A treatment planning study comparison between supine and prone position for different lung tumour locations using CyberKnife TPS

Published online by Cambridge University Press:  17 May 2021

Hiroaki Akasaka
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan Division of Radiation Oncology, Kobe University Graduate School of Medicine, Chuo-ku Kobe, Hyogo650-0017, Japan
Kazufusa Mizonobe
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Yuya Oki
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Kazuyuki Uehara
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Aya Harada
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Shuichiro Miyazaki
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Keiji Kitatani
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Tomonori Yabuuchi
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
Takeaki Ishihara
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Daisuke Miyawaki
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Masao Nakayama
Affiliation:
Division of Radiation Oncology, Kobe University Graduate School of Medicine, Chuo-ku Kobe, Hyogo650-0017, Japan
Naritoshi Mukumoto
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Ryuichi Yada
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Ai Nakaoka
Affiliation:
Division of Radiation Oncology, Kobe University Graduate School of Medicine, Chuo-ku Kobe, Hyogo650-0017, Japan
Hikaru Kubota
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Hiroki Kawaguchi
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Satoshi Seno
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Ryohei Sasaki
Affiliation:
Division of Radiation Oncology, Kobe University Hospital, Chuo-ku Kobe, Hyogo650-0017, Japan
Hiroshi Mayahara*
Affiliation:
Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan
*
Author for correspondence: Hiroshi Mayahara, Division of Radiation Oncology, Kobe Minimally Invasive Cancer Center, Chuo-ku Kobe, Hyogo650-0046, Japan. Tel: (+81)-78-304-4100. Fax: (+81)-78-304-0041. E-mail: mayahara@k-mcc.net
Rights & Permissions [Opens in a new window]

Abstract

Aim:

CyberKnife is the most advanced form of stereotactic body radiotherapy (SBRT) system that uses a robotic arm to deliver highly focused beams of radiation; however, a limitation is that it only irradiates from ceiling to floor direction. In patients with posterior lungs tumour who are positioned supine, normal lung tissue may suffer undesirable radiation injuries. This study compares the treatment planning between the prone set-up and the supine set-up for lung cancer in CyberKnife SBRT to decrease normal lung dose to avoid radiation side effects.

Materials and methods:

A human phantom was used to generate 108 plans (54 for prone and 54 for supine) using the CyberKnife planning platform. The supine and prone plans were compared in terms of the dosimetric characteristics, delivery efficiency and plan efficiency.

Results:

For posterior targets, the area of low-dose exposure to normal lungs was smaller in the prone set-up than in the supine set-up. V10 of the lungs was 7·53% and 10·47% (p < 0·001) in the anterior region, and 10·78% and 8·03% (p < 0·001) in the posterior region in the supine and prone set-up plans, respectively.

Conclusions:

The comparison between the prone set-up and the supine set-up was investigated with regard to target coverage and dose to organs at risk. Our results may be deployed in CyberKnife treatment planning to monitor normal tissue dose by considering patient positioning. This may assist in the design of better treatment plans and prevention of symptomatic radiation pneumonitis in lung cancer patients.

Type
Original Article
Copyright
© The Author(s), 2021. Published by Cambridge University Press

