INTRODUCTION
Most commercial treatment planning systems (TPS) incorporate electron beam planning programs. However, not all programmes have desirable accuracy. Pencil beam algorithms are common for electron beam dose calculations. In these techniques the energy spread or dose kernel at a point is summed along a line in a phantom to obtain a pencil type beam. By integrating the pencil beam over the patient’s surface to account for changes in primary intensity and by modifying the shape of the pencil beam with depth and tissue density, a dose distribution can be generated.Reference Podgorsak 1 , Reference Hogstrom, Starkschall and Shiu 2 The most accurate way to calculate electron beam dose distributions is through Monte Carlo techniques.Reference Strydom, Parker and Olıvares 3
Varian (Varian Medical Systems, Palo Alto, CA, USA) marketed a partial Monte Carlo solution under the trade name electron Monte Carlo (eMC), for electron beam dose calculations. The fast eMC technique employed by Eclipse (Varian Medical Systems) is based on standard EGS4 Monte Carlo methods, but reduces the number of electron transport steps through the patient.Reference Janssen, Korevaar, van Battum, Storchi and Huizenga 4 , Reference Fix, Frei, Volken, Neuenschwander, Born and Manser 5
The goal of this study is to experimentally validate the accuracy of the eMC algorithm in the peripheral dose region. The percentage peripheral dose distributions of electron beams were measured using parallel plate type ionisation chamber and film dosimetry and the accuracy of dose calculation algorithms of Eclipse, General Gaussian Pencil Beam (GGPB) and the eMC, for calculating out-of-field doses was investigated.
METHODS
Ionisation chamber measurements
The percentage peripheral dose distributions of electron beams were measured for Varian Triology linear accelerator having 6, 9, 12, 15, 18, 20 and 22 MeV electron energies using a parallel plate ionisation chamber (PTW Roos 34001; PTW, Friedberg, Germany) inserted into a 40×15×120 cm3 (width×height×length) water-equivalent plastic phantom (RW3 Slab phantom; PTW). Great care was taken to ensure that there was no air gap while aligning the slabs.
The linear accelerator output was checked and monitored on a daily basis before each set of measurements. The outputs of the beam were measured with Roos ionisation chamber connected with Unidos Webline electrometer in RW3 solid water phantom. Absorbed dose measurements were done at z ref=0·6R50–0·1, SSD=100 cm using 10×10 cm2 cone for each energy. TRS 398 protocol was used to calculate absorbed doses. The absorbed dose at z ref was converted to absorbed dose at z max using PDD data. 1 cy=1 MU at d max of each energy.
Dose distributions were measured for standard 6×6, 10×10 and 25×25 cm2 cones at 100 cm Source to Surface Distance (SSD) by giving 100 monitor unit (MU) at central axis maximum depth (d max) of each energy. The applicator sizes refer to the dimension of the projected light field at SSD=100 cm.
The measurements were done at d max of each energy which are listed in Table 1.
Table 1 The dose maximum depth of each energy used in this study.

