Hostname: page-component-745bb68f8f-v2bm5 Total loading time: 0 Render date: 2025-02-11T08:36:30.510Z Has data issue: false hasContentIssue false

Calypso’s array attenuation

Published online by Cambridge University Press:  05 March 2015

Célia Silva*
Affiliation:
Centro Clínico Fundação Champalimaud, Lisboa, Portugal Escola Superior de Tecnologias da Saúde de Lisboa, Portugal
Dalila Mateus
Affiliation:
Centro Clínico Fundação Champalimaud, Lisboa, Portugal
Sandra Vieira
Affiliation:
Centro Clínico Fundação Champalimaud, Lisboa, Portugal
Milton Rodrigues
Affiliation:
Centro Clínico Fundação Champalimaud, Lisboa, Portugal
Margarida Eiras
Affiliation:
Escola Superior de Tecnologias da Saúde de Lisboa, Portugal
Carlo Greco
Affiliation:
Centro Clínico Fundação Champalimaud, Lisboa, Portugal
*
Correspondence to: Célia Silva, R. Outeiro Cacho, 6, Loureira, 2495-161 Sta Cat Serra, Portugal. Tel: 00351919628745. E-mail: celia.psilva@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Introduction

The Calypso 4D Localization System gives the possibility to track the tumour during treatment, with no additional ionising radiation delivered. To monitor the patient continuously an array is positioned above the patient during the treatment. We intend to study, for various gantry angles, the attenuation effect of the array for 6- and 10 MV and flattening filter free (FFF) 6- and FFF 10 MV photon beams.

Materials and methods

Measurements were performed using an ion chamber placed in a slab phantom positioned at the linac isocenter for 6 MV, 10 MV, FFF 6 MV and FFF 10 MV photon beams. Measurements were performed with and without array above the phantom for 0°, 10°, 20°, 40° and 50° beam angle for a True Beam STx linac, for 5×5 and 10×10 and 15×15 cm2 field size beams to evaluate the attenuation of the array. A VMAT treatment plan was measured using an ArcCheck with and without the array in the beam path.

Results and discussion

Attenuation measured values were up to 3%. Attenuation values were between 1 and 2% with the exception of the 30°–50° gantry angles which were up to 3.3%. The ratio values calculated in the ArcCheck for relative dose and absolute dose 10 were both 1·00.

Conclusion

Attenuation of the treatment beam by the Calypso array is within acceptable limits.

Type
Original Articles
Copyright
© Cambridge University Press 2015 

Introduction

Higher accuracy and reproducibility in radiotherapy has led to great development in imaging and monitoring systems. Megavoltage imaging has been used clinically for many years, and kV imagers have also been installed in linacs all over the world. Monitoring systems for tracking movement during treatment have been used to monitor patient surface – for example infrared tracking of external markers or virtual view of the patient surface – or the tumour movement – for example fluoroscopy.Reference Schweikard, Shiomi and Adler1Reference Stieler, Wenz, Shi and Lohr11

The Calypso 4D Localization System (Varian Medical Systems, Palo Alto, CA, USA) is a monitoring system that gives the possibility to track the tumour during treatment, with no additional ionising radiation delivered, a great advantage when compared with other systems available.Reference Litzenberg, Willoughby and Balter12

This system has five components: Beacon transponders, the console, the array, the optical localisation subsystem and the monitoring station. The array consists of optical targets, 4 sources and 32 receiver coils.Reference Quigley, Mate and Sylvester13 An oscillating signal (25 Hz) through the source coil generates resonance in the transponders. When this signal is turned off, the transponders emit electromagnetic signals, which are detected by the receiver coils in the array, thereby localising their positions relative to the array. Meanwhile, the in-room infrared camera system tracks the array relative to the isocenter.Reference Santanam, Malinowski and Hubenshmidt14

To monitor the patient continuously an array is used. This array is positioned above the patient during the treatment.Reference Quigley, Mate and Sylvester13, Reference Li, Chetty and Enke15

Although the array lies between the patient and the beam, it is not included in the dose calculation of the treatment planning system.

