An institutional review: dosimetry comparison between simultaneous integrated boost IMRT and VMAT for prostate cancer

Published online by Cambridge University Press:  09 June 2020

Raheel Mukhtar
Affiliation:
Department of Medical Physics, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
Sumera Butt
Affiliation:
Clinical and Radiation Oncology Department, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
Muhammad Abdur Rafaye
Affiliation:
Department of Medical Physics, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
Khalid Iqbal*
Affiliation:
Department of Medical Physics, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
Sana Mazhar
Affiliation:
Department of Medical Physics, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
Tabinda Sadaf
Affiliation:
Clinical and Radiation Oncology Department, ShaukatKhanum Memorial Cancer Hospital and Research Center, Lahore, Pakistan
*
Author for correspondence: Khalid Iqbal, Shaukat Khanum Memorial Cancer Hospital and Research Centre, 7A BLOCK R-3 MA, Johar Town Lahore, Lahore, Punjab54000, Pakistan. Tel: +00923006333815. E-mail: kiqbal@skm.org.pk
Rights & Permissions [Opens in a new window]

Abstract

Purpose:

A comparative study was performed about the plan parameters and quality indices between volumetric arc therapy (VMAT) and intensity-modulated radiotherapy (IMRT) for the treatment of high-risk prostate cancer patients. The aim of this retrospective study was to compare the two methods of external beam radiotherapy IMRT and VMAT in terms of plan quality and efficacy.

Material and method:

Fifteen high-risk prostate patients were planned for radiotherapy using 6 MV photon. Three dose levels were contoured having Planning Tumour Volume 1 (PTV1 = 48 Gy), Planning Tumour Volume 2 (PTV2 = 57.6 Gy) and Planning Tumour Volume 3 (PTV3 = 60 Gy). Setup margins were given using the CHIP trial method. The prescribed PTV3 dose was 60 Gy in 20 fractions which is biologically equivalent to 74 Gy in 37 fractions using α/β = 3. In case of IMRT, seven fixed beam angles 30, 60, 105, 180, 255, 300 and 330 were used and the dose was optimised using the sliding window method. In case of rapid arc technique, one or two full arcs were used for dose optimisation while keeping all the dose constraints and other planning parameters same used in IMRT. The plan evaluation parameters and Organ at risks (OARs) doses were calculated using a dose volume histogram (DVH).

Results:

The average D2, D5, D95 and PTVmean for PTV3 were 61.22, 61.13, 58.12, 60.00 Gy and 62.41 62.24 59.53 61.12 Gy for IMRT and VMAT, respectively. The averages V60 for bladder and V30 for rectum were 22.81, 25 and 67, 65% for IMRT and VMAT, respectively. The average homogeneity index (HI), conformity index (CI) and gradient index (GI) were 1.04, 1.4833, 14.79 and 1.04, 1.704, 7.89 for IMRT and VMAT, respectively.

Conclusion:

VMAT takes less dose-delivery time and lesser number of monitoring units than IMRT, thus it compensates the intrafractional movements during dose delivery. The Dose GI in VMAT was much better than IMRT. This indicates sharper dose fall off near the normal tissue. No other major differences were observed in terms of plan evaluation parameters between IMRT and VMAT techniques. So, we conclude that VMAT technique is more efficient than IMRT in terms of plan quality and dose delivery.

Type
Original Article
Copyright
© The Author(s), 2020. Published by Cambridge University Press

Introduction

Prostate cancer is the most common in males. Reference Riou, de la Mothe and Azria1 Radiotherapy is the commonly used treatment modality. Intensity-modulated radiotherapy (IMRT) is superior technique than 3D conformal radiotherapy due to its ability to generate non-homogenous dose flounce according to geometry of target volume. Since IMRT delivers the dose in fixed gantry positions with fixed dose rate, volumetric arc therapy (VMAT) is superior to IMRT because it can deliver the precise dose to target using the enhanced parameters such as gantry speed, variable dose rate and variable multi-leaf collimator position. Reference Gao, Li and Ning2,Reference Gautam3 VMAT offers similar or better organ at risk sparing compared to IMRT in less time and using less number of monitoring units. Reference Mellon, Javedan and Strom4