Introduction

Stereotactic body radiotherapy (SBRT) facilitates the administration of large doses per fraction and a small number of fractions to tumours, with minimal exposure to the surrounding organs. This method, which offers increased stability and precision, has been widely used for lung cancer treatment over the past 20 years. Hypofractionated high-dose SBRT has emerged to treat early-stage non-small-cell lung cancer. Reference Mehta, King, Agazaryan, Steinberg, Hua and Lee1 However, the adverse effects of this treatment often include radiation pneumonitis (RP), which can be fatal. Therefore, in the treatment planning, the spread of low doses to normal lung tissue should be minimised. Several researchers have reported that the percentage of lung volume receiving an excess dose of 5 Gy, 10 Gy, 20 Gy and 25 Gy (V5, V10, V20 and V25) and the mean lung dose (MLD) are predictive factors for RP. Reference Baker, Han and Sarangkasiri2Reference Nakamura, Nishimura and Nakayama5 To avoid RP or respiratory function decline, conventional planning or intensity-modulated radiation therapy (IMRT) Reference Bradley6Reference Kong, Hayman and Griffith8 planning could be employed by considering the combination of the gantry angle. In this regard, the CyberKnife system (Accuray Incorporated, Sunnyvale, CA) is suitable for increasing dose concentration because of its beneficial characteristics. CyberKnife, a specialised medical device for stereotactic radiosurgery treatments, consists of a compact 6-MV linear accelerator (LINAC) mounted on a robotic arm; the LINAC irradiates the target with radiation from different directions. Reference Kilby, Dooley, Kuduvalli, Sayeh and Maurer9 Moreover, it delivers the radiation dose with real-time tumour tracking that corrects for tumour motion during respiration by repositioning the radiation beam to track the moving tumour. Reference Nakamura, Nishikawa and Mayahara10

In conventional planning or IMRT planning using the medical LINAC, irradiation can be rendered through any angle in any direction. However, the irradiation angle is limited in the CyberKnife system, as its rotation is limited to 100 degrees. In a non-coplanar setting, the available beam directions have been shown to affect plan quality. Reference Rossi, Breedveld, Heijmen, Voet, Lanconelli and Aluwini11 Thus, owing to the specific geometry of the CyberKnife layout, irradiation from the space under the treatment table is impossible.

CyberKnife can only irradiate from above to its device characteristics. When a patient is positioned on his/her back, the exposure of normal organs to low-dose radiation becomes a problem when the tumour is in the posterior region. For example, in the case of lung cancer, if the patient is in the supine position, many beams pass through normal lung tissue when the tumour is located on the patient’s posterior lung side. Consequently, a large volume of the normal lung tissue is undesirably irradiated which can cause radiation injury. Thus, it is necessary to consider patient positioning during CyberKnife treatment. In this study, the comparison between the prone set-up and the supine set-up was investigated to decrease normal lung dose to avoid adverse events.

Materials and Methods

Treatment planning

For a human phantom (Kyoto Kagaku Co., Ltd., Kyoto, Japan), simulation computed tomography (CT) scans were acquired in both supine and prone positions. The human phantom included the skin, lungs and bones. A virtual target was contoured in the supine dataset. For consistency, the same contours were copied onto the prone dataset after co-registering the prone and supine CT images. The tumour location was decided as follows. First, lungs were divided into three areas (Figure 1 (a)). Then, each area was further divided into nine cross-sectional regions (Figure 1 (b)). Finally, the virtual tumour was located in each area. Fifty-four regions were generated for the supine and prone set-ups, respectively (Figure 1 (b)). For simplicity, only one tumour was set in the lung for each plan. For all plans, the target volume was set to approximately 30 cm3 in accordance with Ueyama et al. Reference Ueyama, Arimura and Takumi4 In this study, target volume was considered as the same volume as planning target volume (PTV). Using the CyberKnife planning platform (Precision version 2.0.1.1; CyberKnife (VSI) version 9.6.0), 108 plans (54 for prone and 54 for supine) were generated. The plans were designed to deliver a dose of 60 Gy per eight fractions to 95% of the target volume. For a given target, the plans in the prone and supine positions were optimised using the VOLO method. Additionally, for all treatment planning, the Monte Carlo algorithm Reference Murali, Gopalakrishna Kurup and Bhuvaneswari12 was used for the dose calculation because of the lung electron density within and/or around the treatment volume. The plans were generated by a medical physicist specialised in CyberKnife treatment.

Figure 1. CT image of the phantom in this study. (a) Coronal view of the CT image of the phantom that is divided into three regions: upper, middle and lower. (b) Axial view of the CT image of the phantom according to each region, which is further divided into 18 slices (nos. 1 to 18).

