Hostname: page-component-745bb68f8f-lrblm Total loading time: 0 Render date: 2025-02-06T09:17:22.274Z Has data issue: false hasContentIssue false

Atrial septal defect closure with the new Cardia Ultrasept II™ device with interposed Goretex patch: Mexican experience – has the perforation of Ivalon’s membrane been solved?

Published online by Cambridge University Press:  27 February 2018

Mijangos-Vázquez Roberto*
Affiliation:
Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Tuxtla Gutiérrez, Chiapas, México
García-Montes Antonio J.
Affiliation:
Pediatric Interventional Cardiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Soto-López Elena M.
Affiliation:
Aorta Clinic, Immunology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Guarner-Lans Verónica
Affiliation:
Physiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
Zabal Carlos
Affiliation:
Pediatric Interventional Cardiology Department, National Institute of Cardiology “Ignacio Chávez”, Mexico City, Mexico
*
Author for correspondence: M.-V. Roberto, Pediatric Interventional Cardiology Department, Pediatric Specialties Hospital, Blvd. SS Juan Pablo II y Blvd. Antonio Pariente Algarín S/N, CP. 29070, Tuxtla Gutiérrez, Chiapas. Tel: +52 961 6170700; E-mail: dr.rmijangos@gmail.com
Rights & Permissions [Opens in a new window]

Abstract

Objectives

The objective of this study was to demonstrate the safety and feasibility of using the new Cardia Ultrasept II™ device with interposed Goretex patch referring to the perforation of polyvinyl alcohol membrane.

Background

Great advances have been made in the development of devices for closure of atrial septal defect. The Cardia Ultrasept II™ with interposed Goretex patch is the modified last generation of Cardia devices, having the advantage of a super-low profile within the atria and an integral locking delivery-retrieval mechanism that ensures safe deployment. In addition, with the interposition of the Goretex, it has been possible to abolish perforation of Ivalon’s membrane as a complication.

Methods and results

Patients with ostium secundum atrial septal defect with surrounding rims with a minimum length of 5 mm and who underwent atrial septal defect closure with the new Ultrasept II™ with Goretex patch were included from two paediatric cardiac centres. Primary end point was to determine perforation of the Goretex membrane at follow-up; secondary end point included right ventricular diastolic diameter. In total, 30 patients underwent atrial septal defect closure at a median age of 6 (1–29) years. At follow-up for 6 (range, 1–15) months, freedom from perforations was 100%. A continuous decrease in right ventricular diastolic diameter was found with an initial median of 30 (25–49) mm and after catheterisation of 27.5 (18–33) mm, p=0.01, and Z-score of 2.6 (1.7–3.6) versus 1.9 (1–2.9) after procedure, p=0.01.

Conclusions

The new modified generation of the Ultrasept II™ device with interposed Goretex patch is a good alternative to achieve atrial septal defect closure safely and feasibly with no membrane perforation at follow-up.

Type
Original Articles
Copyright
© Cambridge University Press 2018 

Device closure is considered safe and offers many intuitive advantages over surgical closure, which include avoidance of cardiopulmonary bypass and its potential adverse neurologic sequelae, avoidance of sternotomy scar, a potentially lower incidence of post-procedure complications, and a shorter hospital stay.Reference Du, Hijazi, Kleinman, Silverman and Larntz 1 , Reference Visconti, Bichell, Jonas, Newburger and Bellinger 2

