Head mirror versus headlight: illumination, visual identification and visual acuity for otolaryngological examination

Published online by Cambridge University Press:  09 July 2013

C-H Lin
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
H-T Hsu
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
P-Y Chen
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Shuang Ho Hospital, Taipei, Taiwan
L K Huon
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Cathay General Hospital, Taipei, Taiwan
Y-Z Lin
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
S-H Hung*
Affiliation:
Department of Otolaryngology Head and Neck Surgery, Taipei Medical University Hospital, Taipei, Taiwan
*
Address for correspondence: Dr Shih-Han Hung, Department of Otolaryngology, School of Medicine, Taipei Medical University, 250 Wusing St, Sinyi District, Taipei City 110, Taiwan Fax: +886 2 27359903 E-mail: seedturtle@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objective:

To investigate and compare the performance of head mirrors and headlights during otolaryngological examination.

Methods:

The illuminance and illumination field of each device were measured and compared. Visual identification and visual acuity were also measured, in 13 medical students and 10 otolaryngology specialists.

Results:

The illuminance (mean ± standard deviation) of the LumiView, Kimscope 1 W and Kimscope 3 W headlights and a standard head mirror were 352.3 ± 9, 92.3 ± 4.5, 438 ± 15.7 and 68.3 ± 1.2 lux, respectively. The illumination field of the head mirror (mean ± standard deviation) was 348 ± 29.8 grids, significantly greater than that of the Kimscope 3 W headlight (183 ± 9.2 grids) (p = 0.0017). The student group showed no statistically significant difference between visual identification with the best headlight and the head mirror (score means ± standard deviations: 56.2 ± 9 and 53.3 ± 14.1, respectively; p = 0.3). The expert group scored significantly higher for visual identification with head mirrors versus headlights (59.7 ± 3.3 vs 55.2 ± 5.8, respectively; p = 0.0035), but showed no difference for visual acuity.

Conclusion:

Despite the advantages of headlight illumination, head mirrors provided better, shadow-free illumination. Despite no differences amongst students, head mirrors performed better than headlights in experienced hands.

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2013 

Introduction

The head mirror is the most recognisable symbol of an otorhinolaryngologist. In certain cultures, the head mirror is as common as the stethoscope, if not more so, in artists' renderings and popular depictions of doctors. Introduced in 1743 by the French accoucheur Levert for examining the larynx, today's head mirror has withstood the test of time.1 However, an increasing number of doctors are starting to use battery-powered headlights instead of head mirrors. Some otolaryngologists question whether the head mirror can continue to sustain its relevance and trump its younger technological sibling, the battery-powered or fibre-optic headlight.

During our training years, many of us were told of the multiple advantages of the head mirror, the most important of which was the ability to align one's eye sight and the light source axis, which theoretically facilitated the identification of details within narrow and often tubular spaces during the otorhinolaryngological examination.Reference Hannan2 However, there is no clear evidence to support this statement.

In this study, we aimed to evaluate and compare the head mirror and various headlights, through the measurement of illumination, visual identification and visual acuity, while using these instruments under conditions simulating a clinical otorhinolaryngological examination.

Materials and methods

Head mirror and headlights

The head mirror used in the study was manufactured by Nagashima (Tokyo, Japan) and had the following specifications: 89 mm diameter and 13 mm round aperture, with a fibre forehead band.

Three types of headlight were tested: the Kimscope 1 W (model: SLL 01 Warm) (Seahanul Biotech, Anyang, South Korea); the Kimscope 3 W (model: SLL 02 Warm; Seahanul Biotech); and the LumiView Portable Binocular Microscope (Welch Allyn, Skaneatelese Falls, New York, USA). All these headlights emitted warm light, similar to the light source when using a head mirror.

Illuminance

Illuminance was measured using a portable lux meter (Sanpometer LX 1010BS; Sanpo Instrument, Shenzhen, China). The mean spectral sensitivity ± standard deviation (SD) of the meter at 600–620 nm (warm light) was approximately 55 ± 5 per cent, as claimed by the manufacturer.

The lux meter was placed and secured on an examination chair headrest. Illuminance was measured and recorded while the head mirror or headlight directly illuminated the meter sensor from a distance of 50 cm, simulating a common examination condition in daily otorhinolaryngological practice.Reference Mester3

Illumination field

For the rest of the study, the Kimscope 3 W headlight was selected as the representative headlight due to its superior illuminance.

To test illumination field, the examined head was replicated using a full-scale skull model (First Class Human Skull; Anatomical Chart Company, Lippincott, Williams & Wilkins, Philadelphia, Pennsylvania, USA) with highly accurate replication of the nasal cavity, nasal turbinates and nasopharynx. A visual target was attached to the posterior nasopharyngeal wall such that it could be visualised through the nasal cavity and the choana. This visual target was a 4 × 4 cm plain sheet imprinted with a reference grid scale. The distance between the visual target and the light source was 35 cm. The illuminated area was recorded and compared for the head mirror and the Kimscope 3 W headlight.