Plan evaluation

The supine and prone plans were compared in terms of the following: the dosimetric characteristics (maximum PTV dose, percentage of the prescription dose covering 95% of the volume (D95), VS5, V5, V10, V20, V25 and MLD); delivery efficiency (number of beams, treatment time and total plan monitor units (MUs)) and plan efficiency (homogeneity index (HI), conformity index (CI) and new conformity index (nCI)). The HI is defined as follows Reference Cao, Zhu and Ju13 :

$${\rm{HI}} = {{{D_{max}}} \over {{R_{xDose}}}},$$

where ${R_{xDose}}$ denotes the prescription dose. The CI is defined as follows Reference Cao, Zhu and Ju13 :

$${\rm{CI}} = {{prescription\;isodose\;volume\;\left( {c{m^3}} \right)} \over {tumor\;volume\;encompassed\;prescription\;isodose\;line\;\left( {c{m^3}} \right)}}$$

The nCI is calculated as follows Reference Cao, Zhu and Ju13 :

$${\rm{nCI}} = C{I_{coverage}}$$

Here, coverage is defined as the ratio of the target volume covered with the prescription dose to the target volume.

V5, V10, V20, V25 and MLD of the lung were reported as risk factors for RP after stereotactic radiation therapy for lung tumours. Reference Baker, Han and Sarangkasiri2Reference Nakamura, Nishimura and Nakayama5 A dose–volume histogram (DVH) analysis was performed. For the supine and prone position plans, the mean DVHs of the target and critical organs were calculated and compared for each lung region, that is, posterior, middle and anterior region.

Statistical analysis

Data were presented as mean ± standard error. Differences between groups were evaluated using the Student’s t-test. Data were considered statistically significant for p < 0·001.

Results

Plan quality

Table 1 shows the results obtained for the plan quality index. No statistically significant difference was observed in the dose coverage parameter among the each plan. The average HI values and the average CI values were calculated for each region, that is, posterior, middle and anterior region. There was a significant difference in HI, CI and nCI between the supine and prone set-up plans only when the target was seated in the anterior region.

Table 1. Dosimetric parameter in supine and prone position

Dosimetric characteristics

Table 2 shows the dosimetric characteristics of the target, and Table 3 shows the dosimetric characteristics of the lungs. The V10 of the lungs is 7·53% and 10·47% (p < 0·001) in the anterior region, 11·78% and 11·56% (p = 0·382) in the middle region, and 10·78% and 8·03% (p < 0·001) in the posterior region in the supine and prone set-up plans, respectively (Figure 2 (a)–(f)). There was a significant difference in the V5 and V10 to the lung between the supine set-up and prone set-up plans when the target was located in either the anterior or posterior region.

Table 2. Maximum dose and D95 dose in supine and prone position

Table 3. Dose volume metrics of the factors related to symptomatic RP

Figure 2. Dose distribution of the typical plans for different set-ups when (a, b) the target is located in the anterior or posterior region. (c, d) Dose distribution of the plans for the supine set-up when the target is located in the anterior or posterior region. (e, f) Dose distribution of the plans in prone set-up when the target is located in the anterior or posterior region.

Delivery efficiency

A comparable degree of complexity in the planning was observed for both the prone and supine plans. For a given target, plans in the prone and supine position required similar planning efforts and resulted in a similar number of beams. Table 4 shows that no significant difference exists between the supine and prone set-up plans with regard to the average number of beams for each side. However, there is a significant difference in the total planned MUs in the anterior and posterior regions between the supine and prone set-up plans. Lower total planned MUs translate into a faster treatment time.

Table 4. Quality factors of each plan in supine and prone positions

DVH analysis

Figure 3 shows the DVH for the target and the lung. From the DVH characteristics, it appears that the target dose was not much different for each plan. However, the low dose to the lung was lower for plans developed for the supine set-up when the target was located in the anterior region; conversely, the low dose to the lung was lower for the plans developed in the prone set-up when the target was seated in the posterior region.

Figure 3. DVH curve of a typical plan for supine and prone set-ups, respectively. DVH curve of the target (a, b), and DVH curve of the lung (c, d) when the target is located in the anterior or posterior region. The solid line indicates the supine set-up plan; the dotted line indicates the prone set-up plan.

