Electrode design and insertional depth-dependent intra-cochlear pressure changes: a model experiment

Published online by Cambridge University Press:  06 November 2017

P Mittmann*
Affiliation:
Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Germany
A Ernst
Affiliation:
Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Germany
I Todt
Affiliation:
Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Germany
*
Address for correspondence: Dr Philipp Mittmann, Department of Otolaryngology, Head and Neck Surgery, Unfallkrankenhaus Berlin, Warener Str. 7, 12683 Berlin, Germany Fax: +49 30 5681 4303 E-mail: philipp.mittmann@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Background:

Preservation of residual hearing is one of the major goals in modern cochlear implant surgery. Intra-cochlear fluid pressure changes influence residual hearing, and should be kept low before, during and after cochlear implant insertion.

Methods:

Experiments were performed in an artificial cochlear model. A pressure sensor was inserted in the apical part. Five insertions were performed on two electrode arrays. Each insertion was divided into three parts, and statistically evaluated in terms of pressure peak frequency and pressure peak amplitude.

Results:

The peak frequency over each third part of the electrode increased in both electrode arrays. A slight increase was seen in peak amplitude in the lateral wall electrode array, but not in the midscalar electrode array. Significant differences were found in the first third of both electrode arrays.

Conclusion:

The midscalar and lateral wall electrode arrays have different intra-cochlear fluid pressure changes associated with intra-cochlear placement, electrode characteristics and insertion.

Type
Main Articles
Copyright
Copyright © JLO (1984) Limited 2017 

Introduction

Cochlear implantation for the auditory rehabilitation of deaf patients (i.e. those with profound sensorineural hearing loss), and nowadays patients with substantial residual hearing, has become a worldwide-accepted procedure. As the number of implantations rises, the challenges and demands of modern surgery increase, and each sub-step of the surgical procedure is the focus of clinical research.

Recipients with preserved residual low-frequency hearing have significantly improved discrimination scores and hearing benefits in challenging listening surroundings.Reference Miranda, Sampaio, Lopes, Ramos Venosa and de Oliveira 1 Techniques have been developed to minimise the insertion trauma in surgery performed to preserve residual hearing.Reference Adunka, Unkelbach, Mack, Hambek, Gstoettner and Kiefer 2 Reference Sun, Hsu, Chen and Wu 5 In addition to surgical expertise, careful selection of the electrode array is an important sub-step of the procedure. Preservation of residual hearing can be achieved by implanting either short,Reference Gantz, Dunn, Walker, Van Voorst, Gogel and Hansen 6 , Reference Gantz and Turner 7 standardReference Gstottner, Pok, Peters, Kiefer and Adunka 8 or perimodiolar electrodes.Reference Fraysse, Macias, Sterkers, Burdo, Ramsden and Deguine 4 , Reference Balkany, Connell, Hodges, Payne, Telischi and Eshraghi 9 The latter are associated with an increased risk of intra-cochlear trauma,Reference Aschendorff, Kromeier, Klenzner and Laszig 10 and are hence less commonly chosen for hearing preservation. Nevertheless, these electrodes tend to be close to the modiolus, provide a good audiological outcome and have better electrophysiological abilities.

Lateral wall electrodes are known to cause less intra-cochlear trauma and are favourable for hearing preservation.Reference O'Connell, Hunter, Gifford, Rivas, Haynes and Noble 11 Midscalar electrodes have been designed to combine the electrophysiological benefits of perimodiolar electrodes and the atraumatic behaviour of lateral wall electrodes.

Important clinically evaluated factors for the preservation of residual hearing include: the design of the cochlear electrode,Reference Roland 12 Reference Todd, Naghdy and Svehla 14 peri- and post-operative medications,Reference Ye, Tillein, Hartmann, Gstoettner and Kiefer 15 sealing of the cochleostomy,Reference Burghard, Lenarz, Kral and Paasche 16 and electrode insertion speed.Reference Rajan, Kontorinis and Kuthubutheen 17 , Reference Todt, Mittmann and Ernst 18

Reliable hearing preservation is a major goal in modern cochlear implant surgery. The intra-cochlear fluid pressure changes that occur during cochlear implantation are assumed to influence the intra-cochlear structures and can lower residual hearing. These intra-cochlear fluid pressure changes depend on different parameters during the cochlear implant surgery and should therefore be kept low in general.

