Development of a Mobile Laboratory for Sudden Onset Disasters

Published online by Cambridge University Press:  21 April 2020

Ian Marr*
Affiliation:
National Critical Care and Trauma Response Centre, Darwin, Australia
Joshua R. Francis
Affiliation:
Department of Paediatrics, Royal Darwin Hospital, Darwin, Northern Territory, Australia Menzies School of Health Research, Darwin, Northern Territory, Australia
Dianne P. Stephens
Affiliation:
National Critical Care and Trauma Response Centre, Darwin, Australia Menzies School of Health Research, Darwin, Northern Territory, Australia
Kristy Marshall
Affiliation:
Territory Pathology, Northern Territory, Darwin, Australia
David J. Read
Affiliation:
National Critical Care and Trauma Response Centre, Darwin, Australia
Rob W. Baird
Affiliation:
Territory Pathology, Northern Territory, Darwin, Australia
Nicholas Coatsworth
Affiliation:
Infectious Disease Department, Canberra Hospital, ACT, Australia
*
Correspondence and reprint requests to Dr Ian Marr, c/o Menzies School of Health Research, Darwin, Northern Territory, 0810, Australia (e-mail: Ian_Marr@gmx.com).
Rights & Permissions [Opens in a new window]

Abstract

Objectives:

Clinical diagnostics in sudden onset disasters have historically been limited. We set out to design, implement, and evaluate a mobile diagnostic laboratory accompanying a type 2 emergency medical team (EMT) field hospital.

Methods:

Available diagnostic platforms were reviewed and selected against in field need. Platforms included HemoCue301/WBC DIFF, i-STAT, BIOFIRE FILMARRAY multiplex rt-PCR, Olympus BX53 microscopy, ABO/Rh grouping, and specific rapid diagnostic tests. This equipment was trialed in Katherine, Australia, and Dili, Timor-Leste.

Results:

During the initial deployment, an evaluation of FilmArray tests was successful using blood culture identification, gastrointestinal, and respiratory panels. HemoCue301 (n = 20) hemoglobin values were compared on Sysmex XN 550 (r = 0.94). HemoCue WBC DIFF had some variation, dependent on the cell, when compared with Sysmex XN 550 (r = 0.88-0.16). i-STAT showed nonsignificant differences against Vitros 250. Further evaluation of FilmArray in Dili, Timor-Leste, diagnosed 117 pathogens on 168 FilmArray pouches, including 25 separate organisms on blood culture and 4 separate cerebrospinal fluid pathogens.

Conclusion:

This mobile laboratory represents a major advance in sudden onset disaster. Setup of the service was quick (< 24 hr) and transport to site rapid. Future deployment in fragmented health systems after sudden onset disasters with EMT2 will now allow broader diagnostic capability.

Type
Original Research
Copyright
Copyright © 2020 Society for Disaster Medicine and Public Health, Inc.

Sudden onset disasters (SOD) occur with relative frequency in Asia and the Pacific. Between 2010 and 2014, there were 1446 events that could be deemed as “natural disasters. 1 Significantly, of all people affected by SOD between 1970 and 2014, 87% were located within Asia or the Pacific. 1

The disease burden emanating from SOD is varied, including medical and surgical emergencies (Figure 1). Post-disaster outbreaks of communicable diseases have been widely reported (Figure 2). Reference Watson, Gayer and Connolly2 Diagnostics that can provide basic biochemistry, hematology, and microbiology have the potential to significantly improve quality of care in these settings. Reference Kost, Mecozzi, Brock and Curtis30 - Reference Brock, Mecozzi, Sumner and Kost32 While medical teams and field hospitals have operated with limited diagnostics in the past, the availability of novel point-of-care technology offers the prospect of improved diagnostic precision in the field leading to both clinical and disease surveillance benefits. Reference Coarsey, Esiobu and Narayanan33 , Reference Perkins, Dye and Balasegaram34 Additionally, ABO/Rh grouping for whole “walking blood transfusion” requires screening of donors in accordance with World Health Organization (WHO) guidelines for what has been described as “damage control resuscitation.” 35 , 36

FIGURE 1 Disaster, Associated Pathology and Subsequent Testing Requirements

FIGURE 2 Communicable infections post SOD

Australia’s SOD regional medical response is provided by a WHO verified emergency medical team (EMT) type II field hospital. 35 This includes an emergency care unit (10 chairs, 12 stretchers), a 30-bed inpatient unit, 2 surgical theaters, medical, nursing, allied health staff, and logisticians. The field hospital must be operationally fully self-sufficient.

Current standards for diagnostics in the EMT type II field hospital are covered in the WHO guidelines 35 and include the need for equipment that can measure electrolytes, urea and creatinine, full blood count, blood gas analysis, cross matching, and a limited number of rapid tests, including those for human immunodeficiency virus (HIV), syphilis, and hepatitis B and C.

Historically, the provision of more complex testing in such field hospitals has been difficult. Reference Coatsworth37 Restrictions to regular laboratory processes include disruption of supply routes, difficulty with cold chains, and the use of fragile equipment in environmentally harsh conditions. Reference Louie, Ferguson and Curtis38 The average temperature and humidity in Southeast Asia ranges from 20-35 °C and 70-90%, respectively. This creates potential for variation in temperature sensitive equipment. Reference Louie, Ferguson and Curtis38

Setup of EMT2 hospitals in austere environments must confront these issues. Equipment that does not require temperature regulation is desirable. Further, any laboratory requires an independent supply of electricity, water, and consumables. The laboratory must be adaptable to the disaster situation, mobile, robust, and reliable (see Table 1). With these requirements in mind, we designed a modular mobile laboratory and trialed its function using domestic and international deployments.