The percentage peripheral doses were measured up to 20 cm distances from the geometric field edge. Doses were measured with the increments of 0·5 cm. The collimator and gantry angle was 0°. Each measurement was repeated three times and the mean value of the readings were noted. The standard error was found to be within 1%. All the data were normalised to central axis at their respective d max to calculate percentage peripheral dose. These measured doses were then compared with values taken from the TPS and EBT3 film measurement.
Film calibration
GafChromic EBT3 (International Specialty Products, NJ, USA) is a new model of radiochromic film. As Sipila et al.Reference Sipilä, Ojala, Kaijaluoto, Jokelainen and Kosunen 6 found that the energy dependence of the EBT3 film is uniform within 0·5%, with uncertainties close to 1·6% (k=2) with electron beams ranging from 6 to 16 MeV and at reference measurement conditions in water. The GafChromic EBT3 films were used for the measurement of 2D dose distribution.
For calibration of film, one sheet of film was cut to 24 pieces with a size of 2×2 cm. The film pieces were divided into eight groups (three pieces in each group) for calibration. The groups were irradiated with dose steps of 2, 5, 10, 25, 50, 75, 100 and 125 cGy. The calibration was performed for 10×10 cm2 cone and 12 MeV electron energy at a depth of d max in 30×30×20 cm3 RW3 solid water phantom. After irradiation, the films were left for a period of 48 hours for self-development and scanned with a EPSON XL 1000 flatbed scanner. SNC Patient software version 5 of Sun Nuclear (Melbourne, FL, USA) was used for film analysis. The resulted calibration curve of net optical density (netOD) versus dose was obtained.
Comparison of Roos chamber measured profile with film measurement
In order to see the agreement between film and ionisation chamber, in field cross beam profiles were measured in water phantom using Roos chamber for all electron energies and 10×10 cm2 cone. These profiles were compared with the profiles measured in slab phantom using EBT3 film. The cross plane profiles measured both Roos ionisation chamber and EBT3 film were normalised to central axis dose in order to get rid of the effects of different measurement enviroment (water and solid water, respectively). To provide comparisons of multidimensional dose distributions, dose comparison tools such as gamma dose distribution, distance-to-agreement, and dose difference have been developed.Reference Low and Dempsey 7 , Reference Ezzell, Burmeister and Dogan 8 Gamma index analysis with the test criteria of ±3%/±3 mm were used in this study. A gamma value of >1 would be assigned for points failing the gamma criterion. To reduce statistical uncertainty, the measurement for each plan was repeated three times.
Film measurement
The EBT3 films were horizontally positioned and the measurements were performed at d max of each energy as a function of off-axis distances ranging from field edge to 20 cm; 100 MU was given at each irradiation. The film pieces with dimensions of 2×22 cm (from the same batch as calibration films) were utilised to measure peripheral dose distribution of electron beams. As the regions near the cutted edge would produce inaccurate results for the gamma analysis, the film pieces were in 22 cm length. Peripheral dose measurements were done for all energies and cone sizes used in ionisation chamber measurements. The red channel was used to convert the nOD of these films to dose. The region of interest for gamma analysis was chosen as a 2×20 cm. Then, the dose distribution received by the film and calculated by TPS using two algorithms were compared using gamma index analysis. The doses measured at field edge and 15 cm from field edge were compared with the doses measured by ionisation chamber at the same distances. The measurement for each plan was repeated three times.
External beam treatment planning calculations
The 40×15×120 cm3 water-equivalent plastic phantom was scanned with and without ionisation chamber by a computerised tomography (Toshiba Aquilion; Toshiba Medical Systems, Japan) to obtain three dimensional (3D) image data sets of 1 mm slices and transferred to the TPS. The active volume of ionisation chamber was contoured manually on the CT to calculate the mean dose on this contour.
The electron beams were created in Eclipse for each cone size and energy with the collimator and gantry orientation 0° imitating the measurements under real conditions in solid water phantom. We calculated the prescribed doses to give 100 MU to the d max at the central axis for each energy and cone size. All plans were initially calculated with a GGPB algorithm. Plans were then recomputed (keeping everything same) within Eclipse using eMC. The plans were normalised to 100% at their respective d max for analysis.
The eMC has six user-selectable parameters for individual calculations: calculation grid size, accuracy, maximum number of particle histories, random number generator seed, smoothing method and smoothing level. To attain accurate calculations and consistency within a reasonable amount of time, the eMC calculation parameters used in this study are listed in Table 2. Based on research done by Popple et al.Reference Popple, Weinber and Antolak 9 to achieve clinically acceptable results, grid sizes for eMC calculations should be varied based on energy. The grid size should be approximately one-tenth of the distal fall off distance of the electron depth dose curve.
Table 2 Eclipse electron Monte Carlo calculation parameters used in this study.