As intra-cochlear fluid pressure changes occur in the pre- and post-insertion period, as well as during the insertion itself, each sub-step needs specific attention. In the pre-insertion period, the opening of the round window membrane can be associated with significant intra-cochlear fluid pressure changes.Reference Mittmann, Ernst and Todt 19 Reference Todt, Ernst and Mittmann 21 In the post-insertion period, the sealing of the electrode array in the round window and the cable movements after complete insertion influence the intra-cochlear fluid pressure changes.Reference Todt, Karimi, Luger, Ernst and Mittmann 22 The intra-cochlear fluid pressure changes during electrode insertion are tremor dependent, but can be influenced by the insertion speed and stabilisation of the inserting hand.Reference Todt, Mittmann and Ernst 18 , Reference Todt, Ernst and Mittmann 23 Tremor is assumed to have an impact on the clinical outcome in cochlear implant surgery.Reference Todt, Ernst and Mittmann 23 Electrode movements during the insertion can cause uncontrolled fast and static pressure changes that enhance intra-cochlear trauma.

Furthermore, electrodes with different designs, volumes, lengths and characteristics are associated with different absolute intra-cochlear fluid pressure changes and with a different course of intra-cochlear fluid pressure changes.Reference Todt, Mittmann, Ernst and Mittmann 24 In high-volume lateral wall electrodes, the intra-cochlear fluid pressure changes decrease with progressive insertional depth.Reference Mittmann, Ernst, Mittmann and Todt 25

This study aimed to compare how the intra-cochlear pressure changes when increasing the cochlear implant electrode insertion depth in a midscalar electrode and a prototype, small-volume lateral wall electrode.

Materials and methods

An artificial and transparent full-scalar cochlear model, employed in former studies by the same research group,Reference Todt, Mittmann and Ernst 18 Reference Todt, Ernst and Mittmann 21 , Reference Todt, Ernst and Mittmann 23 Reference Mittmann, Ernst, Mittmann and Todt 25 was used in all experiments. The diameter of the round window in the model was 1.5 mm, which is slightly larger than that in the human temporal bone.Reference Paprocki, Biskup, Kozlowska, Kuniszyk, Bien and Niemczyk 26 The total volume of the model was 87 mm3. The sensor tip was inserted in an extra channel (about 800 µm) in the apex.

The model was filled with pure water, and the pressure sensor was sealed and fixed with fibrin glue within the channel. Neither the channel wall nor the floor was in direct contact with the sensor tip. The cochlear model was checked for enclosed air bubbles under the microscope. After every experiment, it was refilled and rechecked for enclosed bubbles. The experiments were performed with moisturised electrodes, and in series with the sensor in an unchanged position, in order to exclude sensor position-related bias and to allow inter-experimental comparability.

Pressure sensor

The 0.8 mm diameter micro-optical pressure sensor (FOP-M; FISO, Quebec City, Canada) was used to measure the intra-cochlear pressure changes. Details regarding the design, fabrication and capacity of the pressure sensor, designed by Olsen, can be found in the literature.Reference Olson 27 Generally, the pressure sensor is a hollow glass tube, which is tipped and sealed by a thin plastic diaphragm, coated with a reflective surface of gold on one side. An optical fibre delivers light from a light-emitting diode, which fans out and is reflected by the gold-coated diaphragm. The gold-coated membrane bends linearly with changes in the pressure and reflects the light. A photodetector senses the emitted light by taking 5000 measurements per second.Reference Olson 27

Electrodes

The Advanced Bionics (Stäfa, Switzerland) HiFocus Midscalar electrode array and the prototype Lateral Wall 23 electrode array were used for all the experiments. The electrode insertions were made using the insertion tool recommended by Advanced Bionics for the HiFocus Midscalar electrode array and forceps for the Lateral Wall 23 electrode array.