King and Mills in 1974 were the first to use a device to successfully close atrial septal defects.Reference King, Thompson and Steiner 3 Since then, great advances have been made in the elaboration of new devices. The most widely used at present is the Amplatzer™ Septal Occluder device developed by Dr. Kurt Amplatz. In the late 1990s, Cardia Inc. (Eagan, Minnesota, United States of America) developed the PFO-Star device.Reference Schraeder, Fassbender and Strasser 4 All generations of Cardia devices had a low-profile Nitinol frame to minimise device material and enhance flexibility, along with a polyvinyl alcohol sail-like coating, which is said to reduce the risk of thrombus formation. The last generation of these devices is the 7th generation (Ultrasept II™)Reference Turner and Forbes 5 having polyvinyl alcohol (Ivalon)Reference Huang, Kong and Venkatraman 6 coating too on its flower-like rounded disks, which showed relatively favourable results and received CE mark approval. However, relatively early malfunctions of membrane have been reported in several publications.Reference Aubry, Brochet and Halna du Fretay 7 Reference Chamie, Maia and Giuliano 11 In these, the findings have generally been reported within the first months after the device was placed (from 1 week to 10 months after implant). After the publication of these adverse events, the manufacturer announced the development of a new consolidated device with the interposition of a Goretex patch between the two nitinol discs in order to avoid this complication. This device maintains the same name as the previous one, Ultrasept II™.

Materials and methods

Patients

A retrospective study was conducted at two paediatric cardiac centres of all infants who underwent atrial septal defect closure with percutaneous devices. Institutional review board approval was obtained at both centres. Procedures performed from December 2015 to July 2017 were included. Patients were included if they had ostium secundum atrial septal defect, were deemed candidates for atrial septal defect closure by the presence of surrounding rims – superior and inferior caval vein rims – with a minimum length of 5 mm measured by transthoracic echocardiogram, and were brought to atrial septal defect closure using the new generation and modified Cardia Ultrasept II™ device with the interposed Goretex patch. Patients were excluded if they had ostium primum atrial septal defect with or without some extension to the venous sinus, had shorter surrounding rims (<5 mm), and were brought to atrial septal defect closure with different percutaneous devices.

Cardiac catheterisation

An informed consent form was given to the relatives of the patients and a complete left and right heart catheterisation with a double femoral venous access was made under general anaesthesia. Anatomy of the atrial septal defect was thoroughly investigated with intracardiac echocardiogram with an AcuNav 10 Fr catheter (Fig 1). The deployment and release procedures were guided by both fluoroscopy and intracardiac echocardiogram (Fig 2). Intravenous heparin was used during the whole procedure at a dose of 100 U/kg – maximum dose of 5000 UI.

Figure 1 Short-axis intracardiac echocardiogram. ( a ) Measured basal diameter (arrow). ( b ) Catheter expanded diameter (tip of the arrow); the movement of the posteroinferior border can be observed (asterisk).

Figure 2 Final sequence once the Cardia Ultrasept II™ device has been placed. ( a ) Intracardiac echocardiogram. The correct placement of the device that is still being held by the cable (tip of the arrow). ( b ) Minnesota Manoeuvre where the correct position of both disks around the posteroinferior border can be observed (asterisk). ( c ) Intracardiac Echocardiogram. The device after being released (arrow). ( d ) Final position of the device seen by fluoroscopy.

Follow-up

Patients left the hospital 24 hours after the procedure with prophylactic antibiotics and aspirin at a dose of 5 mg/kg, with a maximum dose of 100 mg, during 6 months. During the follow-up, visits were scheduled monthly after the procedure to perform a clinical examination and a transthoracic echocardiogram. We chose monthly visits for the purpose of assessing the integrity of the membrane, as well as the right ventricular diastolic diameter.

Statistical analysis

Numbers and percentages expressed categorical variables. Median and ranges expressed continuous variables. Changes in the dimension of the ventricular chambers were analysed using the Wilcoxon test. The significance was defined as p<0.05. The statistical analysis was performed using the SPSS program and Microsoft Excel 2011 for Mac.

Results

A total of 205 patients with ostium secundum atrial septal defect and with appropriate surrounding borders were treated by catheterisation at two paediatric cardiac centres. Of all patients, we included a total of 30 patients who underwent atrial septal defect closure with the new UItrasept™ II device with Goretex patch interposed. This group included 22 female patients (73.3%) with a group median age of 6 years – ranging from 1 to 29 – and weight of 17 kg – ranging from 7.2 to 90; demographic data are shown in Table 1.

Table 1 Demographic characteristics of the patients.