For the ionisation chamber comparison, the percentage peripheral doses at varying off-axis points were recorded from the TPS by transferring contour which represent the active volume of ionisation chamber to measurement point using ‘move structure’ option of the Eclipse and calculating the mean doses at this volume. The dose measured at every location was compared with the given dose as generated by the TPS for two different algorithms at corresponding volumes. The dose difference % was calculated as follows:

For the EBT3 film measurement comparison, plans were made in the same way as described above on the CT scan of slab phantoms without chamber. The dose profiles calculated using two different algorithm were transferred to SNC patient software in order to compare them with EBT3 film measurements.
RESULTS
Comparison of calculated data with ionisation chamber measurements
Figure 1 shows the Roos chamber measured percentage peripheral dose distributions of 6, 12 and 22 MeV electron beams for 10×10 cm2 cone. It is seen from the figure that percentage peripheral dose increases with the increase in energy. The maximum percentage peripheral dose was 83·92±3·2% for 22 MeV and it decreased to 68·54±3·8% for 6 MeV at 25×25 cm2 cone size. For the Varian Linac, the out-of-field dose decreases continuously with increasing distance. These doses, at 15 cm from the beam central axis, are 0·7, 0·8, 0·9, 0·9, 1·2 and 1·4% of d max for 6, 9, 12, 15, 18 and 22 MeV at 10×10 cm2 cone, respectively.

Figure 1 The Roos chamber measured percentage peripheral dose distributions of 6, 12 and 22 MeV electron beams for 10×10 cm2 cone.
Figures 2 and 3 shows the Roos chamber measured percentage peripheral dose distributions of 6 and 22 MeV electron beams for 6×6, 10×10 and 25×25 cm2 cone sizes. According to the figures, percentage peripheral dose increases with the increase in cone size. The ratio of percentage peripheral doses for 10×10 to 6×6 cm2 cones sizes was 1·02 and it was 1·18 for 25×25 to 10×10 cm2 cone sizes for 6 MeV at the field edge. The increase in percentage peripheral dose with cone size increases with electron beam energy. The ratio of percentage peripheral doses for 10×10 to 6×6 cm2 cones sizes was 1·08 and it was 1·25 for 25×25 to 10×10 cm2 cone sizes for 22 MeV at the field edge.

Figure 2 The Roos chamber measured percentage peripheral dose distributions of 6 MeV electron beams for 6×6, 10×10 and 25×25 cm2 cone sizes.

Figure 3 The Roos chamber measured percentage peripheral dose distributions of 22 MeV electron beams for 6×6, 10×10 and 25×25 cm2 cone sizes.
Figures 4a, 4b, 5a, 5b, 6a and 6b show the percentage difference graphs of 6, 9, 12, 15, 18 and 22 MeV electron beams at three different cone sizes, respectively. It is seen from the figures that the biggest difference between measurement and calculation was observed at the field edge. And as the distance from the field edge increased the difference between measurement and calculation decreased. The TPS calculated peripheral dose underestimated the measured ones.

Figure 4 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 6×6 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 6×6 cm2 cone.

Figure 5 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 10×10 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 10×10 cm2 cone.