Discussion

RP is the most severe side effect after SBRT for lung tumours. Some researchers have reported that the incidence of symptomatic RP after SBRT ranges from 9% to 29%. Reference Yamashita, Nakagawa and Nakamura14Reference Stauder, Macdonald and Olivier18 To reduce the incidence of symptomatic RP after SBRT, some studies have assessed the DVH of SBRT and strived to predict the rate of symptomatic RP. Ueyama et al. reported that the cut-off values of V20 and V10 were 4·3% and 9·7%, respectively, for symptomatic RP. Reference Ueyama, Arimura and Takumi4 Furthermore, Nakamura et al. reported that the cut-off values of V5, V10, V20, V25 and MLD to be 21·5%, 9·1%, 3·5%, 3·4% and 3·6 Gy, respectively, for symptomatic RP. Reference Nakamura, Nishimura and Nakayama5

We evaluated set-up position for lung cancer treatment using the CyberKnife system. It was obvious that the prone set-up was better when the target was located on the posterior, and the supine set-up was better when the target was located on the anterior. Ding et al. Reference Ding, Chang, Haslam, Timmerman and Solberg19 noted that CyberKnife SBRT may deliver a lower dose to healthy lung tissue than IMRT when treating tumours in the anterior region of the lung. However, the low-dose volume delivered via CyberKnife is significantly greater than that delivered via LINAC-based delivery when treating tumours in the posterior region of the lung. This is because CyberKnife SBRT plans use more MUs than conventional SBRT treatments. Reference Ding, Chang, Haslam, Timmerman and Solberg19 Mark K.H.Chan reported that the volumetric-modulated arc therapy plans resulted in higher dose around the gantry rotation path compared to the CyberKnife plans for when the tumour located in anterior lesion, whereas the CyberKnife plans were mostly associated with high doses anterior to the lesions due to the missing irradiation coming from underneath the patient when the tumour located in posterior lesions. Reference Chan, Kwong and Law20 Moreover, the CyberKnife system can be used only up to 100 degrees from the top, owing to the limited mobility of the arm. However, our results prove that changing the patient set-up could solve this problem. The prone set-up is highly beneficial in the case of spinal radiosurgery. Reference Descovich, Ma, Chuang, Larson and Barani21,Reference Fürweger, Drexler, Muacevic, Wowra, de Klerck and Hoogeman22 This study revealed that it is also beneficial for lung cancer patients with their tumour located in the posterior region to receive treatment in the prone position.

In the case of lung cancer, for a target that moves with the respiratory motion, CyberKnife SBRT can deliver radiation beams with continuous tumour tracking. Nakayama et al. analysed the synchrony log files of patients with lung tumours who were treated using the Xsight Lung Tracking (XLT) or 1-View tracking system. They reported that the tumour motion amplitude may be one of the factors that affect the model errors; therefore, the errors should be carefully analysed to determine the margins for tumours with large motion amplitudes as accurately as possible. Reference Nakayama, Nishimura and Mayahara23

This study had a limitation that real-time tumour motion tracking is not considered. However, in practice, the use of respiratory synchrony is imperative. Furthermore, the range of respiratory motion varies depending on the site of the lung tumour. When respiratory synchrony is used, there will always be motion errors. In future studies, it is thus necessary to evaluate the set-up position with consideration of respiratory synchrony.

Conclusions

This study compares the treatment planning between the prone set-up and the supine set-up for lung cancer in CyberKnife SBRT, which to the best of our knowledge, is the first such study. The comparison between the prone set-up and the supine set-up was investigated with regard to target coverage and dose to organs at risk. We expect that our findings will help radiation oncologists, radiation technologists, dosimetrists and medical physicists to decrease lung dose when CyberKnife lung SBRT will be performed and in designing optimal radiation therapy plans for patients. To extend our inferences based on a phantom study to real-life cases, we intend to conduct further investigations.

Acknowledgements

The authors thank Keisuke Okumura (Kobe University Hospital) for offering technical assistance.

Financial Support

This research received no specific grant from any funding agency, commercial or not-for-profit sectors.

