Planned reduction in Fontan fenestration size using the Atrial Flow Regulator

Published online by Cambridge University Press:  12 April 2021

E. Oliver Aregullin
Affiliation:
Congenital Heart Center, Helen DeVos Children’s Hospital, Grand Rapids, Michigan, USA
Bennett Samuel
Affiliation:
Congenital Heart Center, Helen DeVos Children’s Hospital, Grand Rapids, Michigan, USA
Joseph Vettukattil*
Affiliation:
Congenital Heart Center, Helen DeVos Children’s Hospital, Grand Rapids, Michigan, USA
*
Author for correspondence: J. Vettukattil, Congenital Heart Center, Helen DeVos Children’s Hospital, 100 Michigan NE, Grand Rapids, Michigan, USA. Tel: +1 616 267 0988; Fax: +1 616 267 1408; Email: joseph.vettukattil@helendevoschildrens.org
Rights & Permissions [Opens in a new window]

Abstract

Fontan fenestration allows right-to-left shunting increasing cardiac output and oxygen delivery. Increased shunting occurs as cardiac function and ventricular end-diastolic pressures improve, potentially decreasing oxygen saturation. Complete closure may result in impaired Fontan haemodynamics and low cardiac output; however, there are no dedicated devices to reduce fenestration size. We describe Fontan fenestration size reduction using the Atrial Flow Regulator.

Type
Brief Report
Copyright
© The Author(s), 2021. Published by Cambridge University Press

A fenestration in the Fontan pathway placed during surgical palliation lowers impedance to systemic venous return. The right-to-left shunting increases cardiac output and oxygen delivery at the expense of desaturation. Shunting may become functionally intolerable with improvement in cardiac function and ventricular end-diastolic pressures, potentially compromising oxygen saturation.Reference Rychik, Atz and Celermajer1,Reference Saiki, Kuwata and Iwamoto2 There are no dedicated devices for partial closure, and complete closure may result in systemic venous hypertension and multi-organ failure. We describe reducing Fontan fenestration size in a child using the Atrial Flow Regulator (Occlutech International AB, Helsingborg, Sweden).

Case Report

A 5-year-old male born with HLHS, DORV, mitral atresia, obstructed TAPVR, and Kabuki syndrome underwent initial pulmonary artery banding, arch reconstruction, and TAPVR repair shortly after birth. Due to the diagnosis of pulmonary lymphangiectasia and pulmonary interstitial glycogenosis, he eventually required tracheostomy and ventilator support. Takedown of PA band, placement of right modified Blalock–Taussig shunt, and Damus–Kaye–Stansel procedure were performed at 5 months followed by shunt revision at 3 years of age. The patient continued to have complications from chronic lung disease, pulmonary hypertension , and tracheal haemorrhage requiring multiple hospitalisations, and comprehensive PH therapy.

Lateral tunnel Fontan palliation was performed at the age of 5 with a 7 mm fenestration. He was discharged 3 weeks later with oxygen saturations in the mid-80s. The oxygen saturations frequently dropped to 50–60% over time. Although he slowly recovered from persistent pleural effusions, fenestration reduction was planned due to persistent hypoxaemia.

Device and delivery system

The AFR is a nitinol wire mesh-based self-expanding double-disc device with a centrally located fenestration and flexible waist connecting the two discs. The device’s proximal disc includes a welded adapter ball, which is inserted in the jaw of the Pistol Pusher (Occlutech) for implantation using a compatible Delivery Set (Occlutech).

Intervention

Under aseptic precautions and general anaesthesia, right and left heart catheterisation were performed 4 months post-surgery to evaluate haemodynamics. Mean Fontan pressure of 17 mmHg, mean PCWP of 14 mmHg, and aortic saturation of 65% on room air were noted. Test occlusion with a Tyshak Mini (B. Braun Interventional Systems Inc., Bethlehem, PA, USA) balloon showed improvement in oxygen saturation to 95% with mean Fontan pressure of 19 mmHg, which supported the decision to reduce the fenestration size (Table 1). An 8F Delivery Set was positioned in the left atrium over the previously positioned Amplatz Support Wire Guide (Cook Medical, LLC., Bloomington, IN, USA). An AFR with 4 mm fenestration was then deployed using the Pistol Pusher under fluoroscopic and transesophageal echocardiogram guidance in a typical fashion (Figs 1 and 2). Oxygen saturations increased to 92% with stable Fontan pressure: mean 18 mmHg. The patient was kept on intravenous heparin for 24 hours and subsequently switched to oral clopidogrel and aspirin. He was discharged 3 weeks later with a saturation of 91%.