TABLE 1 Key Features of the Mobile Laboratory Unit for SOD

METHODS

A number of device platforms were assembled after consultation with field experienced personnel and a review of the literature (Table 2). Parameters required for selection were portability, adaptability, ease of consumable supply, and simplicity of quality control (QC) performance. Standard operating procedures (SOPs) were designed for each platform. A programmatic approach to sample collection, delivery, and processing was created. This was based on operation by 1 senior microbiology trained scientist (as would mimic an infield deployment). All plans incorporated pre-analytical and post-analytical sample processing requirements (based on ISO15189). Stringent screening, drawing, labeling, and issuing standards were enlisted for blood grouping. All non-analytical processes were designed specifically for EMT2 in a SOD environment. 39 A review of rapid diagnostic tests (RDT) was performed and SOD field applicable lateral flow tests chosen for hepatitis B, C, syphilis, HIV, leptospirosis, dengue, and malaria (Table 2). A selection of malaria RDT was based on best operation for Plasmodium falciparum and Plasmodium vivax in the Southeast Asian/Pacific region (as per WHO product testing of malaria RDTs: round 6). Reference Robinson, Wampfler and Betuela40 The BIOFIRE FILMARRAY multiplex real-time (rt) polymerase chain reaction (PCR) system was maintained in an external quality assurance program prior to deployment to ensure correct functionality.

TABLE 2 NCCTRC Mobile Modular Laboratory

mAFB = modified acid-fast bacilli; Pf = Plasmodium falciparum; Pm = Plasmodium malariae; Po = Plasmodium ovale; Pv = Plasmodium vivax;

RG – Risk Group.

# Data on File, BioFire Diagnostics (https://www.biofiredx.com/ [accessed 10/03/2019])

The National Critical Care and Trauma Response Centre’s (NCCTRC) modular mobile laboratory was trialed in 2 locations. The first was a remote hospital in Katherine, Northern Territory (NT), Australia, for 12 days. This hospital is a 60-bed facility located 320 km from the nearest referral center. BIOFIRE® FILMARRAY® multiplex rt-PCR (BioFire Diagnostics, bioMérieux, Salt Lake City, Utah, USA), iSTAT (Abbott, Princeton, New Jersey, USA), HemoCue301, HemoCueWBC DIFF (A Quest Diagnostic Company, Sweden), and the BX53 – Olympus upright microscope were used on pre-collected patient samples by a single microbiology trained staff member. Reproducibility of those systems was tested against accredited validated platforms, including VITROS® 4600 (Ortho-Clinical Diagnostics, Inc., Rochester, NY), Sysmex XN 550 (Sysmex Corporation, Kobe, Japan), and traditional culture-dependent techniques, which included VITEK® 2 (bioMérieux, Marcy l’Étoile, France), VITEK® MS matrix assisted laser desorption ionization time-of-flight (MALDI-TOF) mass spectrometry (bioMérieux, Marcy l’Étoile, France), and where necessary specific in-house PCR at Royal Darwin Hospital, Australia. Blood cultures were performed using both BacT/ALERT® (bioMérieux, Marcy l’Étoile, France) and BACTEC™ (BD, Franklin Lakes, New Jersey, USA) culture media.

The second deployment involved transport of the microbiology module to Dili, Timor-Leste, for operation of the FilmArray system within a national referral laboratory for 84 days by a series of pre-trained Timorese scientists. Training on platform loading, operation, and troubleshooting was completed over 1 day at deployment. This module was embedded within a laboratory with limited microbiology services. Initiation of a blood culture and cerebrospinal fluid (CSF) service occurred for the first time for the local national hospital, Hospital Nacional Guido Valadares (HNGV). FilmArray gastrointestinal and respiratory multiplex panels supplemented pre-existing stool microscopy and influenza testing, respectively. Positive CSF and blood culture samples were run against the corresponding FilmArray panel and concurrently cultured using an on-site incubator and traditional bench top microbiology identification techniques, including API® analytical profile index (bioMérieux, Marcy l’Etoile, France).

All staff involved were trained on biosafety operations prior to use and deployment. Personnel protective equipment (PPE) and biosafety and biosecurity management protocols were implemented for Risk Group 2 organisms. All organisms considered possible Risk Group 3 41 and above were to be isolated and transferred with International Air Transport Association (IATA) Category A packaging for further testing. A special note was made for samples that were thought likely to be containing Mycobacterium tuberculosis. Given the risk to laboratory staff, all high-risk samples were required to be handled in the HEPA filtered film isolator only, and no aerosolization procedures were allowed. QC, necessary for all laboratory diagnostics, occurred on deployment with previously identified organisms for FilmArray, and microscopy. ABO/Rh reagents required strict pre-deployment and in field QC to ensure no diagnostic failures. HemoCue and i-STAT required liquid pre-deployment QC and infield internal QC; however, QC of each RDT batch occurred prior to deployment, not in the field, given the logistics of positive control supply of these infectious agents to a remote site.

During both trials, equipment used uninterruptible power supply (UPS) protected power from the site of deployment rather than generator driven sine wave power as would be provided in a typical disaster scenario. Both sites had temperature and humidity control available, and this was used. Atmosphere control in SOD deployments occurs in a 32-foot Alaska Shelter™ with floor and insulated lining. It has its own environmental control unit and is thermostatically regulated down to 23 °C.

With all equipment assembled, the space required was approximately 3 m x 2.5 m for adequate sample flow (see Table 2 for further details of platforms). Initial costs for laboratory equipment deployment were approximately US $95 000. Ongoing procurement of consumables was estimated at between US $2000 and US $10 000 per deployment (dependent on size and length of period in use).