The HiFocus Midscalar electrodes have a tip size of 0.5 mm and an end size of 0.7 mm, with a total insertion length of 18.5 mm. The total volume is 6.5 mm3 and the mean volume per mm is 0.35 mm3. The Lateral Wall 23 electrodes have a tip size of 0.22 mm × 0.55 mm and an end size of 0.6 mm × 0.8 mm, with a total insertion length of 23 mm. The total volume is 5.19 mm3 and the mean volume per mm is 0.23 mm3.

Insertions were performed using the one-point supported insertion technique, where only the elbow of the inserting arm is rested on the table.Reference Todt, Ernst and Mittmann 23 The insertion speed was 0.5 mm/second for all measurements. The insertions took approximately 37 seconds and 46 seconds respectively, and were performed by the senior author.

Statistical analysis

GNOME Evolution software, Microsoft Excel (Santa Rosa, California, USA) and SPSS software, version 22.0 (SPSS, Chicago, Illinois, USA), were used to record and analyse the data.

Five complete insertions were performed with each electrode array. The individual measurements were split into thirds to further analyse the data. For each electrode third, the number of peaks was counted (pressure peak frequency) and the three greatest amplitudes were analysed (pressure peak amplitude).

Statistical evaluation was performed using one-way analysis of variance (ANOVA) and unpaired t-tests. The data are presented as means ± standard deviations (SDs). P values of less than 0.05 were considered statistically significant.

Results

Pressure peak frequency changes

In order to determine if there were differences in the number of peaks between the two electrode arrays, a one-way ANOVA was conducted for each array. Each electrode array was inserted five times. The data were normally distributed for both electrode arrays, as assessed by the Shapiro–Wilk test (p > 0.05). Homogeneity of variance was observed, as assessed by Levene's test of homogeneity of variances (p > 0.05).

There was no statistically significant difference in the mean number of peaks between each evaluated third for both the HiFocus Midscalar electrode array (F(2,12) = 0.369, p > 0.05) and the Lateral Wall 23 electrode array (F(2,12) = 0.092, p > 0.05). Post-hoc testing revealed no significant differences either. However, there was a slight increase in the number of peaks for the HiFocus Midscalar electrode array from the first third (mean number of peaks ± SD = 6 ± 1.87) to the second third (mean ± SD = 6.4 ± 1.14) to the last third (mean ± SD = 6.8 ± 1.3). For the Lateral Wall 23 electrode array, an increase from the first third (mean number of peaks ± SD = 8 ± 1) to the second third (mean ± SD = 8.4 ± 2.07) and a slight decrease to the last third (mean ± SD = 8.2 ± 1.1) was found (Table I).

Table I Mean peak pressure frequency and amplitude for both electrode arrays

Data represent means ± standard deviations.

An independent t-test was conducted to determine whether there was a significant difference between the peak frequency within the thirds between both electrode arrays. In all the thirds, a higher peak frequency was observed in the Lateral Wall 23 electrode array than in the HiFocus Midscalar electrode array, but the difference was not statistically significant (first third comparison, p = 0.068; second third comparison, p = 0.095; last third comparison, p = 0.103) (Table I).

Pressure peak amplitude changes

A one-way ANOVA was performed to determine whether there were any differences in the measured peak amplitudes between both electrode arrays. Each insertion was measured five times, similar to the peak frequency investigation.

Homogeneity of variance could not be assumed for the measured amplitudes in the last third of the Lateral Wall 23 electrode array group, as assessed by Levene's test of homogeneity of variance (p = 0.003). There was no statistically significant difference in the means of the peak amplitude between each evaluated third for both the HiFocus Midscalar electrode array (F(2,12) = 1.578, p > 0.05) and the Lateral Wall 23 electrode array (F(2,12) = 0.0225, p > 0.05). Post-hoc testing revealed no significant difference either. A slight decrease in peak amplitude was seen for the HiFocus Midscalar electrode array from the first third (0.13 ± 0.06 mmHg) to the second third (0.06 ± 0.02 mmHg) to the last third (0.12 ± 0.09 mmHg). For the Lateral Wall 23 electrode array, a smooth increase from the first third (0.05 ± 0.03 mmHg) to the second third (0.06 ± 0.02 mmHg) and to the last third (0.06 ± 0.01 mmHg) was found (Table I).