Simultaneous integrated boost (SIB) is a way to deliver selective doses to different planning tumour volume (PTV) in a single plan. Reference Franzese, Fogliata and D’Agostino5Reference Ishii, Ogino and Hosokawa7 This can be achieved using IMRT or VMAT. Reference Daoud, Saleh and Elsherbini8 SIB can be delivered to prostate patients without increasing the risk of acute toxicity. Reference Onal, Sonmez and Erbay9

The standard seven gantry angles for prostate IMRT planning technique are 30, 60, 105, 180, 255, 300 and 330. The standard VMAT plan for prostate uses one or two full arc rotations around the patient while the beam is on. Reference Mellon, Javedan and Strom4

VMAT offers superior dose homogeneity index (HI), dose gradient index (GI) as compared to IMRT. Reference Daoud, Saleh and Elsherbini8 A better GI shows sharp dose fall near normal tissue which indicates better Organ at risk (OAR) sparing. Although VMAT offers better dosimetric results Reference Johnson, Osei and Fleck10 but there is no final conclusion. There is much variation in results as it depends upon institutional workflow and practice. Reference Johnson, Osei and Fleck10

Bhoj Gotam has compared several research works in an article Reference Gautam3 about IMRT and VMAT techniques for prostate cancer patients. Tsai et al. Reference Tsai, Wu and Chao11 showed that the results of IMRT and VMAT were comparable with VMAT having little dosimetric advantage over IMRT. Zhang et al. Reference Zhang, Happersett and Hunt12 has described that dose-delivery time is reduced by up to 50–55% in VMAT as compared to IMRT. Ost et al. Reference Ost, Speleers and De Meerleer13 showed that rectum dose is less in VMAT plan. Yoo et al. Reference Yoo, Wu and Lee14 concluded after comparison of ten patient plans that IMRT offers better OAR sparing as compared to VMAT. Wolff et al. Reference Wolff, Stieler and Welzel15 showed that mean rectum dose was less in IMRT plans. The variation in results depends upon certain factors like treatment planning system data, plan optimisation parameters, techniques and heterogeneous structures like femoral heads in beam path and heterogeneity corrections applied in dose calculation algorithm which varies from institute to institute. Reference Gautam3 The variation in IMRT and VMAT results in different studies also depends upon linear accelerator commissioning data and linear accelerator model used in the study. Reference Gautam3 Onal et al. Reference Onal, Sonmez and Erbay9 declared that PTV homogeneity was better in IMRT than VMAT for all the PTV dose energy levels.

The purpose of this study is to have an institutional-based retrospective review between two planning techniques in terms of plan evaluation parameters and efficient dose delivery and also this work has been intended to provide methods to develop alternative tool to enhance the quality of the treatment plans.

Material and Method

Fifteen high-risk prostate patients were selected randomly who were treated with SIB–VMAT plans during 2018–19. Sixteen slice CT scanner (Model: Aquilion LB Toshiba Medical Systems) was used to scan the patients with scan slice thickness of 3 mm. For bladder and rectum, standard institutional protocols were followed which were empty rectum and comfortably full bladder. All patients were scanned in supine position with Vac-lock being used as immobilisation device.

PTV 48, PTV 57.6 and PTV 60 were delineated, and margins were assigned according to CHIP trial protocols. Reference David, Syndikus and Mossop16 Bladder, rectum, femoral heads and small bowel were contoured as organs at risk.

Total prescribed dose was 60 Gy in 20 fractions. It is biologically equivalent to 74 Gy in 33 fractions according to alpha beta ratio 3. Reference Lee17,Reference Myrehaug, Chan and Craig18

PTV contouring

PTV 48, PTV 57.6 and PTV 60 were contoured according to Conventional or hypo-fractionated high dose intensity modulated radiotherapy for prostate cancer (CHHiP) trial guidelines. Reference David, Syndikus and Mossop16

Treatment planning

IMRT and VMAT plans were optimised using newly developed photon optimiser (PO). The exact IGRT couch top medium was incorporated in the dose calculation grid. Dose calculations were done using anisotropic analytic algorithm (AAA) on ARIA Version 15.6. SIB plans were optimised to deliver the dose 60Gy in 20 fractions which is according to linear quadratic model (ELQ) is equivalent to 74 Gy in 37 fractions in conventional method.

VMAT and IMRT plans for prostate were created using Varian Eclipse ARIA version 15.6 (Varian Medical Systems, Palo Alto, CA, USA). Dose constraints were applied for both IMRT and VMAT plans according to Table 1.