Visual identification and acuity

Under similar settings as mentioned above, the visual target was replaced with a down-scaled Landolt C table.Reference Danilova and Bondarko4 This table was reduced to 1/12 of its original size in order to match the requirements for a standard visual acuity test at a distance of 50 cm. Sixty-four ‘c' letters ranging in diameter from 2.5 to 16.67 mm were included in the chart, arranged in various orientations (see Figure 1).

Fig. 1 Diagram showing positioning of the target for visual identification (a down-scaled Landolt C table) at the nasopharynx within a full-scale skull model.

Thirteen medical students and 10 otolaryngology specialists were enrolled as volunteer subjects. The participants were asked to identify the visual target at the nasopharynx, as effectively as possible, using the head mirror and headlight, within a period of 3 minutes. In each group, half started the test with the head mirror and the other half started with the headlight. Participants were asked to read out the target Landolt C table as far as possible, without assistance. In the student group, only the number of correctly identified letters was recorded and compared. In the consultant group, visual acuity (i.e. the correct interpretation of the direction of the c letter) was also recorded and compared.

Statistical methods

Data from each group were compared using Student's t-test (paired, two-tailed). A p value of less than 0.05 was taken to indicate a statistically significant difference.

Results and analysis

Illuminance

The mean illuminance ± SD for the LumiView, Kimscope 1 W and Kimscope 3 W headlights and the head mirror was 352.3 ± 9, 92.3 ± 4.5, 438 ± 15.7 and 68.3 ± 1.2 lux, respectively. The Kimscope 3 W headlight gave the highest illuminance of all the devices tested (Figure 2).

Fig. 2 Comparison of illuminance for the three headlights and the head mirror. P < 0.05, comparing all sources.

Illumination field

Figures 3 and 4 show the typical illumination fields generated using the Kimscope 3 W headlight and the head mirror, respectively. The mean ± SD illumination field of the head mirror was 348 ± 29.8 grid squares, significantly greater than that of the Kimscope 3 W headlight (183 ± 9.2 grid squares) (p = 0.0017, two-tailed t-test) (Figure 5). Each grid square was equal to 3.579 mm2.

Fig. 3 Photograph showing typical illumination field projected through the nasospharynx by the Kimscope 3 W headlight.

Fig. 4 Photograph showing typical illumination field projected through the nasospharynx by the head mirror.

Fig. 5 Comparison of the illumination field of the head mirror and Kimscope 3 W headlight. P < 0.05, comparing both sources.

Visual identification and acuity

In the student group, the mean ± SD number of letters identified with the headlight and head mirror were respectively 56.2 ± 9 and 53.3 ± 14.1; this difference was not statistically significant (p = 0.3) (Figure 6).

Fig. 6 Comparison of visual identification percentage for the head mirror and the Kimscope 3 W headlight in the student group.

In the consultant group, surprisingly, the mean ± SD number of letters identified with the head mirror was significantly higher than with the headlight (59.7 ± 3.3 vs 55.2 ± 5.8, respectively; p = 0.0035) (Figure 7). Despite this difference, the consultants' visual acuity was the same regardless of the device used (i.e. whether using a head mirror or headlight, the line number of the smallest correctly identified letters remained identical).

Fig. 7 Comparison of visual identification percentage for the head mirror and the Kimscope 3 W headlight in the consultant group.

Discussion

In 1585, Aranzi was the first to use a light source for an endoscopic procedure, by focusing sunlight through a flask of water and projecting the light into the nasal cavity.

In 1806, Philip Bozzini built an instrument that could be introduced into the human body to visualise the internal organs, and is thus regarded as the father of endoscopy. He used an aluminium tube to visualise the interior body in a novel way: the tube, illuminated by a wax candle, had a fitted mirror to reflect images. In 1901, Dimitri Ott wore a headband mirror to reflect light and augment his visualisation. To this end, he used an access technique whereby a speculum was introduced through an incision in the posterior vaginal fornix.

Thus, head mirrors have a time-honoured history, and are typically used today for the examination of the ear, nose and throat. In recent times, the head mirror has gradually become a symbolic tool used by otorhinolaryngologists in daily practice.

Commonly, the head mirror comprises a circular, concave mirror with a small hole in the middle, which is attached to a head band. The key feature of the head mirror is the combination of the concave mirror and the centre aperture, allowing the axis of the light source to align with the visual axis. This feature theoretically allows the examiner to look deeply into a narrow cavity.

Typically, the patient sits and faces the physician. A bright lamp is positioned adjacent to the patient's head, pointing toward the physician's face and thus toward the head mirror. The light from the lamp reflects off the mirror, along the line of sight of the user. Theoretically, the head mirror provides excellent, shadow-free illumination when used properly.

The main drawback of head mirrors is generally that they require some skill to use well.

Head mirrors are rarely used outside the otorhinolaryngology setting, having been largely replaced by pen lights among general practitioners.