Figure 6 (a) The percentage difference between ionisation chamber measured and GGPB calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 25×25 cm2 cone. (b) The percentage difference between ionisation chamber measured and eMC calculated data of 6, 9, 12, 15, 18 and 22 MeV electron beams for 25×25 cm2 cone.
The maximum difference between GGPB calculated peripheral dose and measurement was 8·72% for 22 MeV and 25×25 cm2 cone. The difference between GGPB calculated and measured percentage peripheral dose decreases with increasing energy up to 3 cm away from field edge. However, as the distance from field edge increases, the difference remains constant. The minimum difference was 5·01% for 6 MeV and 6×6 cm2 cone.
The difference between measurement and GGPB results increases with field size. The maximum difference between measurement and GGPB results for 22 MeV was 7·51, 8·21 and 8·72% for 6×6, 10×10 and 25×25 cm2 cones, respectively. And the minimum difference between measurement and GGPB results for 6 MeV was 5·01, 5·51 and 6·02% for 6×6, 10×10 and 25×25 cm2 cones, respectively.
The results of eMC algorithm were much closer to measurement results than GGPB algorithm. The maximum difference between eMC and measurement results was 4·29% for 22 MeV and 25×25 cm2 cone. Except for 22 MeV, the difference between eMC and measurement results decreases with increase in energy. The minimum difference was 2·79% for 18 MeV and 6×6 cm2 cone.
The difference between measurement and eMC results increases with field size. The maximum difference between measurement and eMC results for 22 MeV was 3·59, 4·19 and 4·29% for 6×6, 10×10 and 25×25 cm2 cone sizes, respectively. And the minimum difference between measurement and eMC results for 18 MeV was 2·79, 2·89 and 3·0% for 6×6, 10×10 and 25×25 cm2 cone sizes, respectively.
Comparison of Roos chamber measured profile with film measurement
The cross beam profiles measured using Roos ionisation chamber and EBT3 film for 10×10 cm2 cone size were in agreement between 2·4%. The region of interest was chosen 8×8 cm2 in order not to take into account field edges. The gamma index passing rates for ±3%/3 mm criteria were given in Table 3. The EBT3 film measurements overestimated the dose in comparison with ionisation chamber measurements.
Table 3 The gamma index passing rates of in field cross plane profiles measured by ionisation chamber and film dosimetry for ±3%/3 mm criteria

Comparison of calculated data with EBT3 film measurements
Percentage peripheral dose distributions were measured by EBT3 films by the red channel calibration curve for 6, 9, 12, 15, 18 and 22 MeV electron beams using 6×6 cm2, 10×10 cm2and 25×25 cm2 cone sizes.
Table 4 shows a comparison of ionisation chamber measurements with film measurements, at two different off-axis distances. The EBT3 film measurements overestimated the dose in comparison with ionisation chamber measurements for all points. The measurement discrepancies between ionisation chamber and the film is about 7–8% at field edge, and it reaches up to 20% at 15 cm off-axis distance.
Table 4 Percentage dose difference (film measurement−ionisation chamber measurement/film measurement) between out-of-field doses measured with EBT3 film and Roos ionisation chamber at 10×10 cone

Gamma index passing rates for EBT3 measured and GGPB and eMC calculated doses for all energies and cone sizes are shown in Tables 5 and 6, respectively.
Table 5 Gamma index passing rates for EBT3 film measurement and GGPB calculation for ±3%/3 gamma index passing criteria

Table 6 Gamma index passing rates for EBT3 film measurement and eMC calculation for ±3%/3 gamma index passing criteria

There is no clinical acceptable agreement found between EBT3 measured and GGPB calculated percentage peripheral doses. For all energies and cone sizes analysed the gamma index passing rates were below 90%.
The agreement between measured and eMC calculated values was clinically acceptable for the 12, 15, 18 MeV with >97% of pixels passing our gamma requirements for all cone sizes tested. The 6 and 22 MeV showed poor results; the numbers of pixels passing our gamma requirements were below 95% for all cone sizes used.
The difference between ionisation chamber and film measurements can be arisen from the fact that spatial resolution of Roos chamber is lower than EBT3 film.
Uncertainity analysis of the ionisation chamber and EBT3 film measurements were given at Tables 7 and 8, respectively. Source of the uncertainty for intrinsic characteristics of EBT3 film measurement for red channel were taken from the study León et al.Reference León, Herrera, Camacho, Villarreal and García-Garduño 10
Table 7 Source of the uncertainty for ionisation chamber measurement

Table 8 Source of the uncertainty for intrinsic characteristics of EBT3 film measurement