Ethical approval was obtained from the Cabinet of Quality Control and Ministry of Health for Timor-Leste and the Menzies School of Health and Research (Darwin, NT, Australia), HREC Ref: 2017-2838.

RESULTS

Deployment of all modules to the domestic site (Katherine) was by a single vehicle. Setup at the site was rapid with QC on all platforms started within 2 hours of arrival.

Over the 12-day Australian deployment, 11 FilmArray multiplex rt-PCR pouches were used, including 4 blood culture, 4 gastrointestinal, and 3 respiratory pouches. These detected a total of 9 pathogens with 3 pouches being negative and 1 panel detecting 2 pathogens concurrently. All results were confirmed with approved validated testing platforms, including VITEK 2, VITEKMS, and API. One blood culture sample required a transfer to Royal Darwin Hospital for confirmation as Burkholderia pseudomallei – a pathogen not included on the FilmArray panel. There were 20 HemoCue301 tests performed, all displaying non-clinically significant differences on each parameter when compared with Sysmex XN 550 (r = 0.94). HemoCueWBC DIFF did display some differences on each WBC parameter when compared with Sysmex XN 550. Pearson’s correlation coefficient showed variable effect dependent on the parameter in question (neutrophils r = 0.88, lymphocytes r = 0.49, monocytes r = 0.16, eosinophils r = 0.70, basophils r = 0.16). One sample failed HemoCueWBC DIFF testing and had abnormal film results requiring a detailed BX-53 microscopy review. This case was later diagnosed as chronic lymphocytic leukemia. iSTAT tests were run in parallel against Vitros 250 showing non-clinically significant differences for CHEM4 (n = 10), CG8 (n = 10), and TnI (n = 5) cards. Pearson’s correlation coefficient for Na+, K+, and creatinine were 0.93, 0.95, and 0.99, respectively.

The iSTAT and HemoCue devices fail to perform at high ambient temperatures. The trial at 30 °C caused a machine error requiring the platforms to be reset and cooled for operation at a lower temperature. This is in keeping with known product specifications. The FilmArray and microscopy were not evaluated at these ambient air temperatures and instead were used in a temperature-controlled environment that would mimic the NCCTRC EMT2 Alaska Shelter™.

Deployment to the National Laboratory in Dili, Timor-Leste, occurred on a commercial flight as standard luggage allowance and was operational within 24 hours of arrival. After passing QC, 168 FilmArray pouches were run successfully, with 117 pathogens identified over the period. This included positive blood culture specimens using Oxoid media bottles 59 (Oxoid, Basingstoke, UK) plus an accompanying SIGNAL trigger system, Reference Himmelreich, Orlando and Storch60 CSF, respiratory (nasopharyngeal swabs), and stool samples. All samples were run through corresponding FilmArray multiplex panels. Culture was performed concurrently on blood culture and CSF samples with verification of 92% of culturable organisms using standard bench top tests and API. See Table 3 for further details of organisms isolated.

TABLE 3 BioFire FilmArray Multiplex PCR Results

Safety protocols were audited, and no safety events occurred. The incorporation of SOPs was deemed adequate for the field location, with no major adjustments needed. Malfunctioning rt-PCR runs occurred, with 5 due to power failures (despite functional uninterrupted power supply (UPS), as power delays were persistent not allowing for adequate charging), a further 5 “software errors,” and two pouches that failed to demonstrate vacuum sealing. All samples from failed tests were re-run with normal performance of the assay.

DISCUSSION

Previous use of rapid diagnostics for the NCCTRC was limited. Reference McDermott, Hardstaff and Alpen61 Development of this modular mobile laboratory represents a significant improved diagnostic capability, with an opportunity to impact on patient care. We have shown that the ability to make a reliable and moderately complex diagnosis is now available to an EMT2 unit while attending an SOD.

Adaptability and transportability were a major focus in design. Feasibility of modular deployment of this laboratory means that certain equipment can be sent depending on the support necessary. Setup of the service is quick (< 24 hours), and transport to the site can be rapid. Training for most platforms was straightforward for scientists, and performance was largely consistent in both domestic and international settings. A trained microbiology scientist is deemed necessary for operation, including maintenance of equipment, the skilled microscopy work, and ongoing quality assurance and QC. An interpretation of results requires a background in the basics of microbiology and blood grouping to avoid errors. Use of all modules concurrently is possible for 1 microbiology-trained senior scientist trained in blood grouping; however, the basic operation of i-STAT and HemoCue equipment can be operated by adequately trained nursing staff, lessening the workload on scientists and enabling rapid turnaround times of results.

Infectious diseases can cause significant morbidity and mortality in SODs. This is limited to not only communicable disease spread, but also soft tissue infections, which can be limb- and life-threatening. Antibiotic resistance is now well established in the environment in many Southeast Asian locations. Reference Marr, Sarmento and Brien62 This was seen in data reviewed by Uçkay et al., where repatriated travelers post natural disasters were found to have highly resistant organisms present (including but not restricted to) Enterobacteriaceae, Aeromonas spp., Pseudomonas aeruginosa, Acinetobacter spp., and Burkholderia pseudomallei. 8 Use of our current diagnostics can lead to more accurate treatment and directed antibiotic use for many of these. 8

Currently, disease surveillance in SODs is syndromic and based on presentation features only. This is achieved using data collection tools such as the Surveillance for Post Extreme Emergencies and Disasters (SPEED). Reference Salazar, Law, Pesigan and Winkler63 Supplementation of this information with PCR-based infectious disease diagnostics could enhance disease surveillance in SODs. For many communicable causes of gastroenteritis, on-site PCR will allow for early intervention and limitation of spread by infection control procedures. Reference Ahern, Kovats and Wilkinson12