In recent decades, the development and the usage of battery-powered headlights have increased remarkably. This rise in popularity may be related to progress in the field of electronics, especially the introduction of light-weight, rechargeable batteries and light-emitting diode light sources. Headlights no longer require a connection to an external light source with optical fibre cables. Nevertheless, no matter how the headlight is designed, it is difficult to render the axis of the light source and the visual axis coaxial, which can lead to a decrease in visible field while examining a narrow, deep space. However, these opinions remain to be tested scientifically.

Few related studies have been published to date. One study that attempted to justify and compare the use of headlights and head mirrors was performed by Rowlands et al. Reference Rowlands, Mal and Thornton5 These authors recruited 48 medical students and asked them to complete a simple task using a headlight and using a head mirror. The students required more time to complete the task when using the mirror, and overwhelmingly preferred the headlight to the mirror. The authors concluded that headlights should replace head mirrors in routine clinical practice, particularly for newcomers to the specialty.

We observed similar trends in our study. Although the student group's visual identification was slightly better using the headlight rather than the head mirror, this difference was not statistically significant. However, surprisingly, the consultant group had significantly better visual identification when using the head mirror as opposed to the headlight. We presume that, in skilful hands, the light reflected from the mirror is coaxial with the visual axis, and the large area of the concave mirror surface converges a broader area of light into narrow cavities, significantly reducing shadows and ‘blind spots’.

  • Head mirrors and headlights are frequently used in otolaryngology

  • There has been little scientific evaluation of their performance

  • The ideal device would provide maximal illuminance and illumination field, aiding detailed visualisation in narrow spaces

  • In this study, headlights provided best illumination but head mirrors threw fewer shadows

  • Head mirrors performed better in experienced hands

There is evidence supporting a large difference in vision quality under conditions of strong versus weak light intensity.Reference Sjostrand6 Adequate illumination is crucial to achieving good diagnostic accuracy.Reference Hausner7, Reference Perrig, Kazennikov and Wiesendanger8 However, the present study found that, despite the headlight's greater illuminance, visual acuity was similar with both the head mirror and the headlight.

Conclusion

Despite the headlight's greater illuminance, head mirrors provide illumination with fewer shadows. While there seemed to be little difference between headlights and head mirrors amongst inexperienced students, surprisingly, head mirrors performed better in experienced hands.

Acknowledgment

The authors greatly appreciate the assistance of Professor Roger W Chan in the statistical analysis and revision of this manuscript.

References

1ENT Today: An iconic tool: is there still a place for the head mirror? In: http://www.enttoday.org/details/article/836587/An_Iconic_Tool_Is_there_still_a_place_for_the_head_mirror.html [19 March 2012]Google Scholar
2Hannan, SA. The instrument that determined my practice. BMJ 2003;326:747CrossRefGoogle Scholar
3Mester, AF. Six great men in Hungarian otorhinolaryngology. Am J Otolaryngol 1982;3:6772CrossRefGoogle ScholarPubMed
4Danilova, MV, Bondarko, VM. Foveal contour interactions and crowding effects at the resolution limit of the visual system. J Vis 2007;7:118CrossRefGoogle ScholarPubMed
5Rowlands, J, Mal, R, Thornton, M. Headlight or head mirror? Eur Arch Otorhinolaryngol 2006;263:601–2CrossRefGoogle ScholarPubMed
6Sjostrand, FS. Color vision at low light intensity, dark adaptation, Purkinje shift, critical flicker frequency and the deterioration of vision at low illumination. Neurophysiology at the nanometer range of neural structure. J Submicrosc Cytol Pathol 2003;35:117–27Google ScholarPubMed
7Hausner, HH. Human efficiency as a function of light and illumination. Ann N Y Acad Sci 1951;51:1166–78CrossRefGoogle ScholarPubMed
8Perrig, S, Kazennikov, O, Wiesendanger, M. Time structure of a goal-directed bimanual skill and its dependence on task constraints. Behav Brain Res 1999;103:95104CrossRefGoogle ScholarPubMed
Figure 0

Fig. 1 Diagram showing positioning of the target for visual identification (a down-scaled Landolt C table) at the nasopharynx within a full-scale skull model.

Figure 1

Fig. 2 Comparison of illuminance for the three headlights and the head mirror. P < 0.05, comparing all sources.

Figure 2

Fig. 3 Photograph showing typical illumination field projected through the nasospharynx by the Kimscope 3 W headlight.

Figure 3

Fig. 4 Photograph showing typical illumination field projected through the nasospharynx by the head mirror.

Figure 4

Fig. 5 Comparison of the illumination field of the head mirror and Kimscope 3 W headlight. P < 0.05, comparing both sources.

Figure 5

Fig. 6 Comparison of visual identification percentage for the head mirror and the Kimscope 3 W headlight in the student group.

Figure 6

Fig. 7 Comparison of visual identification percentage for the head mirror and the Kimscope 3 W headlight in the consultant group.