Conflicts of Interest

None.

References

Mehta, N, King, C R, Agazaryan, N, Steinberg, M, Hua, A, Lee, P. Stereotactic body radiation therapy and 3-dimensional conformal radiotherapy for stage I non-small cell lung cancer: a pooled analysis of biological equivalent dose and local control. Pract Radiat Oncol 2012; 2 (4): 288295.CrossRefGoogle Scholar
Baker, R, Han, G, Sarangkasiri, S et al. Clinical and dosimetric predictors of radiation pneumonitis in a large series of patients treated with stereotactic body radiation therapy to the lung. Int J Radiat Oncol Biol Phys 2013; 85 (1): 190195.CrossRefGoogle Scholar
Matsuo, Y, Shibuya, K, Nakamura, M et al. Dose-Volume metrics associated with radiation pneumonitis after stereotactic body radiation therapy for lung cancer. Int J Radiat Oncol Biol Phys 2012; 83 (4): e545e549.CrossRefGoogle ScholarPubMed
Ueyama, T, Arimura, T, Takumi, K et al. Risk factors for radiation pneumonitis after stereotactic radiation therapy for lung tumours: clinical usefulness of the planning target volume to total lung volume ratio. Br J Radiol 2018; 91 (1086): 20170453.CrossRefGoogle ScholarPubMed
Nakamura, M, Nishimura, H, Nakayama, M et al. Dosimetric factors predicting radiation pneumonitis after CyberKnife stereotactic body radiotherapy for peripheral lung cancer. Br J Radiol 2016; 89 (1068): 20160560.CrossRefGoogle ScholarPubMed
Bradley, J. A review of radiation dose escalation trials for non-small cell lung cancer within the Radiation Therapy Oncology Group. Semin Oncol 2005; 32 (2 Suppl 3): S111S113.CrossRefGoogle ScholarPubMed
Lee, C B, Stinchcombe, T E, Rosenman, J G, Socinski, M A. Therapeutic advances in local-regional therapy for stage III non-small-cell lung cancer: evolving role of dose-escalated conformal (3-dimensional) radiation therapy. Clin Lung Cancer 2006; 8 (3): 195202.CrossRefGoogle ScholarPubMed
Kong, F M, Hayman, J A, Griffith, K A et al. Final toxicity results of a radiation-dose escalation study in patients with non-small-cell lung cancer (NSCLC): predictors for radiation pneumonitis and fibrosis. Int J Radiat Oncol Biol Phys 2006; 65 (4): 10751086.CrossRefGoogle ScholarPubMed
Kilby, W, Dooley, J R, Kuduvalli, G, Sayeh, S, Maurer, C R Jr The CyberKnife robotic radiosurgery system in 2010. Technol Cancer Res Treat 2010; 9 (5): 433452.CrossRefGoogle ScholarPubMed
Nakamura, M, Nishikawa, R, Mayahara, H et al. Pattern of recurrence after CyberKnife stereotactic body radiotherapy for peripheral early non-small cell lung cancer. J Thorac Dis 2019; 11 (1): 214221.CrossRefGoogle ScholarPubMed
Rossi, L, Breedveld, S, Heijmen, B J, Voet, P W, Lanconelli, N, Aluwini, S. On the beam direction search space in computerized non-coplanar beam angle optimization for IMRT-prostate SBRT. Phys Med Biol 2012; 57 (17): 54415458.CrossRefGoogle ScholarPubMed
Murali, V, Gopalakrishna Kurup, P G, Bhuvaneswari, N et al. Monte Carlo and ray tracing algorithms in the CyberKnife treatment planning for lung tumours—Comparison and validation. J Radiosurg SBRT 2013; 2 (2): 8598.Google ScholarPubMed
Cao, Y, Zhu, X, Ju, X et al. Optimization of dose distributions of target volumes and organs at risk during stereotactic body radiation therapy for pancreatic cancer with dose-limiting auto-shells. Radiat Oncol 2018; 13 (1): 11.CrossRefGoogle ScholarPubMed
Yamashita, H, Nakagawa, K, Nakamura, N et al. Exceptionally high incidence of symptomatic grade 2-5 radiation pneumonitis after stereotactic radiation therapy for lung tumors. Radiat Oncol 2007; 2: 21.CrossRefGoogle ScholarPubMed
