. & Bieniosek, F.M. (2012). Polarization pyrometry – An improvement to multi-wavelength optical pyrometry. Rev. Sci.c Instrum. 83, 123501.CrossRefGoogle ScholarPubMed
Ni, P.A., W.L., F.M (2011). Waldron,“Multi-channel optical pyrometer for sub-nanosecond temperature measurements at NDCX-I/II.” Hi. Temp. Hi. Pres. 40, 151160.Google Scholar
Sasaki, T., Yano, Y., Nakajima, M., Kawamura, T. & Horioka, K. (2006). Warm-dense-matter studies using pulse-powered wire discharges in water. Laser Particle Beams 24, 371380.CrossRefGoogle Scholar
Watanabe, H., Susa, M., Fukuyama, H. & Nagata, K. (2003). Phase Dependence (Liquid/Solid) of Normal Spectral Emissivities of Noble Metals at Melting Points. Internat. J. Thermophys. 24, 223.Google Scholar
Figure 0

Fig. 1. (Color online) Absolute emission spectrum as a function of time from a beam heated Au target by NDCX-I. Temperature is determined from spectrum at each moment of time. Lines correspond to collisional Au-I vapor excitation by ion beam.

Figure 1

Fig. 2. (Color online) (a) Best fit (with a fitted temperature T = 7200 K) of the spectrometer data (from the data in Fig. 1b) using an emissivity assumed to have a quadratic form. (b) Inferred best-fit emissivity as function of wavelength.

Figure 2

Fig. 3. (Color online) (a) Second “best” fit of the spectrometer data (with a fitted temperature T = 2400 K) (from the data in Fig. 1b) using an emissivity assumed to have a quadratic form. (b) Inferred fit emissivity as function of wavelength. The parameters derived in figure are also at a minimum in the mean square deviation, but have a larger deviation (worse fit) than those in Figure 2.