BMI=body mass index; F=female; M=male

The mean systolic pulmonary arterial pressure was normal at 29.9±8.4 mmHg. The rate of the pulmonary and systemic flow (Qp/Qs) was 2±0.4 and the measured diameter of the septal defect was 14.9±6.6 mm, whereas the balloon expanded diameter of the defect was 16.7±6.7 mm.

The Ultrasept II™ devices used were 8 mm (two patients), 10 mm (three patients), 12 mm (five patients), 14 mm (four patients), 16 mm (three patients), 18, 20 mm (four patients), 22 mm (two patients), 24, 26, 30 mm (two patients), and 32 mm (two patients), with a mean device diameter of 17.5±7.1 mm (Fig 3). All of the devices selected were successfully placed. After the placement of the device, there were no residual shunts evidenced by the angiography or intracardiac echocardiogram.

Figure 3 New Cardia Ultrasept II™ device with interposed Goretex patch.

Median fluoroscopy time was 6 minutes, ranging from 5 to 16.5, and the median procedure time was 48 minutes, ranging from 29 to 125. All patients were discharged 24 hours after the procedure. During follow-up, at 7.1±4.7 months, all patients showed complete closure of the defect and no evidence of perforation was found with transthoracic echocardiogram examination. The longest follow-up was 15 months in two patients (Fig 4) and the shortest follow-up was 1 month. In addition, a continuous decrease of right ventricular diastolic diameter was found with a basal median of 30 mm (25–49) and post-procedure value of 27.5 mm (18–33), p=0.01, and Z-score of 2.6 (1.7–3.6) versus 1.9 (1–2.9) after catheterisation, p=0.01. These data are shown in Table 2.

Figure 4 Transthoracic echocardiogram. ( a , b ) Two-dimensional echo shows complete device integrity. ( c , d ) Color Doppler echo. No perforations observed.

Table 2 Changes at the end of the diastole of the right ventricle diameters before and after the closure of the atrial septal defect using the new Cardia Ultrasept II™ device with Goretex patch.

RVDD=right ventricular diastolic diameter

* To date, the follow-up has not reached 6 months

Discussion

The atrial septal defect is one of the most common heart diseases, comprising 7–10% of all of the CHD.Reference Feltes, Bacha and Beekman 12 The percutaneous closure of the ostium secundum atrial septal defect is actually considered as the elective treatment, and therefore many occluder devices have been developed to accomplish an adequate closure of this kind of defects. Since 1998, Cardia has created prostheses, which have been re-designed in multiple occasions. The Ultrasept II™ is the last generation of these devices created to close defects that measure between 6 and 38 mm in diameter.Reference Huang, Kong and Venkatraman 6

When performing the percutaneous closure of atrial septal defect, several post-procedure complications have been reported. In a review published in 2011 by Butera et alReference Butera, Biondi-Zoccai and Sangiorgi 13 where 1812 patients were catheterised, the percentage of complications was 5.4%, of which only 1.9% were major and in which embolisation and erosion with perforation were discarded. Chan et alReference Chan, Godman, Walsh, Wilson, Redington and Gibbs 14 reported a total of 100 patients with minor complications such as one transient ST elevation, one transient AV block, one presumed deep-vein thrombosis, and one presumed transient ischaemic attack. Chessa et alReference Chessa, Carminati and Butera 15 concluded that atrial septal defect closure is safe, with a low rate of early and late complications. However, some aspects need further and longer investigations, especially the possibility of late thrombus formation on the device, systemic embolisation, and, the most worrisome event, sudden death presented 1.5 years after the device was placed. In addition, with regard to follow-up, no follow-up has been reported in relation to patients who underwent atrial septal defect closure with the Cardia Ultrasept II™ device. However, the maximum limit at which the perforation of the polyvinyl alcohol membrane has been detected is at 10 months after closure.Reference Bhattacharyya, Ilsley and Baltabaeva 8 In our series, there were no post-procedure complications in the short- or medium-term follow-up, and although this is relatively short it far exceeds the usual time of encountering this entity.