Zou et al. studied the array attenuation effect for the regular energies 6 and 15 MV photon beams for various gantry angles – and concluded that the dose difference due to the placement of Calypso array was clinically insignificant to the treatment.Reference Zou, Betancourt, Yin, Metz, Avery and Kassaee16 In our institute the Calypso system is mainly used in the irradiation of flattening filter free (FFF) beams. Given that the removal of the flattening filter lowers the mean energy of the beam we propose to study, for various gantry angles, the attenuation effect of the array for FFF 6 and FFF 10 MV photon beams. It is also to be noticed that point measurements of an inhomogeneous array may lead to uncertainties, as it contains source coils, sensors and infrared targets. Considering that the Calypso system has been used mainly in prostate treatments, a QA of a prostate VMAT treatment plan was performed with and without the Calypso array in the beam.

MATERIALS AND METHODS

All transmission measurements were performed on a True Beam STx linear accelerator (Varian Medical Systems) using a CC13 ionisation chamber of 0·13 cm3 of sensitive volume (IBA Dosimetry, Germany) connected to a Dose 1 electrometer (IBA Dosimetry, Germany). Corrections for temperature and pressure were applied.

Attenuation effect of the array for 6- and 10 MV and FFF 6- and FFF 10 MV photon beams

The ionisation chamber was inserted in a slab phantom and positioned in the isocenter at 5 cm depth. The array was positioned above the phantom in the (0, 0, 0) position indicated by Calypso software system, in the same way it is positioned above the patient during treatment (see Figure 1).

Figure 1 Gantry angle measurements acquisition scheme.

Measurements were performed for regular 6- and 10 MV and FFF 6- and FFF 10 MV energies, for 5×5 and 10×10 and 15×15 cm2 square field sizes. The readings were obtained in six different gantry angles: 0°, 10°, 20°, 30°, 40° and 50°. Further gantry angles were not considered in this study as the centre of the beam would not traverse the array. For each measurement, 200 monitor units (MU) were delivered at a dose rate of 600 MU/minute for regular beam energies and 800 MU/minute for FFF energies.

Measures were performed with and without the array in the beam path. Each measurement was repeated five times. The transmission measurements were registered in a table. The attenuation was calculated according to the formula:

$$\eqalign{{\rm Attenuation (\,\%\,}) \cr =\left( {1{\minus}{{{\rm Measurement \ with \ array}} \over {{\rm Measurement \ without \ array}}}} \right){\times}100$$

The attenuation calculated values were registered and analysed. Mean and standard deviation were calculated.

A fit was done to the attenuation curves to evaluate the goodness of the fit.

Attenuation effect of the array for a FFF 10 MV prostate treatment

An ArcCheck system (SunNuclear, Melbourne, FL, USA) was used to measure the dose delivered and to compare with the expected dose.

A computed tomography of the ArcCheck was imported in our Treatment Planning System (TPS) Eclipse (Varian Medical Systems). A prostate patient previously treated was selected randomly. The treatment consisted of a VMAT plan for a 28×2·25 Gy irradiation of prostate and seminal vesicles. This plan was recalculated in the ArcCheck CT.

The ArcCheck was positioned in the isocenter, and the array was positioned above the phantom in the (0, 0, 0) position indicated by Calypso software system, as performed during patient treatment.

Repeated ArcCheck measurements of the patient treatment were performed with and without the array in the beam path. A standard deviation was calculated for both conditions.

All measured maps were compared to the dose map calculated. Comparison was evaluated for 3% dose for 3 mm for relative dose (RD) and absolute dose (AD) for all measured points except those with 10% dose or less, according to the local protocol.

The ratio percentage dose (between measures with and without the array) was calculated and analysed.

RESULTS AND DISCUSSION

Attenuation effect of the array for 6- and 10 MV and FFF 6- and FFF 10 MV photon beams

The attenuation values measured were higher for 5×5 cm2 fields than for 10×10 cm2 fields for all energies and for the same measurement conditions. Also, the attenuation values measured were higher for 10×10 cm2 fields than for 15×15 cm2 fields for all energies and for the same measurement conditions. Therefore, the data shows that the beam attenuation is field size dependent. This dependency was not calculated. Field size dependency has been previously reported in other devices attenuation studies, although this dependency was also not quantifiable in those reports.Reference Myint, Niedbala, Wilkins and Gerig17Reference Seppälä and Kulmala19

These studies usually also report an angular dependence on the attenuation of the beam by devices. A second degree polynomial fit was applied to the attenuation curves. For the 5×5 cm2 field size curves, the r 2 value for 6 MV, 10 MV, FFF 6 MV and FFF 10 MV of 0·98; 0·97; 0·97 and 0·97, respectively. For 10×10 cm2 field size curves, the r 2 value was 0·96; 0·98; 0·98 and 0·99, for the same energies, respectively. For 15×15 cm2 field size curves, the r 2 value was 0·97; 0·96; 0·98 and 0·96, for the same energies, respectively. Therefore, there is a tendency for higher attenuation values as the gantry angle increases, as it is shown in Figures 2, 3 and 4.