Guckenberger, M, Baier, K, Polat, B et al. Dose-Response relationship for radiation-induced pneumonitis after pulmonary stereotactic body radiotherapy. Radiother Oncol 2010; 97 (1): 6570.CrossRefGoogle ScholarPubMed
Ong, C L, Palma, D, Verbakel, W F, Slotman, B J, Senan, S. Treatment of large stage I-II lung tumors using stereotactic body radiotherapy (SBRT): planning considerations and early toxicity. Radiother Oncol 2010; 97 (3): 431436.CrossRefGoogle ScholarPubMed
Barriger, R B, Forquer, J A, Brabham, J G, et al. A dose-volume analysis of radiation pneumonitis in non-small cell lung cancer patients treated with stereotactic body radiation therapy. Int J Radiat Oncol Biol Phys 2012; 82 (1): 457462.CrossRefGoogle ScholarPubMed
Stauder, M C, Macdonald, O K, Olivier, K R et al. Early pulmonary toxicity following lung stereotactic body radiation therapy delivered in consecutive daily fractions. Radiother Oncol 2011; 99 (2): 166171.CrossRefGoogle ScholarPubMed
Ding, C, Chang, C H, Haslam, J, Timmerman, R, Solberg, T. A dosimetric comparison of stereotactic body radiation therapy techniques for lung cancer: robotic versus conventional linac-based systems. J Appl Clin Med Phys 2010; 11 (3): 3223.CrossRefGoogle ScholarPubMed
Chan, M K H, Kwong, D L W, Law, G M L et al. Dosimetric evaluation of four-dimensional dose distributions of CyberKnife and volumetric-modulated arc radiotherapy in stereotactic body lung radiotherapy. J Appl Clin Med Phys 2013;14 (4): 4229.CrossRefGoogle ScholarPubMed
Descovich, M, Ma, L, Chuang, C F, Larson, D A, Barani, I J. Comparison between prone and supine patient setup for spine stereotactic body radiosurgery. Technol Cancer Res Treat 2012; 11 (3): 229236.CrossRefGoogle ScholarPubMed
Fürweger, C, Drexler, C, Muacevic, A, Wowra, B, de Klerck, E C, Hoogeman, M S. CyberKnife robotic spinal radiosurgery in prone position: dosimetric advantage due to posterior radiation access?. J Appl Clin Med Phys 2014; 15 (4): 4427.CrossRefGoogle ScholarPubMed
Nakayama, M, Nishimura, H, Mayahara, H et al. Clinical log data analysis for assessing the accuracy of the CyberKnife fiducial-free lung tumor tracking system. Pract Radiat Oncol 2018; 8 (2): e63e70.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1. CT image of the phantom in this study. (a) Coronal view of the CT image of the phantom that is divided into three regions: upper, middle and lower. (b) Axial view of the CT image of the phantom according to each region, which is further divided into 18 slices (nos. 1 to 18).

Figure 1

Table 1. Dosimetric parameter in supine and prone position

Figure 2

Table 2. Maximum dose and D95 dose in supine and prone position

Figure 3

Table 3. Dose volume metrics of the factors related to symptomatic RP

Figure 4

Figure 2. Dose distribution of the typical plans for different set-ups when (a, b) the target is located in the anterior or posterior region. (c, d) Dose distribution of the plans for the supine set-up when the target is located in the anterior or posterior region. (e, f) Dose distribution of the plans in prone set-up when the target is located in the anterior or posterior region.

Figure 5

Table 4. Quality factors of each plan in supine and prone positions

Figure 6

Figure 3. DVH curve of a typical plan for supine and prone set-ups, respectively. DVH curve of the target (a, b), and DVH curve of the lung (c, d) when the target is located in the anterior or posterior region. The solid line indicates the supine set-up plan; the dotted line indicates the prone set-up plan.