It has been reported in several publications about the malfunctioning (perforation) of the polyvinyl alcohol membrane early after the closure of the atrial septal defect causing significant clinical deterioration in some of the patients who have been subjected to closure with the Ultrasept II™ device. Aubry et alReference Aubry, Brochet and Halna du Fretay 7 reported two cases in a series of nine consecutive patients: a 77-year-old man and a 41-year-old woman with dysfunction of 24 and 32 mm prostheses, respectively. Perforation zones in the Ivalon membrane were found with echocardiography 3 months after the implant. Bhattacharyya et alReference Bhattacharyya, Ilsley and Baltabaeva 8 reported the same complication in a 69-year-old man, 10 months after the placement of a 28-mm device. Ramoglu et alReference Ramoglu, Ucar and Tutar 9 also reported a case of early dysfunction after 1 week of the placement of a 20-mm Ultrasept II™ device in a 4-year-old boy, demonstrating the presence of multiple perforations in the polyvinyl alcohol membrane. Finally, LabombardaReference Labombarda, Roule and Beygui 10 reported in 2016 the largest series of patients with the presentation of this complication. They reported a total of four relatively early malfunctions of the polyvinyl alcohol membrane observed in a series of six consecutive patients treated with atrial septal defect Ultrasept II™ closure device from January to December 2014. Chamié et alReference Chamie, Maia and Giuliano 11 informed about four cases of early failure of the polyvinyl alcohol membrane and, in fact, they proposed a new and less invasive treatment for patients who present with early to mid-term failure of the polyvinyl alcohol sail-based devices.

After 2015, Cardia Inc. developed a new version of this same device – Ultrasept II™ – in which a Goretex patch has been interposed between both discs. In our series of 30 cases, patients have been monitored monthly with transthoracic echocardiogram, and at a follow-up of 7.1±4.7months no residual shunts evidencing perforations in the Ivalon membrane have been found. This is probably because of the new Goretex patch placed in the new modified Ultrasept II™ device.

In addition, our follow-up data have corroborated the continuous decrease in the right ventricular diastolic diameter reaching even normal indexed diameters in most of the patients. To our knowledge, there are no reports to date informing the evolution of patients regarding the measurement of these diameters and the Z-score, which might be considered as a direct evidence of improvement after the implantation of the device.

These results might change the perspective of many institutions regarding this device.

Conclusion

The new and modified generation device from Cardia is a safe and effective alternative for atrial septal defect closure. The main difference with other devices is the Ivalon cover and the interposed Goretex patch, which allows for an easy display of the disks, a subtle contact with adjacent structures within the atria while the device is being placed, and, most importantly, avoids major complications such as perforation of the polyvinyl alcohol membrane.

Acknowledgements

The authors thank SUMED for the distribution of Cardia Ultrasept II™ device in our country. The authors also thank Dr. Carlos Zabal, Dr. José Antonio García Montes, and Dr. Juan Pablo Sandoval for the teachings in the technique of closing atrial septal defects.

Financial Support

This research received no specific grant from any funding agency or from commercial or not-for-profit sectors.

Conflicts of Interest

None.