An independent t-test was conducted to determine whether there was a significant difference within the thirds between both electrode arrays, similar to peak frequency. Higher peak amplitudes were seen in the HiFocus Midscalar electrode array group in all the thirds. Regarding the first third, the difference of 0.08 mmHg was statistically significant (95 per cent confidence interval (CI) = 0.01 to 0.15, p = 0.031). In the second and last thirds, the difference was not statistically significant (second third comparison, p = 0.768; last third comparison, p = 0.191) (Table I, Figures 1 and 2).

Fig. 1 Typical pressure course associated with insertion of the HiFocus Midscalar electrode array.

Fig. 2 Typical pressure course associated with insertion of the Lateral Wall 23 electrode array.

Discussion

The value of cochlear implantation for the hearing rehabilitation of patients with severe to profound hearing loss has been established worldwide. The indication criteria have changed during the past few decades; nowadays, in many countries it includes patients with unilateral deafness as well as patients with sustainable residual hearing. Preservation of residual hearing has become one of the major goals in modern cochlear implant surgery.Reference Miranda, Sampaio, Lopes, Ramos Venosa and de Oliveira 1 , Reference Fraysse, Macias, Sterkers, Burdo, Ramsden and Deguine 4 , Reference Sun, Hsu, Chen and Wu 5 , Reference Havenith, Lammers, Tange, Trabalzini, della Volpe and van der Heijden 28

Insertion of the cochlear implant electrode arrays leads to intra-cochlear fluid pressure changes,Reference Todt, Ernst and Mittmann 23 Reference Mittmann, Ernst, Mittmann and Todt 25 which may affect residual hearing preservation and hence the audiological outcome.Reference Roland 12 Therefore, the intra-cochlear fluid pressure changes should be minimised before, during and after cochlear implantation.Reference Todt, Mittmann and Ernst 18 , Reference Mittmann, Ernst and Todt 19 , Reference Todt, Ernst and Mittmann 23 , Reference Todt, Mittmann, Ernst and Mittmann 24

The intra-cochlear fluid pressure changes during cochlear implant insertion are influenced by several factors. Significant reductions in terms of maximum pressure gain, peak frequency and peak amplitude can be achieved by inserting the electrode array with a supported techniqueReference Todt, Ernst and Mittmann 23 at a low insertional speed.Reference Todt, Mittmann and Ernst 18

Variations in the volume, tip size, shape, design and intra-cochlear position of the electrode array itself can influence the intra-cochlear fluid pressure. Usage of a stylet or sheath stabilises the electrodes, and reduces electrode-induced intra-cochlear fluid pressure changes by minimising intra-cochlear movement.Reference Todt, Mittmann, Ernst and Mittmann 24 With increasing insertional depth, the lateral wall electrodes stabilise by increasing contact with the lateral wall. As a result, the intra-cochlear movement reduces, and the number of pressure peaks and the pressure amplitude decreases with progressive insertional depth.Reference Mittmann, Ernst, Mittmann and Todt 25

The Advanced Bionics HiFocus Midscalar® electrode array and the prototype Advanced Bionics Lateral Wall 23® electrode array were used in this study. The HiFocus Midscalar electrode array was inserted with the provided insertion tool that pushes the electrode off a stylet. The first third of the electrode is inserted manually into the cochlea; the tool then smoothly pushes the electrode off the stylet.

Regarding peak amplitude, there is a primary high peak amplitude in the first third (manually inserted electrode part), followed by a sudden decrease when the electrode is stabilised by the stylet. The natural tremor is reduced. For the final third, the electrode loses guidance from the tool and stylet; therefore, the peak amplitude rises.

The HiFocus Midscalar electrode array showed higher peak amplitudes than the Lateral Wall 23 electrode array in all the thirds, but the difference was significant only in the first third. The latter finding is assumed to be a result of the higher volume of the HiFocus Midscalar electrode array. The HiFocus Midscalar electrode array, inserted with the tool, is not guided by the lateral wall, which could increase the absolute movement within the cochlea. In contrast, the lateral wall stabilises the Lateral Wall 23 electrode array.