DISCUSSION
The peripheral dose is the radiation dose received at points beyond the collimated radiotherapy field edge. The peripheral dose has two components; photons and electrons.Reference Chow and Grigorov 11 – Reference Shimozato, Okudaira, Fuse and Tabushi 13 The first component, Bremsstrahlung photons, occurs when relatively high energy electron beams interact with the applicator.Reference Zhu, Das and Bjärngard 14 The second component, scattered electrons, occur when electron beams penetrate with applicator and direct leakage into air from scatter foils during interaction with the applicator. In high energy electron beams, the direct escape to air mainly occurs during penetration and/or activation with the applicator. In lower energy electron beams scattered electrons occur.Reference Iktueren, Bilge, Karacam and Atkovar 15
TPSs are not commissioned for out-of-field dose calculations and the accuracy of TPS dose calculations is known to decrease beyond the field edge. While TPSs work well to calculate treatment dose in high-dose regions, they lack the ability to accurately model the doses to the periphery well beyond the extent of the field. Given the potential consequences of under or overestimation of dose it would be useful to document the accuracy of peripheral dose for specific TPSs.Reference Howell, Scarboro, Kry and Yaldo 16 Therefore, the objective of this work was to quantify the accuracy of out-of-field dose for a commonly used TPS, Eclipse version 13.0.28. The biggest difference between measurement and calculation was observed at the field edge and as the distance from the field edge increased the difference between measurement and calculation decreased. The difference between measured and calculated percentage peripheral dose increases as the cone size increases and except for 22 MeV it decreases as the energy of the beam increases. The TPS calculated peripheral dose underestimated the measured ones. In all cases, it is easy to note that the eMC models the dose distribution more accurately than the GGPB does. The TPS underestimated the out-of-field doses. The differences between measured and calculated doses increase with the cone size. The largest deviation between calculated and measured dose distribution is <4·29% using the eMC, but can increase up to 8·72% using GGPB. eMC algorithms models medium energies best. The agreement is better for 18 MeV and small field sizes. However, for high energy and big field sizes the disagreement can reach 4·29%. One reason could be that in the current implementation of eMC algorithm in Eclipse, the EGSnrc precalculations for the MMC database have been carried out for monoenergetic electrons in certain steps which might not be fine enough.
Alabdoaburas et al.Reference Alabdoaburas, Mege and Chavaudra 17 measured the peripheral dose of electron beams by thermoluminescent dosimeter and EBT3 film for a Varian-type applicator. They found that the out-of-field dose from electron beams increases with the beam energy and the applicator size, and decreases with the distance from the beam central axis and the depth in water. The results of our work is in agreement with this work.
Xu et al.Reference Xu, Walsh, Telivala, Meek and Yang 18 compared eMC calculations and measurements of depth doses, isodose distributions and monitor units for several different energy and small field cutout size combinations at different SSDs. They measured using EBT film and a PinPoint ion chamber. They found that the eMC algorithm can accurately predict depth doses, isodose distributions, and monitor units (within 2·5%) for field sizes as small as 3·0 cm diameter for energies in the 6–20 MeV range at 100 cm SSD. However, at extended SSDs (105–110 cm), their results show good agreement (within 4%) only for higher energies (12, 16 and 20 MeV) for a field size of 3 cm.
Zhang et al.Reference Zhang, Wen, Nurushev, Burmeister and Chetty 19 investigated dose calculation accuracy between pencil beam (PB) and eMC algorithms in the same treatment planning system for heterogeneous phantoms. They found that the eMC calculations generally agreed with the measurements to within 3% or 2 mm in both homogeneous and heterogeneous phantoms and large difference was observed for 20 MeV. However, the PB algorithm cannot predict the sharp dose gradients adjacent to the inhomogeneity and results in large errors (up to 25%) beneath the inhomogeneities such as bone and lung.
CONCLUSION
Our results show that the Monte Carlo algorithm for electron planning in Eclipse is more accurate than previous algorithms for peripheral dose distributions. It must be emphasised that the use of GGPB for planning large field treatments with 6 MeV could lead to inaccuracies of clinical significance. Even eMC algorithm underestimates the measured values. We concluded that peripheral dose data from TPS should only be used with a clear understanding of the accuracy of dose calculations outside the treatment field.