References

1. Du, ZD, Hijazi, ZM, Kleinman, CS, Silverman, NH, Larntz, K, The Amplatzer Investigators. Comparison between transcatheter and surgical closure of secundum atrial septal defect in children and adults: results of a multicenter nonrandomized trial. J Am Coll Cardiol 2002; 39: 18361844.CrossRefGoogle ScholarPubMed
2. Visconti, KJ, Bichell, DP, Jonas, RA, Newburger, JW, Bellinger, DC. Developmental outcome after surgical versus interventional closure of secundum atrial septal defect in children. Circulation 1999; 100 (Suppl): II145II150.Google Scholar
3. King, TD, Thompson, SL, Steiner, C, et al. Secundum atrial septal defect: non operative closure during cardiac catheterization. J Am Med Assoc 1976; 235: 25062509.CrossRefGoogle Scholar
4. Schraeder, R, Fassbender, D, Strasser, RH. PFO-star for closure of patent foramen ovale in patients with presumed paradoxical embolism. In: PS Rao, MJ Kern (eds) Catheter Based Devices for the Treatment of Non-Coronary Cardiovascular Disease in Adults and Children. Lippincott Williams & Wilkins, Philadelphia, PA, USA, 2003: 103109.Google Scholar
5. Turner, DR, Forbes, TJ. Cardia devices. In: Hijazi Z, Feldman T, Abdullah Al-Qbandi MH, et al (eds). Transcatheter Closure of ASDs and PFOs: A Comprehensive Assessment. Cardiotext, Minneapolis, MN, 2010: 407–416.Google Scholar
6. Huang, Y, Kong, JF, Venkatraman, SS. Biomaterials and design in occlusion devices for cardiac defects: a review. Acta Biomaterialia 2014; 10: 10881101.Google Scholar
7. Aubry, P, Brochet, E, Halna du Fretay, X, et al. Early malfunction of polyvinyl alcohol membrane-covered atrial septal defect closure devices. Circ Cardiovasc Intrerv 2014; 7: 721722.Google Scholar
8. Bhattacharyya, S, Ilsley, CD, Baltabaeva, A. Disintegration of polyvinyl alcohol membrane covering atrial septal defect closure device. Eur Heart J 2015: 1153.Google ScholarPubMed
9. Ramoglu, MG, Ucar, T, Tutar, E. Early malfunction of polyvinyl alcohol membrane of septal occluder. Cathet Cardiovasc Intervent 2016; 87: E151E153.Google Scholar
10. Labombarda, F, Roule, V, Beygui, F. Delayed spontaneous perforation of polyvinyl alcohol membrane – covered atrial septal defect closure devices. Catheter Cardiovasc Interv 2016; 89: E141E144.Google Scholar
11. Chamie, F, Maia, J, Giuliano, LC. Device-in-device: a transcatheter alternative to surgical explantation of a failing atrial septal defect intracardiac prosthesis. Catheter Cardiovasc Interv 2016; 88: 239243.Google Scholar
12. Feltes, TF, Bacha, E, Beekman, RH, et al. Indications for cardiac catheterization and intervention in pediatric cardiac disease: a scientific statement from the American Heart Association. Circulation 2011; 123: 26072652.Google Scholar
13. Butera, G, Biondi-Zoccai, G, Sangiorgi, G, et al. Percutaneous versus surgical closure of secundum atrial septal defects: a systematic review and meta-analysis of currently available clinical evidence. Euro Intervention 2011; 7: 377385.Google Scholar
14. Chan, KC, Godman, MJ, Walsh, K, Wilson, N, Redington, A, Gibbs, JL. Transcatheter closure of atrial septal defect and interatrial communications with a new self expanding Nitinol double disc device (Amplatzer septal occluder): multicentre UK experience. Heart 1999; 82: 300306.Google Scholar
15. Chessa, M, Carminati, M, Butera, G, et al. Early and late complications associated with transcatheter occlusion of secundum atrial septal defect. J Am Coll Cardiol 2002; 39: 10611065.Google Scholar
Figure 0

Figure 1 Short-axis intracardiac echocardiogram. (a) Measured basal diameter (arrow). (b) Catheter expanded diameter (tip of the arrow); the movement of the posteroinferior border can be observed (asterisk).

Figure 1

Figure 2 Final sequence once the Cardia Ultrasept II™ device has been placed. (a) Intracardiac echocardiogram. The correct placement of the device that is still being held by the cable (tip of the arrow). (b) Minnesota Manoeuvre where the correct position of both disks around the posteroinferior border can be observed (asterisk). (c) Intracardiac Echocardiogram. The device after being released (arrow). (d) Final position of the device seen by fluoroscopy.

Figure 2

Table 1 Demographic characteristics of the patients.

Figure 3

Figure 3 New Cardia Ultrasept II™ device with interposed Goretex patch.

Figure 4

Figure 4 Transthoracic echocardiogram. (a, b) Two-dimensional echo shows complete device integrity. (c, d) Color Doppler echo. No perforations observed.

Figure 5

Table 2 Changes at the end of the diastole of the right ventricle diameters before and after the closure of the atrial septal defect using the new Cardia Ultrasept II™ device with Goretex patch.