Compared with the Advanced Bionics IJ® electrode array,Reference Mittmann, Ernst, Mittmann and Todt 25 for which the peak amplitude continuously decreases with ongoing insertional depth, the peak amplitude in the Lateral Wall 23 electrode array increases. Despite the fact that both are lateral wall electrode arrays, stiffness, volume and tip size appear to influence intra-cochlear fluid pressure changes; this is more distinct in smaller volume electrodes. This is in contrast to our previous study,Reference Mittmann, Ernst, Mittmann and Todt 25 and we had expected to find an intra-cochlear fluid pressure reduction with increasing insertional depth, associated with the stabilisation of the electrodes by the cochlear wall. The Lateral Wall 23 electrode array has a smaller volume and is less stiff than the Advanced Bionics IJ electrode array, which might increase the intra-cochlear fluid pressure. Intra-cochlear fluid pressure changes are influenced by the insertion design (forceps vs Advanced Bionics IJ insertion tool), as well as the electrode array dimensions and characteristics.

Regarding peak frequency, the Lateral Wall 23 electrode array showed a greater number of peaks in every third than the HiFocus Midscalar electrode array. Although the differences between the thirds were not statistically significant, there was a visible tendency for reduced intra-cochlear pressure changes when using the HiFocus Midscalar electrode array. The peak frequency reflects the manual tremor. As the peak frequency is almost the same, a comparable manual tremor in all the insertions can be assumed. Compared with the Advanced Bionics IJ® electrode array,Reference Mittmann, Ernst, Mittmann and Todt 25 the course of the number of peaks with progressive insertional depth of the Lateral Wall 23 electrode array was similar, but increased in the HiFocus Midscalar electrode array group.

Our study has some limitations associated with use of the artificial cochlear model. The comparisons between both electrode arrays is challenging as they differ in insertion design, as well as in volume, diameter and stiffness. Furthermore, the transformation of the rapid intra-cochlear fluid pressure changes in the human cochlea might be challenging to assess. The absolute intra-cochlear fluid pressure values in vivo might be different, as the relationship between the artificial model and the electrode array volume is different from the in vivo situation. The surface of the cochlear model is different from the human cochlea, which adds a certain resistance to the insertion. The resistance of the artificial model has to be overcome; it remains unknown how this resistance is comparable to that of the human cochlea.Reference Kontorinis, Paasche, Lenarz and Stover 29

  • Preservation of residual hearing is a major goal in cochlear implant surgery

  • Intra-cochlear fluid pressure changes influence residual hearing

  • Intra-cochlear fluid pressure changes should be kept low before, during and after cochlear implant insertion

  • The electrode array itself has an impact on intra-cochlear fluid pressure changes

  • Perimodiolar and lateral cochlear implant electrode arrays have variable intra-cochlear fluid pressure effects

Another aspect is the drain of the human cochlea. The model was sealed with fibrin glue in the apex; therefore, fluid loss was only observed through the round window. The relationship between the round window opening and the electrode array influences the amount of intra-cochlear fluid leakage and pressure, as it is slightly wider than the relationship in vivo. This smaller relationship might increase the in vivo intra-cochlear fluid pressure and this should be kept in mind. The natural main pathway for pressure equilibration is the cochlear aqueduct.Reference Ciuman 30 The intra-cochlear fluid pressure changes can be transferred between the different labyrinthine compartments.Reference Park, Boeven, Vogel, Leonhardt, Wit and Westhofen 31 However, the direct pressure transfer into these compartments has to be assumed to be limited in vivo. Reference Laurens-Thalen, Wit, Segenhout and Albers 32 Furthermore, our measurements consider only the intra-cochlear fluid pressure changes after the round window insertion, and were performed by one cochlear implantation surgeon only. These results may only be applicable to the surgeons who use the round window approach and who are familiar with these electrode arrays.