Table 1. Optimisation objectives

Abbreviation: LT/RT, left and right.

Structures and objective parameters used for VMAT and IMRT plan optimisation

A relatively new feature has been introduced in the latest version of ARIA treatment planning system version 15.6 known as generalised equivalent uniform dose (gEUD) which has been applied as upper dose constraint to bladder, rectum and small bowel. All the plans were made using the beam data of Varian Linear Accelerator Clinac-DHX.

IMRT plans were created using seven gantry angles 0, 30, 60, 105, 180, 255, 300 and 330. Diffident angles can also be used as described in other literature, for example, Sale et al. Reference Sale and Moloney19 suggested gantry angles 0, 75, 105,135, 225 and 270. But, there is no final agreement on the IMRT gantry angles and number of fields. Reference David, Syndikus and Mossop16,Reference Sale and Moloney19,Reference Pasler, Georg and Wirtz20 We followed our institutional practice for IMRT gantry angles and number of fields. VMAT plans were made using single full arc 179–181 or two full arcs as some studies suggested using two full arcs spare OAR in better way. Reference Gautam3 All VMAT plans were made using default collimator angles automatically adjusted by ARIA arc geometry tool feature. All the plans were made to be run on Varian’s Clinac DHX linear accelerator.

Photons of 6 MV were used for creating all the plans. Using high energy has in general no greater benefit in case of IMRT or VMAT. Reference Onal, Sonmez and Erbay9,Reference Pasler, Georg and Wirtz20,Reference Jo, Kim and Baek21 Eclipse treatment planning system 15.6.04 has been used for IMRT and VMAT optimisation and dose calculation.

HI was calculated according to the following formula Reference Yoon, Park and Shin22

(1) $${\rm{HI}} = {{\rm{D}}_{\rm{5}}}{\rm{/}}{{\rm{D}}_{{\rm{95}}}}$$

where D5 and D95 are the doses received by 5 and 95% of the PTV volume, respectively. Smaller and value closer to 1 indicates a homogenous PTV coverage.

Conformity index (CI) was calculated according to the following formula Reference Chow, Jiang and Kiciak23

(2) $${\rm{CI}} = {\rm{PT}}{{\rm{V}}_{{\rm{95\% P}}{\rm{.D}}}}{\rm{/}}{{\rm{V}}_{{\rm{PTV}}}}$$

It is defined as volume of PTV receiving 95% of prescribed dose divided by total volume of PTV.

GI has been calculated as

(3) $${\rm{GI}}=\hskip -1pt{{\rm{V}}_{50\% }}{\rm{/}}{{\rm{V}}_{100\% }}$$

V50% and V100% are simply the isodose volumes of 50 and 100% of the prescribed dose. Reference Chow, Jiang and Kiciak23Reference Paddick and Lippitz25

CI closer to 1 indicates conformal dose to the target volume while GI closer to 1 shows a faster dose fall off and better OAR sparing. Reference Maraghechi, Davis and Badu26

Dose verification and quality assurance

Each plan was evaluated for quality assurance using electronic portal imaging device (EPID), and DTA setting was 3 mm 3% dose–distance criteria to make sure the reproducibility of the plan. Reference Schreibmann, Dhabaan and Elder27Reference Xu, Kang and Jiang29

Statistical analysis

All the statistical analyses were done using MS EXCEL 2013 (Microsoft Corporations). Statistical analysis of the collected data includes the average values of all the plan evaluation parameters that were collected for individual patient from the dose volume histogram (DVH).

Results

The mean PTV 44, 57.6 and 60 volumes were 225.22 cm3 (range: 178.8–275.5 cm3), 199.50 cm3 (range: 147.5–260 cm3) and 116.74 cm3 (range: 79.3–152.5 cm3), respectively.

The plan evaluation parameters are described in Tables 2 and 3.

Table 2. IMRT pkan evaluation parameters

Table 3. VMAT plan evaluation parameters

All the plan evaluation parameters represent the mean values.

Since there were three dose levels (48, 57.6 and 60 Gy), so plan quality parameters D2, D5, D95, Dmean, HI, CI and dose GI were calculated for each individual dose level as shown in Tables 2 and 3 for IMRT and VMAT plans.

