The contaminated surface environment of hospital rooms has been demonstrated to be a source of patient-to-patient transmission of multidrug-resistant organisms (MDROs), including methicillin-resistant Staphylococcus aureus (MRSA).Reference Weber, Rutala, Miller, Huslage and Sickbert-Bennett 1 , Reference Weber, Anderson and Rutala 2 MDROs can survive on surfaces for days to weeks (for months in some cases).Reference Kramer, Schwebke and Kampf 3 Following terminal room cleaning and disinfection, frequent contamination by these pathogens has been reported.Reference Weber, Rutala, Miller, Huslage and Sickbert-Bennett 1 , Reference Weber, Anderson and Rutala 2 Patients admitted to a room where the previous occupant was also colonized or infected with an MDRO are at substantial risk for acquisition of this MDRO.Reference Otter, Yezli, Salkeld and French 4 For these reasons, “no-touch” methods of room disinfection have been evaluated, including devices that emit ultraviolet (UV) light or produce hydrogen peroxide vapor.Reference Weber and Rutala 5 Use of UV devices has been demonstrated to reduce healthcare-associated pathogens.Reference Weber and Rutala 5
Carbapenem-resistant Enterobacteriaceae (CRE) are a growing problem in the United States and worldwide. 6 CRE infections are difficult to treat and have been associated with substantial mortality. Frequent contamination of room surfaces near CRE carriers has been described.Reference Lerner, Adler and Abu-Hanna 7 For this reason, we assessed the efficacy of a UV-C device to inactivate MRSA and CRE. We also assessed whether 2 cycles of the device (in which 1 cycle was completed on the right side of the bed and 1 cycle on the left side of the bed) were superior to a single cycle (in which the device was placed at the foot of the bed).
In this study, we evaluated the Optimum UV (Clorox, Pleasanton, CA), a fixed cycle-time UV-C (254 nm) device with four 157-cm lamps in an unoccupied patient room (24.76 m2). To assess the efficacy of 2 different decontamination methods, the UV-C device was placed either at the end of the bed in the center of room in a single cycle (design A) or at both sides of the bed in the center of room with a single cycle at the right side and another cycle at left side (design B). The cycle time of the device for room decontamination was fixed at 5 minutes. Each design was tested in triplicate. Locations where test Formica sheets were placed were categorized as the UV-C direct group (laser point visible on site) or the UV-C indirect group (laser point not visible on site). Testing was performed using clinical isolates of MRSA and carbapenem-resistant Klebsiella pneumoniae (CRKP). A 10–15 µL inoculum containing ~106–107 test organisms per replicate organism detection and counting (Rodac) template was spread separately on each Formica sheet using a sterile glass hockey loop. Inoculated Formica sheets were placed and attached with tape in the following 10 locations throughout the room for both designs: (1) the back of the computer (indirect/indirect); (2) the side of the sink counter (indirect/direct); (3) the vertical end of the foot of the bed (direct/indirect); (4) the horizontal top of the bed (direct/direct); (5) the vertical back of the visitor’s chair (indirect/indirect); (6) the horizontal counter in front of the refrigerator (direct/direct); (7) the vertical end of the head of the bed (indirect/indirect); (8) the vertical side of the bedside table facing the window (indirect/indirect); (9) the horizontal top of the toilet rim in the bathroom (indirect/indirect); and (10) the horizontal top of the couch table (direct/direct). After UV-C cycle completion, each Formica sheet was cultured using Rodac plates containing Day-Engley (DE) neutralizing agar. These plates were incubated aerobically at 37°C for 48 hours for CRKP and MRSA. The colony-forming units (CFU) of the test organisms on each plate were then counted. When Rodac plates had confluent growth, the number of CFUs per plate was set at 300 for analysis. Statistical significance was determined by paired t test; P<.05 was considered significant.