Transfusion support is also enabled by this laboratory. Rapid blood ABO/Rh grouping with bloodborne virus (BBV) screening allows for a whole blood transfusion expanding capability of surgical and obstetric management options. Current trauma resuscitation guidelines recommend a balanced component therapy (1:1:1) of packed red cells, fresh frozen plasma, and apheresed platelets for massive transfusion. The logistics of stored component therapy, particularly for platelets, can preclude this in the austere environment. The use of whole blood transfusion is well established as a therapeutic option in many low- and middle-income countries (LMICs), 36 is used in humanitarian actions by non-governmental organizations such as the International committee of the Red Cross (ICRC), Reference Giannou and Baldan64 and has been used extensively in military conflicts since World War I. Reference Spinella, Perkins and Grathwohl65 - Reference Spinella, Perkins and Grathwohl67 While not a US Food and Drug Administration approved therapy, they provide a lifesaving management option for massive exsanguination in the restricted field hospital environment that adequately supplies coagulation factors, hemoglobin, and platelets. There are data on successful outcomes in Afghanistan and Iraq, Reference Spinella, Perkins and Grathwohl67 and a pilot study of early whole blood in trauma has been reported in the United States. Reference Cotton, Podbielski and Camp68 The risks of use of whole blood, like any transfusion, should be weighed against the benefits. A relative risk of transmission of infectious agents is dependent on the prevalence in the community in question. A retrospective review of whole blood transfusion products in Iraq by the US military found very few (0.11%) transmissible agents present in any of the units donated. Reference Giannou and Baldan64

Electrolyte and blood differential analysis can give valuable information in the setting of crush injuries (which requires close potassium monitoring) and blood loss, or large electrolyte imbalances. The lack of renal replacement therapy on deployment would require careful monitoring of creatinine to detect any renal impairment arising from crush injury or overwhelming sepsis. The growing burden of non-communicable diseases, including type 2 diabetes, its vascular complications, as well as ischemic heart disease are well known in the Pacific and Southeast Asian regions. Reference Giannou and Baldan64 Rise in this baseline prevalence means that the monitoring of troponin, creatinine, and blood glucose is necessary, widening the scope of electrolytes needed by a mobile laboratory.

The availability of electrolyte and acid base assessment is reported to have proven useful in previous disaster events. Reference Blasetti, Petrucci and Co69 The i-STAT platform was limited by the failure of cartridges at ambient temperature (requiring cold chain if > 2 weeks storage). Alternative options include the epoc® microfluidic blood analysis system (Epocal Inc., Ottawa, Ontario, Canada), which uses room temperature stored cartridges but has similar environmental limitations to platform use. Previous use has shown linearity when compared with both iSTAT and larger biochemistry analyzers in controlled environments. Reference Gilstad, Roschewski and Wells70

Biosafety against potential infectious agents requires thought and prior planning in any laboratory work. Mycobacterium tuberculosis is a bacterium known to cause laboratory acquired infections and can be present in a variety of specimens. Reference Singh71 This as well as other agents deemed high risk are classified by the internationally known Risk Group classification. 41 In our region, this includes (but is not limited to) Risk Group 3: Mycobacterium tuberculosis, Brucella spp., Burkholderia pseudomallei, Neisseria meningitidis, dimorphic fungi; and Risk Group 4 viruses: Hendra and Nipah. Amplifying the inoculum load with use of blood culture exponentially increases transmission risk and hence requires adequate biosafety cabinet/isolator use with HEPA filtration. Any agent in Risk Group 3 or above cannot be managed in this current laboratory and requires transfer via an internationally specified IATA Category A logistics team, if further testing is deemed necessary. Trained scientists are required to know how to identify and avoid further spread of these agents early in the diagnostic process concurrently avoiding all aerosolizing activities. Any use of laboratory equipment without prior planning for these agents risks spread and laboratory staff health. Reference Singh71

Diagnostic errors can also risk patient health. Most diagnostic errors occur in the steps leading up to the sample analysis (pre-analytical) or in the post-analytical period of result transfer and interpretation. Reference Linden, Wagner, Voytovich and Sheehan72 Limitations in pre-analytical and post-analytical collection and processing were controlled in our study with protocols based on international standards and the integrity of sample flow and detailed documentation. 39 Failures in this area are well described in ABO/Rh testing with mismatching of details and samples leading to catastrophic life-ending events (transfusion incompatibility). Reference Linden, Wagner, Voytovich and Sheehan72 Efforts were employed to ensure that this is avoided, with a strict “hemovigilance” pathway implemented that must be followed at all times.

Elements of this mobile laboratory can also prove useful in Australia. With a broad range of diagnostics, it can serve to provide service where an established laboratory may not be functional due to an SOD. The mobile laboratory is capable of temporarily providing rapid on-site results that would have been available pre-disaster at many Australian hospital sites.

There are limitations. These must be understood when considering the use of platforms for near patient testing. Portability can come at the expense of greater analytical imprecision. Reference Kok, Ng and Li42 As our results show, there is lack of correlation with known gold standards on the biochemistry and hematology platforms. Further environmental limitations have been described previously. Reference Louie, Ferguson and Curtis38 Many sites within Asia and the Pacific have high ambient temperatures with high humidity. This is a current limitation to the functional capabilities of at least 3 of the platforms in this lab – see Table 2. Reference Louie, Ferguson and Curtis38 Both of these issues mean use and interpretation needs to be tempered against the possibility of false negatives, false positives, and differences that may or may not be clinically relevant. Given the inherent individual variation of each of these platforms, adequate external QCs are needed and further verification of the result (if available) should be sought. Limitations to RDTs are already well documented. Reference Robinson, Wampfler and Betuela40 Malaria rapid test kits do not currently provide adequate diagnostic’s capability for Plasmodium knowlesi (a potentially deadly malaria present in some locations in Southeast Asia) and lowered sensitivity for Plasmodium malariae and ovale. For this reason, microscopy is still necessary.