As described by Xu et al. Reference Xu, Kang and Jiang29 and definition from Paddick. Reference Paddick and Lippitz25 CI shows the portion of contoured PTV covered by 95% of isodose line. CI should be between 0 and 1.

HI is the ratio between maximum and minimum dose of PTV and its value should be closer to 1. Lower value of HI indicates better PTV homogeneity. Reference Xu, Kang and Jiang29,Reference Quan, Li and Li30

Dose GI describes the dose fall off near the organs at risk. Lower value indicates better dose fall off. All these plan evaluation parameters have been calculated to compare the IMRT and VMAT plans as shown in Tables 1 and 2.

Dose constraints to organ at risk were applied to organ at risk as given in CHHiP trial protocol. Reference David, Syndikus and Mossop16

For PTV 1 (48 Gy), D2(Gy) was higher in VMAT (mean 60.90 vs. 62.30), similarly values of D5 and D95 were higher in VMAT as compared to IMRT as shown in Tables 2 and 3. Dmean was also higher in VMAT (mean 57.51 vs. 59.17 Gy).

Mean value of HI was same for both techniques (1.17). CI was lower and thus better in VMAT than IMRT (mean 1.85 vs. 1.60). GI was also better in VMAT as compared to IMRT (3.82 vs. 4.89). A trend of HI, CI and GI has been shown in Figures 13 for all 15 patients.

Figure 1. Homogeneity index for PTV 48 Gy.

Figure 2. Conformity index for PTV 48 Gy.

Figure 3. Dose gradient index for PTV 48 Gy.

For PTV 2 (57.6 Gy), similar behaviour has been observed where D2, D5, D95 and Dmean were slightly higher in VMAT plans as compared to IMRT as shown in Table 4.

Table 4. OAR dose comparison between IMRT and VMAT

Abbreviation: LT, left; RT, right.

Mean values of CI, HI and GI were lower and thus better in VMAT as compared to IMRT. (1.19, 1.11, 4.78 vs 1.18, 1.09, 7.56) (Figures 46).

Figure 4. Homogeneity index for PTV 57.6 Gy.

Figure 5. Conformity index for PTV 57.6 Gy.

Figure 6. Dose gradient index for PTV 57.6 Gy.

For PTV 3 (60 Gy), D2, D5, D95 and Dmean were slightly higher in case of VMAT. (61.22, 61.13, 58.12, 60.0 vs. 62.41, 62.24, 59.53, 61.12 Gy, respectively).

In terms of CI, IMRT was slightly better than VMAT (1.47 vs. 1.64). HI was same for both techniques (1.04) while GI was much better in VMAT as compared to IMRT (7.89 vs. 14.79) (Figures 79). A lower value of GI indicates sharper dose fall off near OAR. Reference Tsai, Wu and Chao11

Figure 7. Homogeneity index for PTV 60 Gy.

Figure 8. Conformity index for PTV 60 Gy.

Figure 9. Dose gradient index PTV 60 Gy.

It is interesting to note that for higher dose PTV that is PTV 60, dose GI is much better in VMAT.

Organs at risk

In terms of OARs, the dose constraint objectives were easily achieved. For bladder, V40 < 50% was achieved for all patients as shown in Figure 10. There were few exceptions in some patients, for example, as shown in Figure 11, the objective V48 < 25% could not be met for all cases. Similarly, V59.20 < 5% could not be met in all cases as shown in Figure 12.

Figure 10. Bladder V40 Gy IMRT versus VMAT.

Figure 11. Bladder V48Gy IMRT versus VMAT.

Figure 12. Bladder V59.20 Gy IMRT versus VMAT.

IMRT and VMAT showed a mixed conflicting behaviour in terms of which technique is superior for OAR sparing as Tsai et al. Reference Tsai, Wu and Chao11 had described that IMRT and VMAT showed comparable results with VMAT slightly better than IMRT in some cases.

Figures 1012 are showing the comparison of volume of bladder receiving 40, 48 and 59.20 Gy for IMRT and VMAT.

According to CHHiP trial guidelines for OAR doses, Reference David, Syndikus and Mossop16 the volume of bladder receiving 40, 48 and 59.20 Gy (if prescribed dose is 60 Gy in 20 fractions) should be less than 50, 25 and 5%, respectively. In case of 59.20 Gy, IMRT is showing better results compared to VMAT as shown in Figure 13.

Figure 13. Rectum V26.25 Gy IMRT versus VMAT.