Figure 2 Attenuation by Calypso array of 5×5 cm2 field size beams.

Figure 3 Attenuation by Calypso array of 10×10 cm2 field size beams.

Figure 4 Attenuation by Calypso array of 15×15 cm2 field size beams.

All points measured showed 0·0 or 0·1% standard deviation. Measurements can be considered precise.

The array attenuation calculated values are comparable to attenuation values presented previously. Zou et al. reported that the attenuation on the array was about 2–3% for both 6 and 15 MV energies, for 1×1 cm2 field size beams at gantry angles between 0° and 40°. The calculated attenuation slowly increased above these values for angles around 50°–60°.Reference Zou, Betancourt, Yin, Metz, Avery and Kassaee16

Here the calculated attenuation values were between 1% and 2% for gantry angles 0°, 10° and 20°, for both field sizes for all energy beams. Acquisitions at 30°, 40° and 50° gantry angles showed higher attenuation values. The higher attenuation calculated value was 3·4% for a 5×5 cm2 field for a FFF 6 MV beam (gantry angle: 50°), and 3·3% for a 10×10 cm2 field and 3·1% for a 15×15 cm2, both for the same energy beam, at the same gantry angle.

Array attenuation values can be considered acceptable. Nevertheless it is to be noticed that point measurements were performed and because the FFF energy beams are not flat by definition, positioning accuracy of the ionisation chamber can be challenging. Furthermore, the array is also inhomogeneous, as it contains source coils, sensors and infrared targets as it is shown in Figure 5. A 2D Electronic Portal Imaging Device detector could be used to assess that, however it has to be compatible with the use of FFF beams.

Figure 5 Portal image of Calypso’s array.

Attenuation effect of the array for a FFF 10 MV prostate treatment

The repeated measurements without array above the ArcCheck showed less than 0·4% standard deviation for both RD and AD10 evaluations. The repeated measurements with the array above the ArcCheck showed less than 0·5% standard deviation for the same evaluations.

The ratio values calculated for RD were all 1·00. For the AD there was one ratio value of 1·01. The mean values for the two evaluation methods were all 1·00. Therefore the presence of the array in the beam path is negligible.

Conclusion

The behaviour of the array attenuation curves is important to study due to its inhomogeneous structure.

Dose attenuations were measured to be within 1–2% with the exception of the 30°–50° gantry angles which were up to 3·4%. The results indicate that the dose attenuation of the Calypso array may be within acceptable limits.

Future work should assess the Calypso attenuation of radiotherapy treatment beams with more detail.

Acknowledgements

The authors thank the staff of the Radiotherapy Department at Centro Clinico Champalimaud for their personal encouragement and Dr Andrew Macann for helping with English language revision. This review was proposed in the master course Radiations applied to Health Technologies – Radiation Therapy, at Escola Superior de Tecnologias da Saúde de Lisboa. Revision of the review was kindly made by Sandra Vieira and Dalila Mateus, who have worked with Calypso, and also by Prof. Margarida Eiras.

Financial support

No monetary costs were involved in this review. Any cost related to publication is supported by the primary author herself.

Conflicts of Interest

None.