Further limitations are due to the requirement for constant power. Electricity requirements for the FilmArray rt-PCR instrument and accompanying laptop include an input voltage 90-264 VAC and input current 10 A. Power supply from the EMT2 field hospital is via a UPS connected to a Genset 22KVa 3 phase diesel generator. Microscopy BX53 requires 12 V for a 100-W lamp illumination and 240 V for digital camera/monitor function and is connected to this same supply. A critical Genset power failure would restrict the use of much of the axillary laboratory setup but still allow for rapid test, iSTAT, and HemoCue use. The iSTAT operation has versatile power requirements but limitations at high ambient temperatures (requires T < 30 °C). Similarly, optimal thermocycling for multiplex rt-PCR requires a temperature-controlled environment with operation between 15-30 °C and 20-80% humidity. This is available in the NCCTRC’s EMT2 designed climate-controlled facility, but this may not be available to all EMT2 teams.

These trials did not test the suitability of the ABO/Rh whole blood transfusion techniques, as there was little role for this at either location. Neither were the RDT/lateral flow assays assessed in field. These have been assessed previously (as per Table 2) and are known also to have variable sensitivity and specificity dependent on the assay in use.

With the acknowledgment of these caveats, this laboratory is seen as a major advance that can help in the diagnosis and management of patients in SODs. Ongoing horizon scanning to detect and evaluate new platforms that may be more suitable to the SOD environment is important.

CONCLUSION

The development of this mobile laboratory represents a significant increase in diagnostic capability for AUSMAT health assets deploying post-SOD. This will likely translate into more accurate care for future patients. This development of individual diagnostic modules allows customization to individual SODs. It is portable, intuitive to learn, and validated on 2 separate high-fidelity trial deployments. It is anticipated that future diagnostic possibilities will continue to expand with the development of novel, potentially more accurate and portable platforms.

Author Contribution

IM, KM, DR, and NC designed laboratory modules; IM, NC, and JF designed and reviewed the studies; IM, DPS, DR, JF, RWB, and NC drafted and reviewed the manuscript.

Financial Support

The company bioMérieux contributed consumable supplies at both sites.

Conflict of Interest Statement

The authors have no conflicts of interest to declare.