Conclusion

Intra-cochlear fluid pressure changes are assumed to affect residual hearing preservation and should be minimised. The midscalar and lateral wall electrode arrays have different intra-cochlear fluid pressure effects associated with intra-cochlear placement, electrode characteristics and insertion. Nevertheless, implant insertion into the human cochlea remains challenging and requires further investigation.

References

1 Miranda, PC, Sampaio, AL, Lopes, RA, Ramos Venosa, A, de Oliveira, CA. Hearing preservation in cochlear implant surgery. Int J Otolaryngol 2014;2014:468515 Google Scholar
2 Adunka, O, Unkelbach, MH, Mack, M, Hambek, M, Gstoettner, W, Kiefer, J. Cochlear implantation via the round window membrane minimizes trauma to cochlear structures: a histologically controlled insertion study. Acta Otolaryngol 2004;124:807–12Google Scholar
3 Carlson, ML, Driscoll, CL, Gifford, RH, Service, GJ, Tombers, NM, Hughes-Borst, BJ et al. Implications of minimizing trauma during conventional cochlear implantation. Otol Neurotol 2011;32:962–8Google Scholar
4 Fraysse, B, Macias, AR, Sterkers, O, Burdo, S, Ramsden, R, Deguine, O et al. Residual hearing conservation and electroacoustic stimulation with the Nucleus 24 Contour Advance cochlear implant. Otol Neurotol 2006;27:624–33Google Scholar
5 Sun, CH, Hsu, CJ, Chen, PR, Wu, HP. Residual hearing preservation after cochlear implantation via round window or cochleostomy approach. Laryngoscope 2015;125:1715–19CrossRefGoogle ScholarPubMed
6 Gantz, BJ, Dunn, C, Walker, E, Van Voorst, T, Gogel, S, Hansen, M. Outcomes of adolescents with a short electrode cochlear implant with preserved residual hearing. Otol Neurotol 2016;37:e11825 Google Scholar
7 Gantz, BJ, Turner, CW. Combining acoustic and electrical hearing. Laryngoscope 2003;113:1726–30Google Scholar
8 Gstottner, W, Pok, SM, Peters, S, Kiefer, J, Adunka, O. Cochlear implantation with preservation of residual deep frequency hearing [in German]. HNO 2005;53:784–90Google Scholar
9 Balkany, TJ, Connell, SS, Hodges, AV, Payne, SL, Telischi, FF, Eshraghi, AA et al. Conservation of residual acoustic hearing after cochlear implantation. Otol Neurotol 2006;27:1083–8Google Scholar
10 Aschendorff, A, Kromeier, J, Klenzner, T, Laszig, R. Quality control after insertion of the Nucleus Contour and Contour Advance electrode in adults. Ear Hear 2007;28:75S79S Google Scholar
11 O'Connell, BP, Hunter, JB, Gifford, RH, Rivas, A, Haynes, DS, Noble, JH et al. Electrode location and audiologic performance after cochlear implantation: a comparative study between Nucleus CI422 and CI512 electrode arrays. Otol Neurotol 2016;37:1032–5Google Scholar
12 Roland, JT Jr. A model for cochlear implant electrode insertion and force evaluation: results with a new electrode design and insertion technique. Laryngoscope 2005;115:1325–39Google Scholar
13 Sennaroglu, L, Atay, G, Bajin, MD. A new cochlear implant electrode with a “cork”-type stopper for inner ear malformations. Auris Nasus Larynx 2014;41:331–6Google Scholar
14 Todd, CA, Naghdy, F, Svehla, MJ. Force application during cochlear implant insertion: an analysis for improvement of surgeon technique. IEEE Trans Biomed Eng 2007;54:1247–55CrossRefGoogle ScholarPubMed
15 Ye, Q, Tillein, J, Hartmann, R, Gstoettner, W, Kiefer, J. Application of a corticosteroid (Triamcinolon) protects inner ear function after surgical intervention. Ear Hear 2007;28:361–9Google Scholar
16 Burghard, A, Lenarz, T, Kral, A, Paasche, G. Insertion site and sealing technique affect residual hearing and tissue formation after cochlear implantation. Hear Res 2014;312:21–7Google Scholar