In case of rectum, the dose constraint objectives V26.25 < 80%, V43.75 < 60% and V61.25 < 15% were easily achieved as shown in Figures 1315.

Figure 14. Rectum V43.75 Gy IMRT versus VMAT.

Figure 15. Rectum V61.25 Gy IMRT versus VMAT.

VMAT offered better rectum sparing as shown in Figures 13 and 14, respectively. Although there is a spike in graph as shown in Figure 15, but this is not clinically significant since target dose constraint was V70 (61.25) < 15% which was easily achieved for all patients.

Figure 16 is showing a trend of small bowel volume in cubic centimetre (cc) receiving 48 Gy. It should be between 6 and 17 cc as an acceptable value. Reference David, Syndikus and Mossop16 In some patients, small bowel dose was very good in VMAT or RapidArc (RA) technique.

Figure 16. Small bowel V48 Gy IMRT versus VMAT.

Similarly, we measured the maximum dose for right and left femoral heads of 15 patients (Figures 17 and 18) and results showed that VMAT offered much lower femoral head doses than IMRT (35.67 Gy ± 637.20 vs. 40.03 Gy ± 674.23 for RT femoral head and 35.56 Gy ± 561 vs. 40.13 Gy ± 538 for left (LT) femoral head, respectively). This shows VMAT offers better plan quality in terms of OAR sparing as has been described by Quan et al. Reference Quan, Li and Li30

Figure 17. LT femoral head Dmax.

Figure 18. RT femoral head Dmax.

Discussion

Two advanced radiotherapy techniques IMRT and VMAT have replaced the need of 3D conformal radiotherapy as they offer more conformal isodoses to the target while minimising the toxicity risk to organs at risk. Reference Iqbal, Isa and Buzdar31,Reference Zulkafal, Khan and Ahmad32

In this research work, we have compared two planning techniques which are widely used to treat the prostate cancer patients. We compared these two techniques in terms of plan quality parameters which are HI, CI, dose GI and doses of organ at risk.

Fifteen high-risk prostate patients were randomly selected and planned with IMRT and VMAT for hypofratioinated radiotherapy. Plans were optimised and normalised so that PTV 1 should receive 76% of prescribed dose, PTV 2 should receive 91% of prescribed dose and PTV 3 which is high-dose PTV should get at least 95% of prescribed dose. The organs at risk were kept as spare as possible so that coverage of each PTV dose level should not be compromised. All the plans were optimised using AAA and PO on Eclipse treatment planning system ARIA version 15.6.

IMRT and VMAT plans were optimised and dosimetric comparison was made between these two techniques in terms of plan evaluation parameters. These plan evaluation parameters were calculated for each dose level (PTV 48, PTV 57.6 and PTV 60).

If we talk about the results of organ at risk doses, IMRT offered better OAR sparing as compared to VMAT in case of bladder and rectum doses as shown in Table 4. Quan et al. Reference Quan, Li and Li30 has shown the similar behaviour of IMRT plans offering better bladder and rectum doses and he concluded that this is due to greater number of monitor units (MUs) of IMRT plan as compared to VMAT plan. All IMRT plans generated in this study uses seven gantry angles and VMAT plans were generated using two full arcs. Two full arcs offer better OAR sparing as compared to one full arc since two full arc rotations gives more room for beam modulation. Reference Quan, Li and Li30

We cannot conclude which technique is better in terms of OAR sparing based on our conflicting results as in some patients, IMRT provided better bladder and rectum sparing but higher femoral head doses as shown in Table 4. This conflicting behaviour in results could be due to applied dose constraints and PTV definitions which entirely depend upon the position of bladder and rectum. Onal et al. Reference Onal, Sonmez and Erbay9 has also discussed the conflicting behaviour of IMRT and VMAT in terms of OAR sparing. He has described the effect of planning strategies, applied dose calculation algorithms and treatment planning system data on the plan quality. The dose splash in IMRT was greater as compared to VMAT and our data of femoral head maximum doses show that VMAT has better ability to spare femoral heads than IMRT.