References

REFERENCES

ESCAP. Overview of natural disasters and their impacts in Asia and the Pacific, 1970-2014. ESCAP Tech Pap. 2015;1:1-30. http://www.unescap.org/sites/default/files/Technical%20paper-Overview%20of%20natural%20hazards%20and%20their%20impacts_final.pdf.Google Scholar
Watson, JT, Gayer, M, Connolly, MA. Epidemics after natural disasters. Emerg Infect Dis. 2007;13:1-5. doi:10.3201/eid1301.060779.CrossRefGoogle ScholarPubMed
Centers for Disease Control and Prevention. Increase in reported coccidioidomycosis -- United States, 1998-2011. Morb Mortal Wkly Rep. 2013;62(12):217-221.Google Scholar
ReliefWeb. Haiti: Chikungunya outbreak July 2014. 2014. https://reliefweb.int/disaster/ep-2014-000082-hti. Accessed September 26, 2018.Google Scholar
Sáenz, R, Bissell, RA, Paniagua, F. Post-disaster malaria in Costa Rica. Prehosp Disaster Med. 1995;10:154-160. doi:10.1017/S1049023X00041935.CrossRefGoogle ScholarPubMed
Ortiz, RM, Le, NK, Sharma, V, et al. Post-earthquake Zika virus surge: disaster and public health threat amid climatic conduciveness. Sci Rep. 2017;7(1):1-10. doi:10.1038/s41598-017-15706-w.Google Scholar
Uçkay, I, Sax, H, Harbarth, S, et al. Multi-resistant infections in repatriated patients after natural disasters: lessons learned from the 2004 tsunami for hospital infection control. J Hosp Infect. 2008;68:1-8. doi:10.1016/j.jhin.2007.10.018.CrossRefGoogle ScholarPubMed
Centers for Disease Control and Prevention. Two cases of toxigenic Vibrio cholerae O1 infection after Hurricanes Katrina and Rita – Louisiana, October 2005. Morb Mortal Wkly Rep. 2006;55:31-32.Google Scholar
Yee, EL, Palacio, H, Atmar, RL, et al. Widespread outbreak of norovirus gastroenteritis among evacuees of Hurricane Katrina residing in a large ‘megashelter’ in Houston, Texas: lessons learned for prevention. Clin Infect Dis. 2007;44:1032-1039. doi:10.1086/512195.CrossRefGoogle Scholar
Green, JP, Karras, DJ. Update on emerging infections: news from the Centers for Disease Control and Prevention. Notes from the field: fatal fungal soft-tissue infections after a tornado – Joplin, Missouri, 2011. Ann Emerg Med. 2012;59:53-55. doi:10.1016/j.annemergmed.2011.10.004.Google Scholar
Ko, AI, Reis, MG, Dourado, CMR, Johnson, WD, Riley, LW. Urban epidemic of severe leptospirosis in Brazil. Lancet. 1999;354:820-825. doi:10.1016/S0140-6736(99)80012-9.CrossRefGoogle ScholarPubMed
Ahern, M, Kovats, RS, Wilkinson, P, et al. Global health impacts of floods: epidemiologic evidence. Epidemiol Rev. 2005;27:36-46. doi:10.1093/epirev/mxi004.CrossRefGoogle ScholarPubMed
Campanella, N. Infectious diseases and natural disasters: the effects of Hurricane Mitch over Villanueva municipal area, Nicaragua. Public Health Rev. 1999;27:311-319.Google ScholarPubMed
Bechah, Y, Capo, C, Mege, JL, Raoult, D. Epidemic typhus. Lancet Infect Dis. 2008;8:417-426. doi:10.1016/S1473-3099(08)70150-6.CrossRefGoogle ScholarPubMed
Gaspar, M, Leite, F, Brumana, L, et al. Epidemiology of meningococcal meningitis in Angola, 1994-2000. Epidemiol Infect. 2001;127:421-424. doi:10.1017/S0950268801006318.CrossRefGoogle Scholar
World Health Organization. Acute jaundice syndrome. Morb Mortal Wkly Rep. 2006;23:1-14.Google Scholar
Sur, D, Dutta, P, Nair, GB, Bhattacharya, SK. Severe cholera outbreak following floods in a northern district of West Bengal. Indian J Med Res. 2000;112:178-182.Google Scholar
Afshar, M, Raju, M, Ansell, D, Bleck, TP. Narrative review: tetanus – a health threat after natural disasters in developing countries. Ann Intern Med. 2011;154:329-336. doi:10.1059/0003-4819-154-5-201103010-00007.CrossRefGoogle ScholarPubMed
Vollaard, AM, Ali, S, van Asten, HAGH, et al. Risk factors for typhoid and paratyphoid fever in Jakarta, Indonesia. JAMA. 2004;291:2607-2615. doi:10.1001/jama.291.21.2607.CrossRefGoogle ScholarPubMed
Katsumata, T, Hosea, D, Wasito, EB, et al. Cryptosporidiosis in Indonesia: a hospital-based study and a community-based survey. Am J Trop Med Hyg. 1998;59:628-632.10.4269/ajtmh.1998.59.628CrossRefGoogle Scholar
Heath, C, Orrell, C, Lee, R, et al. A review of the Royal Perth Hospital Bali experience: an infection control perspective. Aust Infect Control. 2003;8(2):42-54.10.1071/HI03043CrossRefGoogle Scholar
Italiano, CM, Speers, DJ, Chidlow, GR, et al. Hepatitis B outbreak following a mass-casualty incident, Australia. J Infect Dis. 2011;204:400-407. doi:10.1093/infdis/jir269.CrossRefGoogle ScholarPubMed
Aumentado, C, Cerro, BR, Olobia, L, et al. The prevention and control of dengue after Typhoon Haiyan. Western Pac Surveill Response J. 2015;6(Suppl 1):60-65. doi:10.5365/wpsar.2015.6.3.HYN_018.CrossRefGoogle ScholarPubMed
Surmieda, MR, Lopez, JM, Abad-Viola, G, et al. Surveillance in evacuation camps after the eruption of Mt. Pinatubo, Philippines. Morb Mortal Wkly Rep. 1992;41:9-12.Google ScholarPubMed
Maegele, M, Gregor, S, Yuecel, N, et al. One year ago not business as usual: wound management, infection and psychoemotional control during tertiary medical care following the 2004 Tsunami disaster in southeast Asia. Crit Care. 2006;10(2):R50. doi:10.1186/cc4868.CrossRefGoogle Scholar
World Health Organization. Epidemic-prone disease surveillance and response after the tsunami in Aceh province, Indonesia. Wkly Epidemiol Rec. 2005;80(18):160-164.Google Scholar