Table 1. Haemodynamic assessment under test conditions

Figure 1. 2DTEE+colour Doppler: 7 mm lateral tunnel Fontan fenestration and shunting from Fontan to atrial chamber.

Figure 2. 2DTEE+colour Doppler: AFR with 4 mm fenestration and shunting from Fontan to atrial chamber.

Discussion

Fontan fenestration decreases morbidity and mortality by increasing cardiac index and systemic oxygen transport, and lowering systemic venous pressures.Reference Lemler, Scott, Leonard, Stromberg and Ramaciotti3,Reference Gewillig, Boshoff, Voelkel and Schranz4 Creation of a fenestration improves outcomes in failing non-fenestrated Fontan circulation.Reference Rupp, Schieke and Kerst5

In our case, Fontan palliation carried high risk due to comorbidities. The alternative option of increasing systemic-to-PA shunt size would have caused worsening haemodynamic profile with higher flow and resistance, eliminating the feasibility of Fontan palliation. A smaller fenestration (4–5 mm) may have led to multi-organ dysfunction from high venous pressure and poor perfusion gradients, possibly resulting in post-operative collapse, and large pleural effusions. Although PA pressures were elevated, the benefits of higher oxygen saturation including a decrease in the development of collaterals outweighed the risk of reduction in fenestration size. An AFR with 4 mm fenestration was selected as there are no approved devices for the reduction of fenestration sizeReference Zablah and Morgan6; however, the device was previously reported for creation of Fontan fenestration.Reference Lehner, Schulze-Neick and Haas7

Conclusion

The AFR device provides a minimally invasive approach to reduce Fontan fenestration size and improve oxygen saturation without a significant increase in Fontan pressures. Further studies are required to evaluate long-term outcomes.

Acknowledgements

None.

Financial support

This research received no specific grant from any funding agency, commercial, or not-for-profit sectors.

Conflicts of interest

Aregullin and Samuel have no conflicts of interest. Vettukattil invented the AFR device and is a paid consultant for the manufacturer, Occlutech.

Ethical standards

The authors assert all procedures contributing to this work comply with ethical standards of relevant national guidelines on human experimentation (US CFR) and with the 1975 Helsinki Declaration (revised 2008). This case report does not meet the definition of research and is not subject to IRB oversight as per the Spectrum Health IRB. Informed consent was obtained from the parents to use the child’s information in online/printed publications.

References

Rychik, J, Atz, AM, Celermajer, DS, et al. Evaluation and management of the child and adult with Fontan circulation: a scientific statement from the American Heart Association. Circulation 2019; 140: e234e284.10.1161/CIR.0000000000000696CrossRefGoogle Scholar
Saiki, H, Kuwata, S, Iwamoto, Y, et al. Fenestration in the Fontan circulation as a strategy for chronic cardioprotection. Heart 2018; 105: 12661272.10.1136/heartjnl-2018-314183CrossRefGoogle Scholar
Lemler, MS, Scott, WA, Leonard, SR, Stromberg, D, Ramaciotti, C. Fenestration improves clinical outcome of the Fontan procedure: a prospective, randomized study. Circulation 2002; 105: 207212.CrossRefGoogle ScholarPubMed
Gewillig, M, Boshoff, D. Missing a sub-pulmonary ventricle: the Fontan circulation. In: Voelkel, N, Schranz, D (eds). The Right Ventricle in Health and Disease. Springer, New York, 2014: 135157.Google Scholar
Rupp, S, Schieke, C, Kerst, G, et al. Creation of a transcatheter fenestration in children with failure of Fontan circulation: Focus on extracardiac conduit connection. Catheter Cardiovasc Interv 2015; 86: 11891194.10.1002/ccd.26042CrossRefGoogle ScholarPubMed
Zablah, JE, Morgan, GJ. Use of Occlutech® atrial flow regulator in a single ventricle patient: a 3D view of a successful intervention. Eur Heart J Cardiovasc Imaging 2020; 21: 1151.CrossRefGoogle Scholar
Lehner, A, Schulze-Neick, I, Haas, NA. Creation of a defined and stable Fontan fenestration with the new Occlutech Atrial Flow Regulator (AFR®). Cardiol Young 2018; 28: 10621066.CrossRefGoogle Scholar
Figure 0

Table 1. Haemodynamic assessment under test conditions

Figure 1

Figure 1. 2DTEE+colour Doppler: 7 mm lateral tunnel Fontan fenestration and shunting from Fontan to atrial chamber.

Figure 2

Figure 2. 2DTEE+colour Doppler: AFR with 4 mm fenestration and shunting from Fontan to atrial chamber.