We only used 6 MV photon energy to generate the IMRT and VMAT plans as Onal et al. Reference Onal, Sonmez and Erbay9 has described that high energy of photons has no advantage over low energy photons, and neutron generation at 15 MV photon energy cannot be ignored. Reference Onal, Sonmez and Erbay9

Just like OAR doses, the plan quality parameters summarised in Tables 2 and 3 showed a conflicting behaviour. Since we had planned each prostate patient with three different PTV dose levels using SIB, in some dose levels, VMAT offered better conformity than IMRT. HI was almost same in both techniques. However, the dose GI was much better in VMAT. This may indicate a sharper dose fall off near organ at risk offering less acute OAR toxicity. In a study of 292 patients, Kopp et al. Reference Kopp, Duff and Catalfamo33 has discussed that acute rectum toxicity is lower in VMAT and our data of dose GI are describing the similar story.

This study is obviously has come with some limitations. We compared the OAR doses and plan evaluation parameters for a limited number of patients, in an attempt to find that up to how much extent, VMAT can provide better plans. Hence, based on our results of OAR doses and plan quality parameters which are summarised in Tables 24, we cannot definitely conclude which technique is better. As far as our own very busy radiotherapy centre is concerned, VMAT is more feasible as it requires less patient-specific quality assurance (QA) and dose-delivery time.

Conclusion

VMAT is a more efficient dose-delivery technique for high-risk prostate cancer patients in terms of time and without inducing the risk of toxicity to organs at risk. VMAT technique also uses a less number of monitoring units as compared to IMRT. All the plan evaluation parameters including HI, CI were analysed for each individual dose level of PTV and they were almost same for both techniques except dose GI, which is quite better in VMAT (14.79 vs. 7.89) for high-dose PTV3. Better dose GI indicates a sharper dose fall off and better dose conformity. We conclude that VMAT is better and efficient dose-delivery technique than IMRT.

Acknowledgements

The authors acknowledge the Shaukat Khanum Memorial Cancer Hospital and Research Centre for Institutional Review Board (IRB) approval and utilising its services in conduction of this research activity.

Financial Support

This research received no specific grant from any funding agency, commercial or not-for-profit sectors.

Conflict of Interest

None.

Ethical Standards

The authors assert that all procedures contributing to this work comply with the ethical standards and have been approved by the institutional committees (Institutional Review Board Shaukat Khanum Memorial Cancer Hospital and Research Centre Lahore).

No actual human/animal has been affected in conduction of this research work.