Garzoni, C, Emonet, S, Legout, L, et al. Atypical infections in tsunami survivors. Emerg Infect Dis. 2005;11:1591-1593.10.3201/eid1110.050715CrossRefGoogle ScholarPubMed
Brennan, RJ, Rimba, K. Rapid health assessment in Aceh Jaya District, Indonesia, following the December 26 tsunami. Emerg Med Australas. 2005;17:341-350. doi:10.1111/j.1742-6723.2005.00755.x.Google ScholarPubMed
Wynwood, SJ, Craig, SB, Graham, GC, et al. The emergence of Leptospira borgpetersenii serovar Arborea as the dominant infecting serovar following the summer of natural disasters in Queensland, Australia 2011. Trop Biomed. 2014;31:281-285.Google ScholarPubMed
Kost, GJ, Mecozzi, DM, Brock, TK, Curtis, M. Pathogen detection in emergencies and disasters. Point Care. 2013;11:6-15. doi:10.1097/POC.0b013e31825a25cb.Google Scholar
Kost, GJ, Tran, NK, Tuntideelert, M, et al. Katrina, the tsunami, and point-of-care testing. Am J Clin Pathol. 2006;126:513-520. doi:10.1309/NWU5E6T0L4PFCBD9.CrossRefGoogle ScholarPubMed
Brock, TK, Mecozzi, DM, Sumner, S, Kost, GJ. Evidence-based point-of-care tests and device designs for disaster preparedness. Am J Disaster Med. 2011;5:285-294.10.5055/ajdm.2010.0034CrossRefGoogle Scholar
Coarsey, CT, Esiobu, N, Narayanan, R, et al. Strategies in Ebola virus disease (EVD) diagnostics at the point of care. Crit Rev Microbiol. 2017;43:779-798. doi:10.1080/1040841X.2017.1313814.CrossRefGoogle Scholar
Perkins, MD, Dye, C, Balasegaram, M, et al. Diagnostic preparedness for infectious disease outbreaks. Lancet. 2017;390:2211-2214. doi:10.1016/S0140-6736(17)31224-2.CrossRefGoogle ScholarPubMed
Foreign Medical Team Working Group. Classification and minimum standards for foreign medical teams in sudden onset. World Health Organ. 2013;103. http://www.who.int/hac/global_health_cluster/fmt_guidelines_september2013.pdf?ua=1.Google Scholar
World Health Organization. The clinical use of blood – handbook. Vol 1. 2002.Google Scholar
Coatsworth, NR. The Australian medical response to Typhoon Haiyan. Med J Aust. 2014;201:632-634. doi:10.5694/mja14.00306.CrossRefGoogle ScholarPubMed
Louie, RF, Ferguson, WJ, Curtis, CM, et al. Vulnerability of point-of-care test reagents and instruments to environmental stresses: implications for health professionals and developers. Clin Chem Lab Med. 2014;52:325-335. doi:10.1515/cclm-2013-0440.CrossRefGoogle ScholarPubMed
National Association of Testing Authorities, Australia. Medical testing field application document – requirements for accreditation (Medical Testing ISO 15189 Standard Application Document); 2013.Google Scholar
Robinson, LJ, Wampfler, R, Betuela, I, et al. Malaria rapid diagnostic test performance. PLoS Med. 2015;12:e1001891. doi:10.1371/journal.pmed.1001891.CrossRefGoogle Scholar
WHO. Laboratory biosafety manual. 3rd ed. Geneva: World Health Organization; 2004.Google Scholar
Kok, J, Ng, J, Li, SC, et al. Evaluation of point-of-care testing in critically unwell patients: comparison with clinical laboratory analysers and applicability to patients with Ebola virus infection. Pathology. 2015;47:405-409. doi:10.1097/PAT.0000000000000296.CrossRefGoogle Scholar
Baart, AM, de Kort, WLAM, van den Hurk, K, Pasker-de Jong PCM. Hemoglobin assessment: precision and practicability evaluated in the Netherlands – the HAPPEN study. Transfusion. 2016;56:1984-1993. doi:10.1111/trf.13546.CrossRefGoogle Scholar
Tirimacco, R, Simpson, P, Siew, L, et al. Evaluation of the HemoCueWBC DIFF point of care instrument. Australasian Association of Clinical Biochemists. Conference poster. http://www.aacb.asn.au/documents/item/1004.Google Scholar
Salimnia, H, Fairfax, MR, Lephart, PR, et al. Evaluation of the FilmArray Blood Culture Identification Panel: results of a multicenter controlled trial. J Clin Microbiol. 2016;54(3):687-698. doi:10.1128/JCM.01679-15.CrossRefGoogle ScholarPubMed
Popowitch, EB, O’Neill, SS, Miller, MB. Comparison of the biofire filmarray RP, Genmark eSensor RVP, Luminex xTAG RVPv1, and Luminex xTAG RVP fast multiplex assays for detection of respiratory viruses. J Clin Microbiol. 2013;51:1528-1533. doi:10.1128/JCM.03368-12.CrossRefGoogle ScholarPubMed
Buss, SN, Leber, A, Chapin, K, et al. Multicenter evaluation of the BioFire FilmArray gastrointestinal panel for etiologic diagnosis of infectious gastroenteritis. J Clin Microbiol. 2015;53:915-925. doi: 10.1128/JCM.02674-14.CrossRefGoogle ScholarPubMed
Leber, AL, Everhart, K, Balada-Llasat, J, Cullison, J. Multi-center clinical evaluation of a multiplex meningitis/encephalitis PCR panel for simultaneous detection of bacteria, yeast, and viruses in cerebrospinal fluid specimens. J Clin Microbiol. 2016;54:2251-2261. doi:10.1128/JCM.00730-16.CrossRefGoogle Scholar
Xiaodong, S, Tambo, E, Chun, W. Diagnostic performance of CareStartTM malaria HRP2/pLDH (Pf/pan) combo test versus standard microscopy on falciparum and vivax malaria between China-Myanmar endemic borders. Malar J. 2013;12:6. doi:10.1186/1475-2875-12-6.CrossRefGoogle Scholar
Simonnet, C, Okandze, A, Matheus, S, et al. Prospective evaluation of the SD BIOLINE Dengue Duo rapid test during a dengue virus epidemic. Eur J Clin Microbiol Infect Dis. 2017;36:2441-2447. doi:10.1007/s10096-017-3083-8.CrossRefGoogle Scholar
Sehgal, SC, Vijayachari, P, Sugunan, AP, Umapathi, T. Field application of Lepto lateral flow for rapid diagnosis of leptospirosis. J Med Microbiol. 2003;52:897-901. doi:10.1099/jmm.0.05064-0.CrossRefGoogle ScholarPubMed