Overall, in an unoccupied patient room, we observed a 5.01-log10 reduction using design A and a 5.87-log10 reduction using design B for decontamination of CRKP (P=.0003). We observed a 4.61-log10 reduction and a 5.06-log10 reduction for MRSA (P=.0418) (Table 1). At UV-C direct sites, we measured an additional 0.87-log10 reduction using design B for CRKP (P=.0028) and an additional 0.55-log10 reduction for MRSA (P>.05) compared with design A. When we assessed the correlation of log10 reduction (in direct line of sight) to distance, there was a strong correlation (r=−0.826, P=.0009 for CRKP and r=−0.842, P=.0006 for MRSA); however, we detected a ≥4.89-log10 reduction for CRKP and a ≥4.19-log10 reduction for MRSA at the farthest distance (2.69 m). At UV-C indirect sites, we measured an additional 0.86-log10 reduction using design B for CRKP (P=.0002) and an additional 0.38-log10 reduction for MRSA (P=.0113) compared to design A. The frequency of too-numerous-to-count (TNTC) cultures was as follows: MRSA 42% (25 of 60) and CRKP 32% (19 of 60). Data for microbial inactivation without TNTC results are shown in the Supplemental Material.
TABLE 1 Microbial Reduction in Formica Surfaces Experimentally Contaminated with Multidrug-Resistant Pathogens Using a Fixed Cycle-Time Ultraviolet-C (UV-C) Device and 2 Different Radiation Designs
![](https://static.cambridge.org/binary/version/id/urn:cambridge.org:id:binary:20160921002925677-0598:S0899823X16000805:S0899823X16000805_tab1.gif?pub-status=live)
a Values are shown in mean log10 reductions (95% confidence interval) (No. of samples). Quantitated inoculum was 6.93-log10 per Rodac template for carbapenem-resistant K. pneumoniae and 6.65-log10 per Rodac template for methicillin-resistant S. aureus.
Microbial reduction was calculated by subtracting the number (log10) of test organisms measured on UV-C cycle completion from quantitated inoculum (log10). For design A, the UV-C device was placed at the end of the bed in the center of room with a single cycle (total exposure time: 5 min). For design B, the UV-C device was placed at both sides of the bed in the center of room with a single cycle at the right side and another cycle at left side (total exposure time: 10 min).
Our study demonstrated that the fixed-time UV-C device evaluated in this study was effective in 5–10 minutes in eliminating >5-log10 MRSA and CRKP when the surfaces were in direct line of sight of the UV-C device and >4-log10 MRSA and CRKP when the surfaces were in indirect line of sight. These results are similar to those we have previously published.Reference Rutala, Gergen, Tande and Weber 8 While the use of 2 cycles of the device with placement at either side of the bed for 1 cycle each often resulted in significantly improved inactivation of the vegetative MDROs studied (ie, MRSA, CRKP), a single cycle was highly effective, and the difference is likely not clinically meaningful. Given the frequency of TNTC cultures, our level of microbial kill should be considered a maximum level.
We are aware of several limitations of this study. First, we only inoculated Formica sheets. Our previous study revealed that CRE survived poorly not only on fabrics but also on hard surfaces such as Formica and steel.Reference Weber, Rutala, Kanamori, Gergen and Sickbert-Bennett 9 Second, the size of our study limited our ability to perform subgroup analyses such as inactivation in direct versus indirect line of sight.
In this study, we tested a clinical CRE strain because of the global concern regarding its morbidity and mortality. Importantly, the present study demonstrated that UV-C room decontamination was effective against CRE. The role of hospital environmental surfaces in CRE transmission remains controversial: CRE was infrequently (8%) isolated from environmental surfaces in rooms housing CRE colonized/infected patients in our previous study,Reference Weber, Rutala, Kanamori, Gergen and Sickbert-Bennett 9 while CRE contamination of hospital environment was frequently (88%) identified at surrounding sites among patients with this organism by an Israeli study.Reference Lerner, Adler and Abu-Hanna 7
In conclusion, the present data and our previous findings have demonstrated that UV-C devices can effectively decontaminate hospital surfaces contaminated with MDROs, including CRE. In this experiment, 2 cycles of a UV-C device were often statistically superior to a single cycle, but both designs are likely to be clinically effective (>2-log10 reduction) because epidemiologically important pathogens measured in hospital room surfaces were <100 CFU/Rodac (<2-log10).Reference Huslage, Rutala, Gergen, Sickbert-Bennett and Weber 10