References

Riou, O, de la Mothe, P R, Azria, D et al. Simultaneous integrated boost plan comparison of volumetric-modulated arc therapy and sliding window intensity-modulated radiotherapy for whole pelvis irradiation of locally advanced prostate cancer. J App Clin Med Phys 2013; 14 (4): 2635.CrossRefGoogle ScholarPubMed
Gao, M, Li, Q, Ning, Z et al. Dosimetric comparison between step-shoot intensity-modulated radiotherapy and volumetric-modulated arc therapy for upper thoracic and cervical esophageal carcinoma. Med Dosim 2016; 41 (2): 131135.CrossRefGoogle ScholarPubMed
Gautam, B. Literature review on IMRT and VMAT for prostate cancer. Am J Cancer Rev 2014; 2 (1): 5.Google Scholar
Mellon, E A, Javedan, K, Strom, T J et al. A dosimetric comparison of volumetric modulated arc therapy with step-and-shoot intensity modulated radiation therapy for prostate cancer. Pract Radiat Oncol 2015; 5 (1): 1115.CrossRefGoogle ScholarPubMed
Franzese, C, Fogliata, A, D’Agostino, G R et al. Moderate hypofractionated radiotherapy with volumetric modulated arc therapy and simultaneous integrated boost for pelvic irradiation in prostate cancer. J Cancer Res Clin Oncol 2017; 143 (7): 13011309.CrossRefGoogle ScholarPubMed
Hesselberg, G, Fogarty, G, Haydu, L et al. Volumetric modulated arc therapy of the pelvic lymph nodes to the aortic bifurcation in higher risk prostate cancer: early toxicity outcomes. BioMed Res Int 2015; 2015: 696439.CrossRefGoogle ScholarPubMed
Ishii, K, Ogino, R, Hosokawa, Y et al. Whole-pelvic volumetric-modulated arc therapy for high-risk prostate cancer: treatment planning and acute toxicity. J Radiat Res 2015; 56 (1): 141150.CrossRefGoogle ScholarPubMed
Daoud, M A, Saleh, Y M, Elsherbini, M et al. Evaluation of acute toxicity and dosimetric parameters in high risk prostate cancer patients treated by high radiation doses. J Cancer Ther 2019; 10 (8): 654.CrossRefGoogle Scholar
Onal, C, Sonmez, S, Erbay, G et al. Simultaneous integrated boost to intraprostatic lesions using different energy levels of intensity-modulated radiotherapy and volumetric-arc therapy. Brit J Radiol 2014; 87 (1034): 20130617.CrossRefGoogle ScholarPubMed
Johnson, D, Osei, E, Fleck, A et al. Retrospective dosimetric evaluation of VMAT plans for prostate cancer treatment. J Radiother Pract 2019; 18 (2): 155164.Google Scholar
Tsai, C-L, Wu, J-K, Chao, H-L et al. Treatment and dosimetric advantages between VMAT, IMRT, and helical tomotherapy in prostate cancer. Med Dosim 2011; 36 (3): 264271.CrossRefGoogle ScholarPubMed
Zhang, P, Happersett, L, Hunt, M et al. Volumetric modulated arc therapy: planning and evaluation for prostate cancer cases. Int J Radiat Oncol Biol Phys 2010; 76 (5): 14561462.CrossRefGoogle ScholarPubMed
Ost, P, Speleers, B, De Meerleer, G et al. Volumetric arc therapy and intensity-modulated radiotherapy for primary prostate radiotherapy with simultaneous integrated boost to intraprostatic lesion with 6 and 18 MV: a planning comparison study. Int J Radiat Oncol Biol Phys 2011; 79 (3): 920926.CrossRefGoogle ScholarPubMed
Yoo, S, Wu, Q J, Lee, W R et al. Radiotherapy treatment plans with RapidArc for prostate cancer involving seminal vesicles and lymph nodes. Int J Radiat Oncol Biol Phys 2010; 76 (3): 935942.CrossRefGoogle ScholarPubMed
Wolff, D, Stieler, F, Welzel, G et al. Volumetric modulated arc therapy (VMAT) vs. serial tomotherapy, step-and-shoot IMRT and 3D-conformal RT for treatment of prostate cancer. Radiother Oncol 2009; 93 (2): 226233.CrossRefGoogle ScholarPubMed
David, D, Syndikus, I, Mossop, H et al. Conventional versus hypofractionated high-dose intensity-modulated radiotherapy for prostate cancer: 5-year outcomes of the randomised, non-inferiority, phase 3 CHHiP trial. Lancet Oncol 2016; 17 (8): 10471060.Google Scholar
Lee, W R. Hypofractionation for prostate cancer: tested and proven. Lancet Oncol 2016; 17 (8): 10201022.CrossRefGoogle ScholarPubMed
Myrehaug, S, Chan, G, Craig, T et al. A treatment planning and acute toxicity comparison of two pelvic nodal volume delineation techniques and delivery comparison of intensity-modulated radiotherapy versus volumetric modulated arc therapy for hypofractionated high-risk prostate cancer radiotherapy. Int J Radiat Oncol Biol Phys 2012; 82 (4): e657e662.CrossRefGoogle ScholarPubMed
Sale, C, Moloney, P. Dose comparisons for conformal, IMRT and VMAT prostate plans. J Med Imaging Radiat Oncol 2011; 55 (6): 611621.CrossRefGoogle ScholarPubMed
Pasler, P, Georg, D, Wirtz, H et al. Effect of photon-beam energy on VMAT and IMRT treatment plan quality and dosimetric accuracy for advanced prostate cancer. Strahlentherapie und Onkologie 2011; 187 (12): 792798.CrossRefGoogle ScholarPubMed
Jo, G S, Kim, M W, Baek, M G et al. Dosimetric comparison for Prostate VMAT of weight and photon energy change. J Korean Soc Radiat Ther 2018; 30 (1_2): 1725.Google Scholar
Yoon, M, Park, S Y, Shin, D et al. A new homogeneity index based on statistical analysis of the dose–volume histogram. J Appl Clin Med Phys 2007; 8 (2): 917.CrossRefGoogle ScholarPubMed
Chow, J C, Jiang, R, Kiciak, A et al. Dosimetric comparison between the prostate intensity-modulated radiotherapy (IMRT) and volumetric-modulated arc therapy (VMAT) plans using the planning target volume (PTV) dose–volume factor. J Radiother Pract 2016; 15 (3): 263268.CrossRefGoogle Scholar
Cilla, S, Deodato, F, Digesù, C et al. Assessing the feasibility of volumetric-modulated arc therapy using simultaneous integrated boost (SIB-VMAT): an analysis for complex head-neck, high-risk prostate and rectal cancer cases. Med Dosim 2014; 39 (1): 108116.CrossRefGoogle ScholarPubMed
Paddick, I, Lippitz, B. A simple dose gradient measurement tool to complement the conformity index. J Neurosurg 2006; 105 (Supplement): 194201.CrossRefGoogle ScholarPubMed
Maraghechi, B, Davis, J, Badu, S et al. Retrospective analysis of portal dosimetry pre-treatment quality assurance of prostate volumetric-modulated arc therapy (VMAT) plans. J Radiother Pract 2018; 17 (1): 4452.CrossRefGoogle Scholar
Schreibmann, E, Dhabaan, A, Elder, E et al. Patient-specific quality assurance method for VMAT treatment delivery. Med Phys 2009; 36 (10): 45304535.CrossRefGoogle ScholarPubMed
Esposito, M, Bruschi, A, Bastiani, P et al. Characterization of EPID software for VMAT transit dosimetry. Australas Phys Eng Sci Med 2018; 41 (4): 10211027.CrossRefGoogle ScholarPubMed
Xu, L-M, Kang, M-L, Jiang, B et al. A study of the dosimetric characteristics between different fixed-field IMRT and VMAT in early-stage primary mediastinal B-cell lymphoma. Med Dosim 2018; 43 (1): 9199.CrossRefGoogle ScholarPubMed
Quan, E M, Li, X, Li, Y et al. A comprehensive comparison of IMRT and VMAT plan quality for prostate cancer treatment. Int J Radiat Oncol Biol Phys 2012; 83 (4): 11691178.CrossRefGoogle ScholarPubMed
Iqbal, K, Isa, M, Buzdar, S A et al. Treatment planning evaluation of sliding window and multiple static segments technique in intensity modulated radiotherapy. Rep Pract Oncol Radiother 2013; 18 (2): 101106.CrossRefGoogle Scholar
Zulkafal, H M, Khan, M A, Ahmad, M W et al. Volumetric modulated arc therapy treatment planning assessment for low-risk prostate cancer in radiotherapy. Clin Cancer Investig J 2017; 6 (4): 179.CrossRefGoogle Scholar
Kopp, R W, Duff, M, Catalfamo, F et al. VMAT vs. 7-field-IMRT: assessing the dosimetric parameters of prostate cancer treatment with a 292-patient sample. Med Dosim 2011; 36 (4): 365372.CrossRefGoogle ScholarPubMed
Figure 0