Dittrich, S, Boutthasavong, L, Keokhamhoung, D, et al. A prospective hospital study to evaluate the diagnostic accuracy of rapid diagnostic tests for the early detection of leptospirosis in Laos. Am J Trop Med Hyg. 2018;98:1056-1060. doi:10.4269/ajtmh.17-0702.CrossRefGoogle ScholarPubMed
Conway, DP, Holt, M, McNulty, A, et al. Multi-centre evaluation of the determine HIV combo assay when used for point of care testing in a high risk clinic-based population. PLoS One. 2014;9(4):1-8. doi:10.1371/journal.pone.0094062.CrossRefGoogle Scholar
Causer, LM, Kaldor, JM, Conway, DP, et al. An evaluation of a novel dual treponemal/nontreponemal point-of-care test for syphilis as a tool to distinguish active from past treated infection. Clin Infect Dis. 2015;61(2):184-191. doi:10.1093/cid/civ243.CrossRefGoogle ScholarPubMed
Causer, LM, Kaldor, JM, Fairley, CK, et al. A laboratory-based evaluation of four rapid point-of-care tests for syphilis. PLoS One. 2014;9(3):1-7. doi:10.1371/journal.pone.0091504.CrossRefGoogle ScholarPubMed
Njai, HF, Shimakawa, Y, Sanneh, B, et al. Validation of rapid point-of-care (POC) tests for detection of hepatitis B surface antigen in field and laboratory settings in Gambia, Western Africa. J Clin Microbiol. 2015;53:1156-1163. doi:10.1128/JCM.02980-14.CrossRefGoogle ScholarPubMed
Kim, MH, Kang, SY, Lee, WI. Evaluation of a new rapid test kit to detect hepatitis C virus infection. J Virol Methods. 2013;193:379-382. doi:10.1016/j.jviromet.2013.07.005.CrossRefGoogle Scholar
Yousuf, R, Abdul Ghani, S, Abdul Khalid, N, Leong, C. Study on ABO and RhD blood grouping: comparison between conventional tile method and a new solid phase method (InTec Blood Grouping Test Kit). Malays J Pathol. 2018;40(1):27-32.Google Scholar
Oxoid Limited, Basingstoke. Oxoid SIGNAL blood culture system product detail. 2010;8-10. https://www.fishersci.com/shop/products/thermo-scientific-oxoid-signal-blood-culture-system-blood-culture-system/r65410.Google Scholar
Himmelreich, CA, Orlando, MF, Storch, GA, et al. Comparison of the oxoid signal blood culture system with supplemented peptone broth in a pediatric hospital. J Clin Microbiol. 1989;27:1262-1265.CrossRefGoogle Scholar
McDermott, KM, Hardstaff, RM, Alpen, S, et al. Management of diabetic surgical patients in a deployed field hospital: a model for acute non-communicable disease care in disaster. Prehosp Disaster Med. 2017;32:657-661. doi:10.1017/S1049023X17006707.CrossRefGoogle Scholar
Marr, I, Sarmento, N, Brien, MO, et al. Antimicrobial resistance in urine and skin isolates in Timor-Leste. Integr Med Res. 2018;13:135-138. doi:10.1016/j.jgar.2017.12.010.Google ScholarPubMed
Salazar, MA, Law, R, Pesigan, A, Winkler, V. Health consequences of Typhoon Haiyan in the Eastern Visayas region using a syndromic surveillance database. PLoS Curr Disasters. 2017;1-14. doi:10.1371/currents.dis.4a3d3b4474847b2599aa5c5eefe3a621.Google ScholarPubMed
Giannou, C, Baldan, M. War Surgery. Vol 1. Geneva, ICRC; 2009.10.1308/rcsann.2010.92.2.179bCrossRefGoogle Scholar
Spinella, PC, Perkins, JG, Grathwohl, KW, et al. Fresh whole blood transfusions in coalition military, foreign national, and enemy combatant patients during Operation Iraqi Freedom at a U.S. combat support hospital. World J Surg. 2008;2-6. doi:10.1007/s00268-007-9201-5.CrossRefGoogle Scholar
Spinella, PC. Warm fresh whole blood transfusion for severe hemorrhage: U.S. military and potential civilian applications. Crit Care Med. 2008;36(7). doi:10.1097/CCM.0b013e31817e2ef9.CrossRefGoogle ScholarPubMed
Spinella, PC, Perkins, JG, Grathwohl, KW, et al. Warm fresh whole blood is independently associated with improved survival for patients with combat-related traumatic injuries. J Trauma. 2009;66. doi:10.1097/TA.0b013e31819d85fb.Google ScholarPubMed
Cotton, BA, Podbielski, J, Camp, E, et al. A randomized controlled pilot trial of modified whole blood versus component therapy in severely injured patients requiring large volume transfusions. Ann Surg. 2013;258:527-532. doi:10.1097/SLA.0b013e3182a4ffa0.CrossRefGoogle ScholarPubMed
Blasetti, AG, Petrucci, E, Co, V, et al. First rescue under the rubble: the medical aid in the first hours after the earthquake in Amatrice. 2017;1-5. doi:10.1017/S1049023X17007075.CrossRefGoogle Scholar
Gilstad, C, Roschewski, M, Wells, J, et al. Fatal transfusion-associated graft-versus-host disease with concomitant immune hemolysis in a group A combat trauma patient resuscitated with group O fresh whole blood. Transfusion. 2012;52:930-935. doi:10.1111/j.1537-2995.2011.03365.x.CrossRefGoogle Scholar
Singh, K. Laboratory acquired infections. Clin Infect Dis. 2009;49(3):142. doi:10.1086/599104.CrossRefGoogle ScholarPubMed
Linden, J, Wagner, K, Voytovich, AE, Sheehan, J. Transfusion errors in New York State: an analysis of 10 years’ experience. Transfusion. 2000;40:1207-1213.10.1046/j.1537-2995.2000.40101207.xCrossRefGoogle ScholarPubMed
Figure 0

FIGURE 1 Disaster, Associated Pathology and Subsequent Testing Requirements

Figure 1

FIGURE 2 Communicable infections post SOD

Figure 2

TABLE 1 Key Features of the Mobile Laboratory Unit for SOD

Figure 3

TABLE 2 NCCTRC Mobile Modular Laboratory

Figure 4

TABLE 3 BioFire FilmArray Multiplex PCR Results