References

1.Schweikard, A, Shiomi, H, Adler, J. Respiration tracking in radiosurgery. Med Phys 2004; 31: 27382741.Google Scholar
2.Meeks, S L, Bova, F J, Wagner, T Het al. Image localization for frameless stereotactic radiotherapy. Int J Radiat Oncol Biol Phys 2000; 46: 12911299.Google Scholar
3.Bova, F J, Meeks, S L, Friedman, W Aet al. Optic-guided stereotactic radiotherapy. Med Dosim 1998; 23: 221228.Google Scholar
4.Roberts, D W, Strohbein, J W, Hatch, J F. A frameless stereotactic computerized tomographic imaging and the operating microscope. J Neurosurg 1986; 65: 545549.CrossRefGoogle Scholar
5.Watanabe, E, Mayanagi, Y, Kosugi, Yet al. Open surgery assisted by the articulated, sensitive arm. Neurosurgery 1991; 28: 792800.CrossRefGoogle ScholarPubMed
6.Suess, O, Suess, S, Mularksi, Set al. Study on the clinical application of pulsed DC magnetic technology for tracking of intraoperative head motion during frameless stereotaxy. Head Face Med 2006; 2: 10.CrossRefGoogle Scholar
7.Keall, P J, Todor, A D, Vedam, S Set al. On the use of EPID-based implanted marker tracking for 4D radiotherapy. Med Phys 2004; 31: 34923499.Google Scholar
8.Meeks, S L, Buatti, J M, Bouchet, L Get al. Ultrasound-guided extracranial radiosurgery: technique and application. Int J Radiat Oncol Biol Phys 2003; 55: 10921101.Google Scholar
9.Tome, W A, Meeks, S L, Orton, N Pet al. Commissioning and quality assurance of an optically guided three-dimensional ultrasound target localization system for radiotherapy. Med Phys 2002; 29: 17811788.Google Scholar
10.Sharp, G C, Jiang, S B, Shimizu, Set al. Tracking errors in a prototype real-time tumor tracking system. Phys Med Biol 2004; 49: 53475356.CrossRefGoogle Scholar
11.Stieler, F, Wenz, F, Shi, M, Lohr, F. A novel surface imaging system for patient positioning and surveillance during radiotherapy: a phantom study and clinical evaluation. Strahlenther Onkol 2013; 189: 938944.Google Scholar
12.Litzenberg, D W, Willoughby, T R, Balter, J Met al. Positional stability of electromagnetic transponders used for prostate localization and continuous, real-time tracking system and on-board kilovoltage imaging system. Int J Radiat Oncol Biol Phys 2007; 68 (4): 11991206.Google Scholar
13.Quigley, M M, Mate, T P, Sylvester, J E. Prostate tumor alignment and continuous, real-time adaptive radiation therapy using electromagnetic fiducials: clinical and cost-utility analyses. Urol Oncol 2009; 27: 473482.Google Scholar
14.Santanam, L, Malinowski, K, Hubenshmidt, Jet al. Fiducial-based translational localization accuracy of electromagnetic tracking system and on-board kilovoltage imaging system. Int J Radiat Oncol Biol Phys 2008; 70 (3): 892899.CrossRefGoogle ScholarPubMed
15.Li, H S, Chetty, I J, Enke, C Het al. Dosimetric consequences of intrafraction prostate motion. Int J Radiat Oncol Biol Phys 2008; 71 (3): 801812.CrossRefGoogle ScholarPubMed
16.Zou, W, Betancourt, R, Yin, L, Metz, J, Avery, S, Kassaee, A. Effects on the photon beam from an electromagnetic array used for patient localization and tumor tracking. J Appl Clin Med Phys 2013; 14 (3): 7280.Google Scholar
17.Myint, K, Niedbala, M, Wilkins, D, Gerig, L H. Investigating treatment dose error due to beam attenuation by a carbon fiber tabletop. J Appl Clin Med Phys 2006; 7 (3): 2127.Google Scholar
18.Njeh, C F, Raines, T W, Saunders, M W. Determination of the photon beam attenuation by the BrainLAB imaging couch: angular and field size dependence. J Appl Clin Med Phys 2009; 10 (3): 1627.Google Scholar
19.Seppälä, J K H, Kulmala, J A J. Increased beam attenuation and surface dose by different couch inserts of treatment tables used in megavoltage radiotherapy. J Appl Clin Med Phys 2011; 12 (4): 1523.CrossRefGoogle ScholarPubMed
Figure 0

Figure 1 Gantry angle measurements acquisition scheme.

Figure 1

Figure 2 Attenuation by Calypso array of 5×5 cm2 field size beams.

Figure 2

Figure 3 Attenuation by Calypso array of 10×10 cm2 field size beams.

Figure 3

Figure 4 Attenuation by Calypso array of 15×15 cm2 field size beams.

Figure 4

Figure 5 Portal image of Calypso’s array.