Table 1. Optimisation objectives

Figure 1

Table 2. IMRT pkan evaluation parameters

Figure 2

Table 3. VMAT plan evaluation parameters

Figure 3

Figure 1. Homogeneity index for PTV 48 Gy.

Figure 4

Figure 2. Conformity index for PTV 48 Gy.

Figure 5

Figure 3. Dose gradient index for PTV 48 Gy.

Figure 6

Table 4. OAR dose comparison between IMRT and VMAT

Figure 7

Figure 4. Homogeneity index for PTV 57.6 Gy.

Figure 8

Figure 5. Conformity index for PTV 57.6 Gy.

Figure 9

Figure 6. Dose gradient index for PTV 57.6 Gy.

Figure 10

Figure 7. Homogeneity index for PTV 60 Gy.

Figure 11

Figure 8. Conformity index for PTV 60 Gy.

Figure 12

Figure 9. Dose gradient index PTV 60 Gy.

Figure 13

Figure 10. Bladder V40 Gy IMRT versus VMAT.

Figure 14

Figure 11. Bladder V48Gy IMRT versus VMAT.

Figure 15

Figure 12. Bladder V59.20 Gy IMRT versus VMAT.

Figure 16

Figure 13. Rectum V26.25 Gy IMRT versus VMAT.

Figure 17

Figure 14. Rectum V43.75 Gy IMRT versus VMAT.

Figure 18

Figure 15. Rectum V61.25 Gy IMRT versus VMAT.

Figure 19

Figure 16. Small bowel V48 Gy IMRT versus VMAT.

Figure 20

Figure 17. LT femoral head Dmax.

Figure 21

Figure 18. RT femoral head Dmax.