17 Rajan, GP, Kontorinis, G, Kuthubutheen, J. The effects of insertion speed on inner ear function during cochlear implantation: a comparison study. Audiol Neurootol 2013;18:1722 CrossRefGoogle ScholarPubMed
18 Todt, I, Mittmann, P, Ernst, A. Intracochlear fluid pressure changes related to the insertional speed of a CI electrode. Biomed Res Int 2014;2014:507241 CrossRefGoogle Scholar
19 Mittmann, P, Ernst, A, Todt, I. Intracochlear pressure changes due to round window opening: a model experiment. ScientificWorldJournal 2014;2014:341075 Google Scholar
20 Mittmann, P, Ernst, A, Mittmann, M, Todt, I. Optimisation of the round window opening in cochlear implant surgery in wet and dry conditions: impact on intracochlear pressure changes. Eur Arch Otorhinolaryngol 2016;273:3609–13Google Scholar
21 Todt, I, Ernst, A, Mittmann, P. Effects of round window opening size and moisturized electrodes on the intracochlear pressure related to the insertion of a cochlear implant electrode. Audiol Neurotol Extra 2016;6:18 Google Scholar
22 Todt, I, Karimi, D, Luger, J, Ernst, A, Mittmann, P. Postinsertional cable movements of cochlear implant electrodes and their effects on intracochlear pressure. Biomed Res Int 2016;2016:3937196 Google Scholar
23 Todt, I, Ernst, A, Mittmann, P. Effects of different insertion techniques of a cochlear implant electrode on the intracochlear pressure. Audiol Neurootol 2016;21:30–7Google Scholar
24 Todt, I, Mittmann, M, Ernst, A, Mittmann, P. Comparison of the effects of four different cochlear implant electrodes on intra-cochlear pressure in a model. Acta Otolaryngol 2017;137:235–41Google Scholar
25 Mittmann, M, Ernst, A, Mittmann, P, Todt, I. Insertional depth-dependent intracochlear pressure changes in a model of cochlear implantation. Acta Otolaryngol 2017;137:113–18CrossRefGoogle Scholar
26 Paprocki, A, Biskup, B, Kozlowska, K, Kuniszyk, A, Bien, D, Niemczyk, K. The topographical anatomy of the round window and related structures for the purpose of cochlear implant surgery. Folia Morphol (Warsz) 2004;63:309–12Google Scholar
27 Olson, ES. Observing middle and inner ear mechanics with novel intracochlear pressure sensors. J Acoust Soc Am 1998;103:3445–63Google Scholar
28 Havenith, S, Lammers, MJ, Tange, RA, Trabalzini, F, della Volpe, A, van der Heijden, GJ et al. Hearing preservation surgery: cochleostomy or round window approach? A systematic review. Otol Neurotol 2013;34:667–74Google Scholar
29 Kontorinis, G, Paasche, G, Lenarz, T, Stover, T. The effect of different lubricants on cochlear implant electrode insertion forces. Otol Neurotol 2011;32:1050–6Google Scholar
30 Ciuman, RR. Communication routes between intracranial spaces and inner ear: function, pathophysiologic importance and relations with inner ear diseases. Am J Otolaryngol 2009;30:193202 CrossRefGoogle ScholarPubMed
31 Park, JJ, Boeven, JJ, Vogel, S, Leonhardt, S, Wit, HP, Westhofen, M. Hydrostatic fluid pressure in the vestibular organ of the guinea pig. Eur Arch Otorhinolaryngol 2012;269:1755–8Google Scholar
32 Laurens-Thalen, EO, Wit, HP, Segenhout, JM, Albers, FW. Direct measurement flow resistance of cochlear aqueduct in guinea pigs. Acta Otolaryngol 2004;124:670–4Google Scholar
Figure 0

Table I Mean peak pressure frequency and amplitude for both electrode arrays

Figure 1

Fig. 1 Typical pressure course associated with insertion of the HiFocus Midscalar electrode array.

Figure 2

Fig. 2 Typical pressure course associated with insertion of the Lateral Wall 23 electrode array.